Percentage depth dose evaluation in heterogeneous media using thermoluminescent dosimetry

被引:19
|
作者
da Rosa, L. A. R. [2 ]
Cardoso, S. C. [1 ]
Campos, L. T. [2 ]
Alves, V. G. L. [1 ]
Batista, D. V. S. [3 ]
Facure, A. [4 ]
机构
[1] Univ Fed Rio de Janeiro, Inst Fis, IRD CNEN, BR-21941972 Rio De Janeiro, Brazil
[2] Univ Fed Rio de Janeiro, Inst Radioprotecao & Dosimetria, BR-21941972 Rio De Janeiro, Brazil
[3] INCA, Inst Nacl Canc, Rio De Janeiro, Brazil
[4] Comissao Nacl Energia Nucl, Rio De Janeiro, Brazil
来源
关键词
thermoluminescent detector; lung heterogeneity; small field size; 15; MV; treatment planning system; MONTE-CARLO; STEREOTACTIC RADIOSURGERY; ENERGY; BEAMS; DISTRIBUTIONS; RADIOTHERAPY; ALGORITHMS; TRANSPORT; TLD-100; IMPACT;
D O I
10.1120/jacmp.v11i1.2947
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to investigate the influence of lung heterogeneity inside a soft tissue phantom on percentage depth dose (PDD). PDD curves were obtained experimentally using LiF: Mg,Ti (TLD-100) thermoluminescent detectors and applying Eclipse treatment planning system algorithms Batho, modified Batho (M-Batho or BMod), equivalent TAR (E-TAR or EQTAR), and anisotropic analytical algorithm (AAA) for a 15 MV photon beam and field sizes of 1 x 1, 2 x 2, 5 x 5, and 10 x 10 cm(2). Monte Carlo simulations were performed using the DOSRZnrc user code of EGSnrc. The experimental results agree with Monte Carlo simulations for all irradiation field sizes. Comparisons with Monte Carlo calculations show that the AAA algorithm provides the best simulations of PDD curves for all field sizes investigated. However, even this algorithm cannot accurately predict PDD values in the lung for field sizes of 1 x 1 and 2 x 2 cm(2). An overdos-age in the lung of about 40% and 20% is calculated by the AAA algorithm close to the interface soft tissue/lung for 1 x 1 and 2 x 2 cm(2) field sizes, respectively. It was demonstrated that differences of 100% between Monte Carlo results and the algorithms Batho, modified Batho, and equivalent TAR responses may exist inside the lung region for the 1 x 1 cm(2) field.
引用
收藏
页码:117 / 127
页数:11
相关论文
共 50 条
  • [21] A mammographic phantom to measure mean glandular dose by thermoluminescent dosimetry
    Lopez-Pineda, E.
    Ruiz-Trejo, C.
    Brandan, M. E.
    RADIATION MEASUREMENTS, 2014, 71 : 297 - 299
  • [22] Commercial kitchenware glass as a potential thermoluminescent media for retrospective dosimetry
    Wahib, Norfadira Binti
    Khandaker, Mayeen Uddin
    Ramli, Nur Aqilah Binti Mohamad
    Sani, S. F. Abdul
    Bradley, D. A.
    APPLIED RADIATION AND ISOTOPES, 2019, 148 : 218 - 224
  • [23] REACTOR FLUX MEASUREMENTS USING THERMOLUMINESCENT DOSIMETRY
    MCKLVEEN, JW
    SCHWENK, M
    NUCLEAR TECHNOLOGY, 1976, 31 (02) : 257 - 263
  • [24] Assessment of occupational and patient dose from diagnostic and therapeutic radiation exposure using thermoluminescent dosimetry
    Banu, H
    Alam, MN
    Chowdhury, MI
    Kamal, M
    Bardhan, DK
    Chakraborty, D
    HEALTH PHYSICS, 1998, 74 (04): : 478 - 480
  • [25] PROTON DEPTH-DOSE DOSIMETRY
    MITCHELL, JC
    DALRYMPLE, GV
    WILLIAMS, GH
    HALL, JD
    MORGAN, IL
    RADIATION RESEARCH, 1966, 28 (02) : 390 - +
  • [26] Dosimetric evaluation of electron total skin irradiation using gafchromic film and thermoluminescent dosimetry
    Falahati, Leila
    Nedaie, Hassan Ali
    Esfahani, Mahbod
    Banaee, Nooshin
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 : S115 - S122
  • [27] Evaluation of Transit Dosimetry Software Using Heterogeneous Phantom
    Matulewicz, L.
    Prazmowska, J.
    Stapor-Fudzinska, M.
    Slosarek, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 368 - +
  • [28] Evaluation of Percentage Depth Dose Measurement of High-energy Electron Beams using new TLD Sheet
    Sasaki, K.
    Shiota, Y.
    Miura, M.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S699 - S700
  • [29] Dose optimization of total or partial skin electron irradiation by thermoluminescent dosimetry
    Schuettrumpf, Lars
    Neumaier, Klement
    Maihoefer, Cornelius
    Niyazi, Maximilian
    Ganswindt, Ute
    Li, Minglun
    Lang, Peter
    Reiner, Michael
    Belka, Claus
    Corradini, Stefanie
    STRAHLENTHERAPIE UND ONKOLOGIE, 2018, 194 (05) : 444 - 453
  • [30] Laser heating of thermoluminescent films: Dose mapping applied to beta dosimetry
    Setzkorn, R
    Prevost, H
    Gasiot, J
    Christensen, P
    RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) : 419 - 422