Electron beam quality correction factors for plane-parallel ionization chambers: Monte Carlo calculations using the PENELOPE system

被引:96
|
作者
Sempau, J [1 ]
Andreo, P
Aldana, J
Mazurier, J
Salvat, F
机构
[1] Univ Politecn Cataluna, Inst Tecn Energet, Barcelona, Spain
[2] Stockholm Univ, Karolinska Inst, S-10691 Stockholm, Sweden
[3] Ctr Hautes Energies, Serv Radiotherapie, Toulouse, France
[4] Univ Barcelona, Fac Fis, ECM, E-08028 Barcelona, Spain
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2004年 / 49卷 / 18期
关键词
D O I
10.1088/0031-9155/49/18/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Simulations of three plane-parallel ionization chambers have been used to determine directly the chamber-and quality-dependent factors fc,(Q,) instead of the product (S-w,(air) p)(Q), and k(Q).(Q), (or k(Q),(Qint)) for a broad range of electron beam qualities (4-20 MeV) using divergent monoenergetic beams and phase-space data from two accelerators. An original calculation method has been used which circumvents the weakness of the so far assumed independence between stopping-power ratios and perturbation factors. Very detailed descriptions of the geometry and materials of the chambers have been obtained from the manufacturers, and prepared as input to the PENELOPE 2003 Monte Carlo system using a computer code that includes correlated sampling and particle splitting. Values of the beam quality factors have been determined for the case of an electron reference beam. The calculated values have been compared with those in the IAEA TRS-398 dosimetry protocol and the differences analysed. The results for a NACP-02 chamber show remarkably good agreement with TRS-398 at high electron beam qualities but differ slightly at low energies. Arguments to explain the differences include questioning the undemonstrated assumption that the NACP is a 'perturbation-free' chamber even at very low electron beam energies. Results for Wellhofer PPC-40 and PPC-05 chambers cannot be compared with data from others for these chambers because no calculations or reliable experimental data exist. It has been found that the results for the PPC-40 are very close to those of a Roos chamber, but the values for the PPC-05 are considerably different from those of a Markus chamber, and rather approach those of a Roos chamber. Results for monoenergetic electrons and accelerator phase-space data have been compared to assess the need for detailed and costly simulations, finding very small differences. This questions the emphasis given in recent years to the use of 'realistic' source data for accurate electron beam dosimetry.
引用
收藏
页码:4427 / 4444
页数:18
相关论文
共 50 条
  • [31] Calculations of magnetic field correction factors for ionization chambers in a transverse magnetic field using Monte Carlo code TOPAS
    Mao, Lingli
    Pei, Xi
    Chao, Tsi-Chian
    Xu, Xie George
    RADIATION PHYSICS AND CHEMISTRY, 2021, 183
  • [32] Validation of a Monte Carlo model of a NACP-02 plane-parallel ionization chamber model using electron backscatter experiments
    Chin, E.
    Shipley, D.
    Bailey, M.
    Seuntjens, J.
    Palmans, H.
    DuSautoy, A.
    Verhaegen, F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (08): : N119 - N126
  • [33] EXPERIMENTAL-DETERMINATION OF WALL CORRECTION FACTORS .2. NACP AND MARKUS PLANE-PARALLEL IONIZATION CHAMBERS
    WITTKAMPER, FW
    AALBERS, AHL
    MIJNHEER, BJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04): : 995 - 1004
  • [34] Measurement of Prepl Pwall factors in electron beams and in a 60Co beam for plane-parallel chambers
    Ding, GX
    Cygler, JE
    MEDICAL PHYSICS, 1998, 25 (08) : 1453 - 1457
  • [35] Photon beam quality correction factors for the NE2571A and NE2581A thimble ionization chambers using PENELOPE
    Erazo, Fabian
    Lallena, Antonio M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (01): : 232 - 236
  • [36] Monte Carlo calculations for the distribution function of energetic electrons in a plane-parallel hollow cathode
    Hashiguchi, Seishiro
    Hasikuni, Mitugi
    Japanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers, 1988, 27 (10): : 2007 - 2008
  • [37] POLARITY EFFECT IN PLANE-PARALLEL IONIZATION CHAMBERS USING AIR OR A DIELECTRIC LIQUID AS IONIZATION MEDIUM
    WICKMAN, G
    HOLMSTROM, T
    MEDICAL PHYSICS, 1992, 19 (03) : 637 - 640
  • [38] Determination of kQ factors for cylindrical and plane-parallel ionization chambers in a scanned carbon ion beam by means of cross calibration
    Osinga-Blaettermann, J-M
    Krauss, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (01):
  • [39] Monte Carlo calculation of magnetic field correction factors for two ionization chambers
    Alissa, M.
    Zink, K.
    Schoenfeld, A. A.
    Czarneck, D.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1310 - S1311
  • [40] Monte-Carlo-based perturbation and beam quality correction factors for thimble ionization chambers in high-energy photon beams
    Wulff, J.
    Heverhagen, J. T.
    Zink, K.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11): : 2823 - 2836