A portal image based quality assurance program using a commercial portal dosimetry software

被引:0
|
作者
Guerrero, M. [1 ]
Patel, I. [1 ]
机构
[1] Dept Vet Affairs Med Ctr, Washington, DC USA
关键词
D O I
10.1118/1.2760375
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: There is an increasing trend in Radiation Oncology Departments to replace film as a tool for localization hence Medical Physicists increasingly find themselves without access to a film processor. Films are typically used for standard routine quality assurance (QA) of linear accelerators and IMRT commissioning and verifications. The purpose of this work is to develop a QA program using a commercial electronic portal imager (EPID) and commercial portal dosimetry software available on Varian Linear Accelerators. Method and Materials: Using Varian aS500 Portal Imager with the Portal Dosimetry Calculation (PDC) from Eclipse (commissioned according to manufacturer's recommendations), we perform a series of measurements associated with Linac QA (light‐field coincidence, flatness and symmetry, off‐axis ratios, etc), IMRT verifications and static and dynamic MLC QA. For light‐filed coincidence we use the gradicule to “mark” the edges of the light field. For flatness and symmetry we determine a “baseline” using a phantom measurement at 10cm depth and establishing a correspondence with the Portal Imager measurement. Similar baselines are established for off‐axis ratios and other measurements. Results: We present the results of our QA measurements; analyze their level of difficulty and convenience as compared with the use of film. The QA measurements performed with the portal imager are consistent with the results from our film measurements. The IMRT verifications performed with the EPID are also consistent with the ones performed with our commercial two‐dimensional diode array. We find that after we became familiar with the use of the Portal Imager and the software, the QA measurements with the portal imager are adequate and in many cases more convenient than the use of film. Conclusion: An appropriate Linac and IMRT QA program can be developed using the Varian aS500 EPID in combination with the Eclipse Portal Dosimetry Calculation. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2337 / 2337
页数:1
相关论文
共 50 条
  • [41] A quality assurance phantom for electronic portal imaging devices
    Das, Indra J.
    Cao, Minsong
    Cheng, Chee-Wai
    Misic, Vladimir
    Scheuring, Klaus
    Schuele, Edmund
    Johnstone, Peter A. S.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 391 - 403
  • [42] Analyzing the Quality of the SBRT Treatments Using the Portal Dosimetry (PD) Device.
    Wang, Y.
    Wittenbach, K.
    Jackson, M.
    Ma, J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7884 - 7885
  • [43] Quality assurance of electron beams using a Varian electronic portal imaging device
    Wang, Y.
    Heaton, R.
    Norrlinger, B.
    Islam, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (16): : 5461 - 5475
  • [44] Commissioning and quality assurance of commercial in vivo dosimetry systems
    Esposito, M.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1 - S2
  • [45] Treatment-based portal dosimetry for VMAT using integrated images
    Bedford, J. L.
    Hanson, I. M.
    Hansen, V. Nordmark
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S176 - S177
  • [46] Patient Specific Quality Assurance of Volumetric Modulated Arc Therapy (VMAT): Portal Dosimetry Compared with ArcCheck Phantom
    Utitsarn, K.
    Watthanasarn, T.
    Pimthong, J.
    Krongkietlearts, K.
    Pihusut, C.
    Chaloemchawalit, W.
    Jitmon, J.
    Khuanchana, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E826 - E826
  • [47] Retrospective analysis of portal dosimetry pre-treatment quality assurance of hybrid IMRT breast treatment plans
    Koo, Meghan
    Darko, Johnson
    Osei, Ernest
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (01) : 22 - 29
  • [48] Importance of beam-matching between truebeam stx and novalis tx in pre-treatment quality assurance using portal dosimetry
    Alejandro Rojas-Lopez, Jose
    Venencia, Daniel
    JOURNAL OF MEDICAL PHYSICS, 2021, 46 (03) : 211 - 220
  • [49] A QUALITY ASSURANCE PROGRAM FOR SOFTWARE MAINTENANCE
    CENTER, JW
    AFIPS CONFERENCE PROCEEDINGS, 1982, 51 : 399 - &
  • [50] Validation of Portal Dosimetry As a Patient Specific Quality Assurance (PSQA) Device for Arc-Based Plans in Terms of Robustness to Known Errors
    Badry, F.
    Grelewicz, Z. R.
    Liao, Y.
    Turian, J. V.
    MEDICAL PHYSICS, 2024, 51 (10) : 7748 - 7749