Iodine quantification and detectability thresholds among major dual-energy CT platforms

被引:8
|
作者
Taylor, Ross Edward [1 ]
Mager, Pamela [1 ]
Yu, Nam C. [1 ]
Katz, David P. [1 ]
Brady, Jett R. [1 ]
Gupta, Nakul [1 ]
机构
[1] Houston Methodist Hosp, Dept Radiol, Houston, TX 77030 USA
来源
BRITISH JOURNAL OF RADIOLOGY | 2019年 / 92卷 / 1104期
关键词
COMPUTED TOMOGRAPHIC APPARATUS; TO-NOISE RATIO; ADRENAL NODULES; MULTIDETECTOR CT; RENAL MASSES; ACCURACY; AORTOGRAPHY; PERFORMANCE; PRINCIPLES; BIOMARKERS;
D O I
10.1259/bjr.20190530
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To estimate the minimum detectable iodine concentration on multiple dual-energy CT (DECT) platforms. Methods and materials: A phantom containing iodine concentrations ranging from 0 to 10 mg ml(-1) was scanned with five dual-energy platforms (two rapid kilo volt switching (r-kVs), one dual source (DS), one sequential acquisition and one split-filter). Serial dilutions of 300mg ml( -1) iodinated contrast material were used to generate concentrations below 2 mg ml(-1). Iodine density and virtual monoenergetic images were reviewed by three radiologists to determine the minimum visually detectable iodine concentration. Contrast-to-noise ratios (CNRs) were calculated. Results: 1 mg mL(-1) (-0.8 mg mL(-1) corrected) was the minimum visually detectable concentration among the platforms and could be seen by all readers on the third-generation r-kVs and DS platforms. Conclusions: At low concentrations, CNR for monoenergetic images was highest on the DS platform and lowest in the sequential acquisition and split-filter platforms. Advances in knowledge: The results of this study corroborate previous in vivo estimates of iodine detection limits at DECT and provide a comparison for the performance of different DECT platforms at low iodine concentrations in vitro.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Clinical significance of dual-energy CT-derived iodine quantification in the diagnosis of metastatic LN in colorectal cancer
    Kato, T.
    Uehara, K.
    Ishigaki, S.
    Nihashi, T.
    Arimoto, A.
    Nakamura, H.
    Kamiya, T.
    Oshiro, T.
    Ebata, T.
    Nagino, M.
    EJSO, 2015, 41 (11): : 1464 - 1470
  • [32] Myocardial Late Iodine Enhancement and Extracellular Volume Quantification with DualLayer Spectral Detector Dual-Energy Cardiac CT
    Oda, Seitaro
    Emoto, Takafumi
    Nakaura, Takeshi
    Kidoh, Masafumi
    Utsunomiya, Daisuke
    Funama, Yoshinori
    Nagayama, Yasunori
    Takashio, Seiji
    Ueda, Mitsuharu
    Yamashita, Taro
    Tsujita, Kenichi
    Ando, Yukio
    Yamashita, Yasuyuki
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2019, 1 (01):
  • [33] Editorial Comment: Toward Precise and Reproducible Iodine Quantification With Dual-Energy CT-Getting Closer, but Not There Yet
    Guimaraes, Luis S.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (05) : 840 - 840
  • [34] Use of Dual-Energy CT and Iodine Maps in Evaluation of Bowel Disease
    Fulwadhva, Urvi P.
    Wortman, Jeremy R.
    Sodickson, Aaron D.
    RADIOGRAPHICS, 2016, 36 (02) : 393 - 406
  • [35] Spectral Imaging of Iodine and Gadolinium Nanoparticles Using Dual-Energy CT
    Badea, C. T.
    Holbrook, M.
    Clark, D. P.
    Ghaghada, K.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [36] Iodine Quantification Using Dual-Energy Computed Tomography for Differentiating Thymic Tumors
    Yan, Wei-Qiang
    Xin, Yong-Kang
    Jing, Yong
    Li, Gang-Feng
    Wang, Shu-Mei
    Rong, Wei-Cheng
    Xiao, Gang
    Lei, Xue-Bin
    Li, Bo
    Hu, Yu-Chuan
    Cui, Guang-Bin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 873 - 880
  • [37] Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT
    Gang, Grace J.
    Zbijewski, Wojciech
    Stayman, J. Webster
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2012, 39 (08) : 5145 - 5156
  • [38] Iodine quantification to distinguish hepatic neuroendocrine tumor metastasis from hepatocellular carcinoma at dual-source dual-energy liver CT
    Kaltenbach, Benjamin
    Wichmann, Julian L.
    Pfeifer, Sophia
    Albrecht, Moritz H.
    Booz, Christian
    Lenga, Lukas
    Hammerstingl, Renate
    Dangelo, Tommaso
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 105 : 20 - 24
  • [39] A novel fast kilovoltage switching dual-energy CT with deep learning: Accuracy of CT number on virtual monochromatic imaging and iodine quantification
    Kojima, Tsukasa
    Shirasaka, Takashi
    Kondo, Masatoshi
    Kato, Toyoyuki
    Nishie, Akihiro
    Ishigami, Kousei
    Yabuuchi, Hidetake
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 253 - 261
  • [40] Quantification of Metal Artifact Reduction of Dual-Energy CT: A Phantom Study
    Li, K.
    Leung, W.
    MEDICAL PHYSICS, 2017, 44 (06) : 3036 - 3036