A multicentre and multi-national evaluation of the accuracy of quantitative Lu-177 SPECT/CT imaging performed within the MRTDosimetry project

被引:43
|
作者
Tran-Gia, Johannes [1 ]
Denis-Bacelar, Ana M. [2 ]
Ferreira, Kelley M. [2 ]
Robinson, Andrew P. [2 ,3 ,4 ]
Calvert, Nicholas [3 ]
Fenwick, Andrew J. [2 ,5 ]
Finocchiaro, Domenico [6 ,7 ]
Fioroni, Federica [6 ]
Grassi, Elisa [6 ]
Heetun, Warda [2 ]
Jewitt, Stephanie J. [8 ]
Kotzassarlidou, Maria [9 ]
Ljungberg, Michael [10 ]
McGowan, Daniel R. [8 ,11 ]
Scott, Nathaniel [8 ,13 ]
Scuffham, James [2 ,12 ]
Gleisner, Katarina Sjogreen [10 ]
Tipping, Jill [3 ]
Wevrett, Jill [2 ,12 ,13 ]
Lassmann, Michael [1 ]
机构
[1] Univ Wurzburg, Dept Nucl Med, Oberdurrbacher Str 6, D-97080 Wurzburg, Germany
[2] Natl Phys Lab, Teddington, Middx, England
[3] Christie NHS Fdn Trust, Christie Med Phys & Engn CMPE, Manchester, Lancs, England
[4] Univ Manchester, Manchester, Lancs, England
[5] Cardiff Univ, Cardiff, Wales
[6] Azienda Unita Sanit Locale Reggio Emilia IRCCS, Med Phys Unit, Reggio Emilia, Italy
[7] Univ Bologna, Dept Phys & Astron, Bologna, Italy
[8] Oxford Univ Hosp NHS Fdn Trust, Churchill Hosp, Radiat Phys & Protect, Oxford, England
[9] THEAGENIO Anticanc Hosp, Nucl Med Dept, Thessaloniki, Greece
[10] Lund Univ, Med Radiat Phys, Lund, Sweden
[11] Univ Oxford, Dept Oncol, Oxford, England
[12] Royal Surrey Cty Hosp, Guildford, Surrey, England
[13] Univ Surrey, Dept Phys, Guildford, Surrey, England
基金
欧盟地平线“2020”;
关键词
Quantitative SPECT; CT; Lu-177; SPECT; CT imaging; Standardization of SPECT; Harmonization of SPECT; International multicenter comparison exercise; Traceability of SPECT; Molecular radiotherapy (MRT); 3D printing; Phantom; VOLUME CORRECTION TECHNIQUES; DOSIMETRY; KIDNEY; PET; TOMOGRAPHY; PHANTOMS;
D O I
10.1186/s40658-021-00397-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Patient-specific dosimetry is required to ensure the safety of molecular radiotherapy and to predict response. Dosimetry involves several steps, the first of which is the determination of the activity of the radiopharmaceutical taken up by an organ/lesion over time. As uncertainties propagate along each of the subsequent steps (integration of the time-activity curve, absorbed dose calculation), establishing a reliable activity quantification is essential. The MRTDosimetry project was a European initiative to bring together expertise in metrology and nuclear medicine research, with one main goal of standardizing quantitative Lu-177 SPECT/CT imaging based on a calibration protocol developed and tested in a multicentre inter-comparison. This study presents the setup and results of this comparison exercise. Methods The inter-comparison included nine SPECT/CT systems. Each site performed a set of three measurements with the same setup (system, acquisition and reconstruction): (1) Determination of an image calibration for conversion from counts to activity concentration (large cylinder phantom), (2) determination of recovery coefficients for partial volume correction (IEC NEMA PET body phantom with sphere inserts), (3) validation of the established quantitative imaging setup using a 3D printed two-organ phantom (ICRP110-based kidney and spleen). In contrast to previous efforts, traceability of the activity measurement was required for each participant, and all participants were asked to calculate uncertainties for their SPECT-based activities. Results Similar combinations of imaging system and reconstruction lead to similar image calibration factors. The activity ratio results of the anthropomorphic phantom validation demonstrate significant harmonization of quantitative imaging performance between the sites with all sites falling within one standard deviation of the mean values for all inserts. Activity recovery was underestimated for total kidney, spleen, and kidney cortex, while it was overestimated for the medulla. Conclusion This international comparison exercise demonstrates that harmonization of quantitative SPECT/CT is feasible when following very specific instructions of a dedicated calibration protocol, as developed within the MRTDosimetry project. While quantitative imaging performance demonstrates significant harmonization, an over- and underestimation of the activity recovery highlights the limitations of any partial volume correction in the presence of spill-in and spill-out between two adjacent volumes of interests.
引用
收藏
页数:17
相关论文
共 47 条
  • [21] Quantitative SPECT/CT 177Lu-DOTATATE Imaging in Patients with In Vivo Validation
    Sanders, J. C.
    Kuwert, T.
    Ritt, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S194 - S194
  • [22] Quantitative 177Lu-SPECT/CT imaging and validation of a commercial dosimetry software
    D'ambrosio, L.
    Aloj, L.
    Madesani, D.
    Morisco, A.
    Aurilio, M.
    Prisco, A.
    Di Gennaro, F.
    Lastoria, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S263 - S263
  • [23] Dosimetry of Lu-177 DKFZ-PSMA-617 for the treatment of metastatic prostate cancer using quantitative SPECT/CT
    Delker, Andreas
    Fendler, Wolfgang
    Brunegraf, Anika
    Gosewisch, Astrid
    Gildehaus, Franz Josef
    Bartenstein, Peter
    Boening, Guido
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [24] Dosimetry of Lu-177 DKFZ-PSMA-617 for the treatment of metastatic prostate cancer using quantitative SPECT/CT
    Delker, A.
    Fendler, W.
    Brunegraf, A.
    Gosewisch, A.
    Gildehaus, F.
    Bartenstein, P.
    Boening, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S158 - S159
  • [25] Quantitative evaluation of SPECT/CT performance for therapeutic Lu177 radionuclide with NEMA standards
    Dwivedi, P.
    Sawant, V.
    Jha, A.
    Rangarajan, V.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S86 - S87
  • [26] Evaluation of the quantitative accuracy using Monte Carlo simulations in 177Lu imaging
    Kubota, C.
    Ishikawa, K.
    Yasumoto, Y.
    Sakashita, T.
    Daisaki, H.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S739 - S739
  • [27] Investigation of the biodistribution and dosimetry for combined Lu-177/Ac225-PSMA-I&T therapy using multi-isotope quantitative SPECT imaging
    Delker, A.
    Schleske, M.
    Liubchenko, G.
    Berg, I.
    Zacherl, M. J.
    Brendel, M.
    Gildehaus, F. J.
    Harbach, A.
    Rumiantcev, M.
    Resch, S.
    Kaiser, L.
    Brosch-Lenz, J.
    Wenter, V.
    Unterrainer, L.
    Bartenstein, P.
    Ziegler, S.
    Beyer, L.
    Boning, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S126 - S127
  • [28] SPECT/CT Quantification of Lu-177 Activity Using Phantom Images to Develop a Quantitative 3D Dosimetry for Molecular Radionuclide Therapy
    Wiessalla, S.
    Schuchardt, C.
    Baum, R. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S404 - S404
  • [29] Quantitative Assessment Of Accuracy And Precision For High Count Rate Spect Imaging Of 99mtc And 177lu
    Vija, A. H.
    Zeintl, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S16 - S17
  • [30] Quantitative SPECT/CT Imaging of 177Lu with In Vivo Validation in Patients Undergoing Peptide Receptor Radionuclide Therapy
    J. C. Sanders
    T. Kuwert
    J. Hornegger
    P. Ritt
    Molecular Imaging and Biology, 2015, 17 : 585 - 593