Quality of Micro-CT Images Acquired from Simultaneous Micro-CT and Benchtop X-Ray Fluorescence Computed Tomography (XFCT): A Preliminary Monte Carlo Study

被引:0
|
作者
Manohar, Nivedh [1 ]
Cho, Sang Hyun [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
NANOPARTICLE-LOADED OBJECTS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Integration of micro-CT with benchtop x-ray fluorescence computed tomography (XFCT) for multimodal imaging poses challenges due to significantly different operating conditions between the two modalities. This Monte Carlo (MC) study investigated the quality of micro-CT images obtainable with incident x-ray spectra that were optimized for benchtop XFCT and, consequently, considerably harder than those typically used for micro-CT. An MC model was created in MCNP5, featuring a cylindrical polymethyl methacrylate phantom (3-cm-diameter) with cylindrical inserts (6-mm-diameter) containing water loaded with gold nanoparticles (2, 1, and 0.5% by weight). The phantom was irradiated by a polychromatic cone-beam x-ray source operated at various tube potentials and filtration. Transmitted photons were scored while the phantom was rotated to generate projections for axial image reconstruction. Reconstructed micro-CT images were comparable to experimentally-obtained images under similar imaging conditions. For a given dose, the contrast and image quality generally decreased with increasing hardness of the incident x-ray spectrum. The results suggest that it would be difficult to achieve the contrast typical of micro-CT images with incident x-ray spectra optimized for benchtop cone-beam XFCT. Further optimization efforts and/or developing alternative imaging schemes (e.g., dual-energy scan) will be necessary to improve micro-CT image quality during simultaneous micro-CT/XF'CT scanning.
引用
收藏
页数:3
相关论文
共 50 条
  • [31] Machining accuracy detection of PCB hole by X-ray micro-CT
    Tao, Wenbin
    Xu, Ying
    Liu, Hao
    Wang, Chengyong
    Zheng, Lijuan
    MICRON, 2020, 131
  • [32] A realistic projection simulator for laboratory based X-ray micro-CT
    Dhaene, Jelle
    Pauwels, Elin
    De Schryver, Thomas
    De Muynck, Amelie
    Dierick, Manuel
    Van Hoorebeke, Luc
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2015, 342 : 170 - 178
  • [33] Improvements in micro-CT method for characterizing X-ray monocapillary optics
    Wang, Zhao
    Pan, Kai
    Du, Zelin
    Zhang, Shuang
    Liu, Zhiguo
    OPTICS COMMUNICATIONS, 2022, 504
  • [34] MICROSTRUCTURE CHARACTERIZATION BY MEANS OF X-RAY MICRO-CT AND NANOINDENTATION MEASUREMENTS
    Rajczakowska, Magdalena
    Stefaniuk, Damian
    Lydzba, Dariusz
    STUDIA GEOTECHNICA ET MECHANICA, 2015, 37 (01) : 75 - 84
  • [35] Resolution and Limitations of X-Ray Micro-CT with Applications to Sandstones and Limestones
    Elkhoury, Jean E.
    Shankar, Raji
    Ramakrishnan, T. S.
    TRANSPORT IN POROUS MEDIA, 2019, 129 (01) : 413 - 425
  • [36] Characterization and analysis of alumina clusters in steel by X-ray Micro-CT
    Zhao, Yi
    Ju, Liying
    Guo, Xiaopei
    Li, Tao
    Tan, Min
    Guo, Han
    Gu, Shaopeng
    JOURNAL OF MATERIALS RESEARCH AND TECHNOLOGY-JMR&T, 2023, 24 : 3252 - 3264
  • [37] Resolution and Limitations of X-Ray Micro-CT with Applications to Sandstones and Limestones
    Jean E. Elkhoury
    Raji Shankar
    T. S. Ramakrishnan
    Transport in Porous Media, 2019, 129 : 413 - 425
  • [38] Environmental cell for in situ X-ray synchrotron micro-CT imaging with simultaneous acoustic measurements
    Drobchik, Arkady N.
    Nikitin, Viktor V.
    Fokin, Mikhail, I
    Dugarov, Geser A.
    Shevchenko, Pavel D.
    Deriy, Alex L.
    Manakov, Andrey Yu
    Kuper, Konstantin E.
    Duchkov, Anton A.
    JOURNAL OF SYNCHROTRON RADIATION, 2022, 29 : 515 - 521
  • [39] Fracture evolution in concrete compressive fatigue experiments based on X-ray micro-CT images
    Skarzynski, L.
    Marzec, I
    Tejchman, J.
    INTERNATIONAL JOURNAL OF FATIGUE, 2019, 122 : 256 - 272
  • [40] Quantitative Study of Porosity and Pore Features in Moldavites by Means of X-ray Micro-CT
    Pratesi, Giovanni
    Caporali, Stefano
    Loglio, Francesca
    Giuli, Gabriele
    Dzikova, Lenka
    Skala, Roman
    MATERIALS, 2014, 7 (04): : 3319 - 3336