Dynamic behavior of cryogen spray cooling: Effects of spurt duration and spray distance

被引:54
|
作者
Aguilar, G
Wang, GX
Nelson, JS
机构
[1] Univ Calif Irvine, Beckman Laser Inst & Med Clin, Irvine, CA 92612 USA
[2] Univ Akron, Dept Mech Engn, Akron, OH 44325 USA
[3] Univ Calif Irvine, Dept Biomed Engn, Irvine, CA 92697 USA
关键词
cooling selectivity; cryogen layer; cryogen droplets; droplet momentum; surface heat flux; skin cooling;
D O I
10.1002/lsm.10133
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and Objectives: Cryogen spray cooling (CSC) is used to minimize the risk of epidermal damage during laser dermatologic surgery. Since optimization of CSC permits the safe use of higher light doses, which improves therapeutic outcome in many patients with superficial skin lesions, studies have focused on understanding spray-surface interactions and cooling dynamics. The objective of this study is to measure accurately temperature variations at the sprayed surface and the effects of spurt duration (Deltat) and nozzle-to- sprayed surface distance (L) on cooling dynamics during CSC. Study Design/Materials and Methods: A fast-response temperature measurement sensor is built using thin (20 mum) aluminum foil placed on top of a poly methylmethacrylate resin (Plexiglass((R))) with a 50 mum bead diameter thermocouple positioned in between. Liquid film residence time (t(r)) and minimum surface temperature (T-min) are systematically measured as a function of At and L. Results: Two distinct spray-surface interaction mechanisms are recognized. The transition between them occurs at a critical length L-c approximate to25-30 mm. Noticeable characteristics include: (1) for spurts at L < L-c, shorter t(r), and lower T-min are reached as compared to L > L-c, T-min is dependent on At and L, while tr is a function of Deltat only; (2) for spurts at L > L-c, T-min still depends on L but not on Deltat, while tr becomes a function of both Deltat and L. Finally, for all Deltat, t(r) reaches a maximum at L = 40 mm. Conclusions: Based on our results, a good choice to achieve low T-min and t(r) for the treatment of superficial skin lesions may be met by using Deltat of approximate to30-50 milliseconds and the shortest spray distance that is tolerable by the patient. Spurt durations (Deltat) of more than 30-50 milliseconds at spray distances (L) greater than L-c lead to higher T-min and longer t(r). These parameters may be appropriate for laser therapy of deeper targets. Lasers Surg. Med. 32:152-159, 2003. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:152 / 159
页数:8
相关论文
共 50 条
  • [1] Effect of spurt duration on the heat transfer dynamics during cryogen spray cooling
    Aguilar, G
    Wang, GX
    Nelson, JS
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14): : 2169 - 2181
  • [2] Measurements of laser light attenuation following cryogen spray cooling spurt termination
    Edris, A
    Choi, B
    Aguilar, G
    Nelson, JS
    LASERS IN SURGERY AND MEDICINE, 2003, 32 (02) : 143 - 147
  • [3] Effects of spray distance on spray cooling with ammonia
    Sun, Shao-Peng
    Wang, Hong
    Zhu, Xun
    Wang, Shu-Ping
    Yu, Yong-Sheng
    Liao, Qiang
    Shi, Yong
    Kung Cheng Je Wu Li Hsueh Pao/Journal of Engineering Thermophysics, 2010, 31 (04): : 675 - 678
  • [4] Cryogen spray cooling: Effects of droplet size and spray density on heat removal
    Pikkula, BM
    Torres, JH
    Tunnell, JW
    Anvari, B
    LASERS IN SURGERY AND MEDICINE, 2001, 28 (02) : 103 - 112
  • [5] Cryogen spray cooling of human skin: effects of ambient humidity level, spraying distance, and cryogen boiling point
    Anvari, B
    Steeg, BJV
    Milner, TE
    Tanenbaum, BS
    Klein, TJ
    Gerstner, E
    Kimel, S
    Nelson, JS
    MEDICAL APPLICATIONS OF LASERS IN DERMATOLOGY, OPHTHALMOLOGY, DENTISTRY, AND ENDOSCOPY, PROCEEDINGS OF, 1997, 3192 : 106 - 110
  • [6] Cryogen spray cooling: Effects of cryogen film on heat removal and light transmission
    Pikkula, BM
    Domankevitz, Y
    Tunnell, JW
    Anvari, B
    LASERS IN SURGERY: ADVANCED CHARACTERIZATION, THERAPEUTICS, AND SYSTEMS XII, 2002, 4609 : 50 - 56
  • [7] Comparison of cooling criteria with a cryogen spray and water air spray
    Exley, J
    Dickinson, MR
    King, TA
    Charlton, A
    Falder, S
    Kenealy, J
    LASER-TISSUE INTERACTION X: PHOTOCHEMICAL, PHOTOTHERMAL, AND PHOTOMECHANICAL, PROCEEDINGS OF, 1999, 3601 : 130 - 140
  • [8] Evaluation of tissue effects after cryogen spray cooling exposure
    Kao, BS
    Kelly, KM
    Aguilar, G
    Barr, RJ
    Nelson, JS
    LASERS IN SURGERY AND MEDICINE, 2003, : 33 - 33
  • [9] Cutaneous effects of cryogen spray cooling on in vivo human skin
    Datrice, Nicole
    Ramirez-San-Juan, Julio C.
    Zhang, Rong
    Meshkinpour, Azin
    Aguilar, Guillermo
    Nelson, J. Stuart
    Kelly, Kristen M.
    DERMATOLOGIC SURGERY, 2006, 32 (08) : 1007 - 1012
  • [10] Influence of spraying distance and post-cooling on cryogen spray cooling for dermatologic laser surgery
    Aguilar, G
    Majaron, B
    Viator, JA
    Basinger, B
    Karapetian, E
    Svaasand, LO
    Lavernia, EJ
    Nelson, JS
    LASERS IN SURGERY: ADVANCED CHARACTERIZATION THERAPEUTICS, AND SYSTEMS XI, 2001, 4244 : 82 - 92