Multicentre validation of IMRT pre-treatment verification: Comparison of in-house and external audit

被引:27
|
作者
Jornet, Nuria [1 ]
Carrasco, Pablo [1 ]
Beltran, Merce [2 ]
Francisco Calvo, Juan [3 ]
Escude, Lluis [4 ]
Hernandez, Victor [5 ]
Quera, Jaume [6 ]
Saez, Jordi [2 ]
机构
[1] Hosp Santa Creu & Sant Pau, Serv Radiofis & Radioproteccio, Barcelona, Spain
[2] Hosp Vall dHebro, Serv Fis, Barcelona, Spain
[3] Hosp Quiron, Serv Oncol Radioterap, Barcelona, Spain
[4] Clin Teknon, Serv Oncol Radioterap, Barcelona, Spain
[5] Hosp St Joan, Serv Fis, Reus, Spain
[6] Hosp Esperanza, Serv Oncol Radioterap, Barcelona, Spain
关键词
IMRT; Audit; Quality assurance; MODULATED RADIATION-THERAPY; RADIOCHROMIC FILM DOSIMETRY; QUALITY-ASSURANCE; EBT2; FILM; RADIOTHERAPY; COMPLEXITY; INDEX; QA;
D O I
10.1016/j.radonc.2014.06.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: We performed a multicentre intercomparison of IMRT optimisation and dose planning and IMRT pre-treatment verification methods and results. The aims were to check consistency between dose plans and to validate whether in-house pre-treatment verification results agreed with those of an external audit. Materials and methods: Participating centres used two mock cases (prostate and head and neck) for the intercomparison and audit. Compliance to dosimetric goals and total number of MU per plan were collected. A simple quality index to compare the different plans was proposed. We compared gamma index pass rates using the centre's equipment and methodology to those of an external audit. Results: While for the prostate case, all centres fulfilled the dosimetric goals and plan quality was homogeneous, that was not the case for the head and neck case. The number of MU did not correlate with the plan quality index. Pre-treatment verifications results of the external audit did not agree with those of the in-house measurements for two centres: being within tolerance for in-house measurements and unacceptable for the audit or the other way round. Conclusions: Although all plans fulfilled dosimetric constraints, plan quality is highly dependent on the planner expertise. External audits are an excellent tool to detect errors in IMRT implementation and cannot be replaced by intercomparison using results obtained by centres. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:381 / 388
页数:8
相关论文
共 50 条
  • [1] In-House IMRT QA Versus External Phantom Audit Results
    Kry, S.
    Molineu, A.
    Alvarez, P.
    Faught, A.
    Huang, J.
    Kerns, J.
    Pulliam, K.
    Tonigan, J.
    Followill, D.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [2] Pre-treatment verification of clinical IMRT fields
    Ercan, T.
    Igdem, S.
    Yildirim, S.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S377 - S378
  • [3] EPID dosimetry - configuration and pre-treatment IMRT verification
    Grzadziel, Aleksandra
    Smolinska, Barbara
    Rutkowski, Roman
    Slosarek, Krzysztof
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2007, 12 (06) : 307 - 312
  • [4] EPID Dosimetry - Configuration and Pre-Treatment IMRT Verification
    Grzadziel, A.
    Smolinska, B.
    Rutkowski, R.
    Slosarek, K.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1932 - 1936
  • [5] Comparison of the Efficacy of 2D Dosimetry Systems in the Pre-treatment Verification of IMRT
    Hong, Chae Seon
    Lim, Jongsoo
    Ju, Sang Gyu
    Shin, Eunhyuk
    Han, Youngyih
    Ahn, Yong Chan
    RADIATION ONCOLOGY JOURNAL, 2009, 27 (02): : 91 - 102
  • [6] Validation of a quick three-dimensional dose verification system for pre-treatment IMRT QA
    Nakaguchi Y.
    Araki F.
    Ono T.
    Tomiyama Y.
    Maruyama M.
    Nagasue N.
    Shimohigashi Y.
    Kai Y.
    Radiological Physics and Technology, 2015, 8 (1) : 73 - 80
  • [7] New imaging devices for pre-treatment plan verification in IMRT
    Slosarek, K.
    Smolinska, B.
    Rutkowski, R.
    Grzadziel, A.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S466 - S466
  • [8] Validation Of Elekta Iviewgt A-Si Epid Model For Pre-Treatment Dose Verification Of Imrt Fields
    Herwiningsih, Sri
    Fielding, Andrew
    4TH INTERNATIONAL CONFERENCE ON ADVANCED COMPUTER THEORY AND ENGINEERING ( ICACTE 2011), 2011, : 35 - 38
  • [9] Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience
    van Zijtveld, Mathilda
    Dirkx, Maarten L. P.
    de Boer, Hans C. J.
    Heijmen, Ben J. M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 (02) : 168 - 175
  • [10] A Statistical Approach to Manage Quality for Pre-Treatment Verification in IMRT/VMAT
    Jassal, K.
    Sarkar, B.
    Mohanti, B.
    Roy, S.
    Ganesh, T.
    Munshi, A.
    Chougule, A.
    Sachdev, K.
    MEDICAL PHYSICS, 2015, 42 (06) : 3348 - 3348