Improved Brain MR Imaging from a Compact, Lightweight 3T Scanner with High-Performance Gradients

被引:2
|
作者
Camerucci, Emanuele [1 ]
Campeau, Norbert G. [1 ]
Trzasko, Joshua D. [1 ]
Gray, Erin M. [1 ]
Bernstein, Matt A. [1 ]
Cogswell, Petrice M. [1 ]
Shu, Yunhong [1 ]
Foo, Thomas K. [2 ]
Huston, John, III [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] GE Res, Niskayuna, NY USA
关键词
MRI; brain; 3T; ACCURACY;
D O I
10.1002/jmri.27812
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background A low-cryogen, compact 3T (C3T) MRI scanner with high-performance gradients capable of simultaneously achieving 80 mT/m gradient amplitude and 700 T/m/second slew rate has been in use to study research patients since March 2016 but has not been implemented in the clinical practice. Purpose To compare head MRI examinations obtained with the C3T system and a conventional whole-body 3T (WB3T) scanner in seven parameters across five commonly used brain imaging sequences. Study Type Prospective. Subjects Thirty patients with a clinically indicated head MRI. Sequence 3T; T1 FLAIR, T1 MP-RAGE, 3D T2 FLAIR, T2 FSE, and DWI. Assessment All patients tolerated the scans well. Three board-certified neuroradiologists scored the comparative quality of C3T and WB3T images in blinded fashion using a five-point Likert scale in terms of: signal-to-noise ratio, lesion conspicuity, motion artifact, gray/white matter contrast, cerebellar folia, susceptibility artifact, and overall quality. Statistical Test Left-sided, right-sided, and two-sided Wilcoxon signed rank test; Fisher's method. A P value The C3T system performed better than the WB3T in virtually all comparisons, except for motion artifacts for the T1 FLAIR and T1 MP-RAGE sequences, where the WB3T system was deemed better. When combining all sequences together, the C3T system outperformed the WB3T system in all image quality parameters evaluated, except for motion artifact (P = 0.13). Data Conclusion The C3T scanner provided better overall image quality for all sequences, and performed better in all individual categories, except for motion artifact on the T1 FLAIR and T1 MP-RAGE. Level of Evidence 2 Technical Efficacy Stage 1
引用
收藏
页码:166 / 175
页数:10
相关论文
共 50 条
  • [41] High Resolution Chemical Shift Imaging of the Human Brain without Water Suppression at 3T
    Chadzynski, G. L.
    Klose, U.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 177 - 180
  • [42] Improved geometric accuracy of whole body diffusion-weighted imaging at 1.5T and 3T using reverse polarity gradients
    T. Sjöholm
    J. Kullberg
    R. Strand
    M. Engström
    H. Ahlström
    F. Malmberg
    Scientific Reports, 12
  • [43] Improved geometric accuracy of whole body diffusion-weighted imaging at 1.5T and 3T using reverse polarity gradients
    Sjoholm, T.
    Kullberg, J.
    Strand, R.
    Engstrom, M.
    Ahlstrom, H.
    Malmberg, F.
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [44] A High-performance Compact Photoacoustic Tomography System for In Vivo Small-animal Brain Imaging
    Upputuri, Paul Kumar
    Periyasamy, Vijitha
    Kalva, Sandeep Kumar
    Pramanik, Manojit
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2017, (124):
  • [45] Mechanisms and prevention of thermal injury from gamma radiosurgery headframes during 3T MR imaging
    Bennett, Marcus C.
    Wiant, David B.
    Gersh, Jacob A.
    Dolesh, Wendy
    Ding, X.
    Best, Ryan C. M.
    Bourland, J. D.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (04): : 54 - 70
  • [46] MR Spectroscopic Imaging of Peripheral Zone in Prostate Cancer Using a 3T MRI Scanner: Endorectal versus External Phased Array Coils
    Nagarajan, Rajakumar
    Margolis, Daniel J. A.
    Raman, Steven S.
    Ouellette, David
    Sarma, Manoj K.
    Reiter, Robert E.
    Thomas, M. Albert
    MAGNETIC RESONANCE INSIGHTS, 2013, 6 : 51 - 58
  • [47] Intensity Correction Method Customized for Multi-animal Abdominal MR Imaging with 3T Clinical Scanner and Multi-Array Coil
    Mitsuda, Minoru
    Yamaguchi, Masayuki
    Nakagami, Ryutaro
    Furuta, Toshihiro
    Sekine, Norio
    Niitsu, Mamoru
    Moriyama, Noriyuki
    Fujii, Hirofumi
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2013, 12 (02) : 95 - 103
  • [48] Variance components associated with long-echo-time MR spectroscopic imaging in human brain at 1.5T and 3T
    Thrippleton, Michael J.
    Parikh, Jehill P.
    Semple, Scott I. K.
    Harris, Bridget A.
    Andrews, Peter J. D.
    Wardlaw, Joanna M.
    Marshall, Ian
    PLOS ONE, 2017, 12 (12):
  • [49] A rapid method for brain tissue segmentation: Application to quantitative proton turbo MR spectroscopic imaging at 1.5 and 3T
    Horska, A
    Jacobs, MA
    Calhoun, VD
    Barker, PB
    RADIOLOGY, 2002, 225 : 315 - 315
  • [50] Assessment of Vascular Supply of Hypervascular Extra-Axial Brain Tumors with 3T MR Regional Perfusion Imaging
    Sasao, A.
    Hirai, T.
    Nishimura, S.
    Fukuoka, H.
    Murakami, R.
    Kitajima, M.
    Okuda, T.
    Akter, M.
    Morioka, M.
    Yano, S.
    Nakamura, H.
    Makino, K.
    Kuratsu, J. -i.
    Awai, K.
    Yamashita, Y.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (03) : 554 - 558