Measuring anterior chamber depth with the Orbscan Topography System

被引:81
|
作者
Auffarth, GU
Tetz, MR
Biazid, Y
Volcker, HE
机构
[1] Department of Ophthalmology, Ruprecht-Karls-Univ. of Heidelberg, D-69120 Heidelberg
来源
关键词
D O I
10.1016/S0886-3350(97)80114-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of the Orbscan Topography System in measuring anterior chamber depth (ACD). Setting: Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany. Methods: The Orbscan Topography System is a 3-D, scanning slit beam system for analyzing corneal surfaces as well as structures of the anterior segment, e.g., iris, lens. Reflected surface data points are measured in the x, y, and z axes, creating color-coded true surface topography maps. We measured ACD prior to cataract surgery in 56 eyes of 37 patients (mean age 70.4 years + 13.2 [SD]) using the Orbscan system and, as a control, the Grieshaber Biometric System (ultrasound) and the Jager ACD slitlamp measurement system. Results: Mean ACD values measured with the Orbscan system were 3.23 +/- 0.55 mm; ultrasound measurements were 3.27 +/- 0.54 mm and Jager measurements, 3.10 +/- 0.55 mm. The average difference between Orbscan and ultrasound values was 0.04 +/- 0.15 mm. The correlation coefficient between Orbscan and ultrasound was 0.96 (P < .00001) and between Orbscan and Jager, 0.899 (P < .0001). Conclusion: The ACD measurements by the Orbscan system were equivalent to common reference measurements.
引用
收藏
页码:1351 / 1355
页数:5
相关论文
共 50 条
  • [21] Central Corneal Thickness, Anterior Chamber Depth, and Pupil Diameter Measurements Using Visante OCT, Orbscan, and Pentacam
    Yazici, Ahmet Taylan
    Bozkurt, Ercument
    Alagoz, Cengiz
    Alagoz, Nese
    Pekel, Gokhan
    Kaya, Vedat
    Yilmaz, Omer Faruk
    JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (02) : 127 - 133
  • [22] Comparison of three methods of measuring corneal thickness and anterior chamber depth
    Buehl, W
    Stojanac, D
    Sacu, S
    Drexler, W
    Findl, O
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) : 7 - 12
  • [23] Keratoconus evaluation using the Orbscan Topography System
    Auffarth, GU
    Wang, L
    Völcker, HE
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (02): : 222 - 228
  • [24] Validation of the anterior chamber angle measurements with the Orbscan II system using two extrapolation methods
    Llorens, DP
    Moreno-Montañes, J
    Diez-Cuenca, M
    Fernandez, P
    Rodriguez-Conde, R
    Maldonado, MJ
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U675 - U675
  • [25] Anterior chamber depth studies
    Hoffer, Kenneth J.
    Savini, Giacomo
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (09): : 1898 - 1904
  • [26] Accuracy and repeatability of the Orbscan corneal topography system.
    Beyrouthy, M
    Brunette, I
    Horner, D
    Munger, R
    Demers, P
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2001, 42 (04) : S895 - S895
  • [28] Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode
    B. Meinhardt
    O. Stachs
    J. Stave
    R. Beck
    R. Guthoff
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2006, 244 : 559 - 564
  • [29] Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode
    Meinhardt, B
    Stachs, O
    Stave, J
    Beck, R
    Guthoff, R
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 244 (05) : 559 - 564
  • [30] Accuracy and precision of the Orbscan topography unit in measuring standardized radially aspheric surfaces
    Taylor, CT
    Sutphin, JE
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 2576 - 2576