Dosimetric comparison of peripheral NSCLC SBRT using Acuros XB and AAA calculation algorithms

被引:15
|
作者
Ong, Chloe C. H. [1 ]
Ang, Khong Wei [2 ]
Soh, Roger C. X. [3 ]
Tin, Kah Ming [2 ]
Yap, Jerome H. H. [2 ]
Lee, James C. L. [2 ,3 ]
Bragg, Christopher M. [1 ]
机构
[1] Sheffield Hallam Univ, Fac Hlth & Wellbeing, Sheffield, S Yorkshire, England
[2] Natl Canc Ctr Singapore, Div Radiat Oncol, Singapore, Singapore
[3] Nanyang Technol Univ, Sch Phys & Math Sci, Div Phys & Appl Phys, Singapore, Singapore
关键词
SBRT; AXB; AAA; Calculation algorithms; NSCLC; DOSE CALCULATION; RADIATION-THERAPY; LUNG; CANCER;
D O I
10.1016/j.meddos.2017.05.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
There is a concern for dose calculation in highly heterogenous environments such as the thorax region. This study compares the quality of treatment plans of peripheral non small cell lung cancer (NSCLC) stereotactic body radiation therapy (SBRT) using 2 calculation algorithms, namely, Eclipse Anisotropic Analytical Algorithm (AAA) and Acuros External Beam (AXB), for 3-dimensional conformal radiation therapy (3DCRT) and volumetric-modulated arc therapy (VMAT). Four-dimensional computed tomography (4DCT) data from 20 anonymized patients were studied using Varian Eclipse planning system, AXB, and MA version 10.0.28. A 3DCRT plan and a VMAT plan were generated using AAA and AXB with constant plan parameters for each patient. The prescription and dose constraints were benchmarked against Radiation Therapy Oncology Group (RTOG) 0915 protocol. Planning parameters of the plan were compared statistically using Mann-Whitney U tests. Results showed that 3DCRT and VMAT plans have a lower target coverage up to 8% when calculated using AXB as compared with AAA. The conformity index (CI) for AXB plans was 4.7% lower than AAA plans, but was closer to unity, which indicated better target conformity. AXB produced plans with global maximum doses which were, on average, 2% hotter than AAA plans. Both 3DCRT and VMAT plans were able to achieve D95%. VMAT plans were shown to be more conformal (CI = 1.01) and were at least 3.2% and 1.5% lower in terms of PTV maximum and mean dose, respectively. There was no statistically significant difference for doses received by organs at risk (OARs) regardless of calculation algorithms and treatment techniques. In general, the difference in tissue modeling for AXB and AAA algorithm is responsible for the dose distribution between the AXB and the AAA algorithms. The AXB VMAT plans could be used to benefit patients receiving peripheral NSCLC SBRT. (C) 2017 American Association of Medical Dosimetrists.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 50 条
  • [1] Dosimetric comparison of AAA and Acuros XB algorithms for lung stereotactic radiotherapy
    Senkesen, O.
    Goksel, E.
    Kucucuk, H.
    Yilmaz, M.
    Tezcanli, E.
    Garipagaoglu, M.
    Sengoz, M.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S183 - S183
  • [2] Comparison of AAA and Acuros XB Dose Calculation Algorithms for SRS
    Yeh, F.
    Adamson, J.
    Wang, Z.
    MEDICAL PHYSICS, 2020, 47 (06) : E725 - E725
  • [3] Dosimetric comparison of the two dose reporting modes of Acuros XB and AAA for lung SBRT
    Mampuya, A. W.
    Nakamura, M.
    Hirose, Y.
    Ishigaki, T.
    Mizowaki, T.
    Hiraoka, M.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S718 - S718
  • [4] Dosimetric Impact of Using the Acuros XB for Planning of SBRT of Centrally Located NSCLC
    Hung, H. M.
    JOURNAL OF THORACIC ONCOLOGY, 2021, 16 (03) : S274 - S275
  • [5] Dosimetric impact of Acuros XB Dose Calculation Algorithm on lung SBRT treatments
    Munoz Montplet, C.
    Agramunt Chaler, S.
    Jurado Bruggeman, D.
    Romera Martinez, I.
    Oliva, E.
    Fuentes, R.
    Rosello, A.
    Aunon, C.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S772 - S773
  • [6] Dosimetric Impact of Calculation Grid Size for Spine SBRT Plans Using Acuros XB Algorithm
    Badkul, R.
    Hart, D.
    MEDICAL PHYSICS, 2020, 47 (06) : E756 - E757
  • [7] Dosimetric comparison of AAA and Acuros XB algorithm for breast DIBH VMAT treatment
    Deshpande, S.
    Naidu, S.
    Chavan, K. N.
    Kannan, V.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S741 - S742
  • [8] Influence of Acuros XB and AAA Dose Calculation Algorithms On Plan Quality Metrics and Normal Lung Doses in Lung SBRT
    Yaparpalvi, R.
    Mynampati, D.
    Kuo, H.
    Garg, M.
    Tome, W.
    Kalnicki, S.
    MEDICAL PHYSICS, 2016, 43 (06) : 3601 - 3602
  • [9] Dosimetric accuracy of Acuros® XB and AAA algorithms for stereotactic body radiotherapy (SBRT) lung treatments: evaluation with PRIMO Monte Carlo code
    Sarin, B.
    Bindhu, B.
    Kumar, P. Raghu
    Sumeesh, S.
    Saju, B.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 22
  • [10] Dosimetric comparison of Acuros XB, AAA, and XVMC in stereotactic body radiotherapy for lung cancer
    Tsuruta, Yusuke
    Nakata, Manabu
    Nakamura, Mitsuhiro
    Matsuo, Yukinori
    Higashimura, Kyoji
    Monzen, Hajime
    Mizowaki, Takashi
    Hiraoka, Masahiro
    MEDICAL PHYSICS, 2014, 41 (08) : 189 - 197