Improving the dynamic measurements of ADCs using the 2-ADC method

被引:2
|
作者
Cauvet, P [1 ]
机构
[1] Digital Media Philips Semicond, Test & Prod Engn Dept, F-14043 Caen 5, France
关键词
analog-to-digital converter (ADC); jitter; aperture uncertainty; signal-to-noise ratio (SNR);
D O I
10.1016/S0920-5489(00)00048-9
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
Testing high-resolution analog-to-digital converters (greater than or equal to 12 bit) at high input frequencies (>5 MHz) becomes more difficult, because the disturbances of the automatic test equipment (ATE) become more significant. Although a lot of improvements have been made to modern ATEs during recent years, they are not sufficient, given the components specifications, especially the signal-to-noise ratio (SNR) parameter. This paper briefly describes the errors of the converters, which have to be tested, and those introduced by the equipment. From this description, a model has been built to simulate a new method based on the simultaneous measurement of two converters and which is intended to give an accurate estimation of the SNR, removing the correlated disturbances. The simulation results show the potential benefits of the method, and eventually, the experimental results in an industrial environment are given and discussed. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:281 / 286
页数:6
相关论文
共 50 条
  • [21] A Method for ADC Error Testing and its Compensation in Ratiometric Measurements
    Hariharan, K.
    Vasanthakumar, P.
    Varun, G.
    Abhaikumar, V.
    MEASUREMENT SCIENCE REVIEW, 2010, 10 (02): : 56 - 59
  • [22] Behavioral Modeling of the Static Transfer Function of ADCs Using INL Measurements
    Guindi, Rafik S.
    Saada, Ehal H.
    2010 INTERNATIONAL CONFERENCE ON MICROELECTRONICS, 2010, : 108 - 111
  • [23] Improving Java']Java performance using dynamic method migration on FPGAs
    Lattanzi, Emanuele
    Gayasen, Aman
    Kandemir, Mahmut
    Vijaykrishnan, N.
    Benini, Luca
    Bogliolo, Alessandro
    INTERNATIONAL JOURNAL OF EMBEDDED SYSTEMS, 2005, 1 (3-4) : 228 - 236
  • [24] Testing high resolution ADCS using deterministic dynamic element matching
    Olleta, B
    Jiang, H
    Chen, D
    Geiger, R
    2004 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOL 1, PROCEEDINGS, 2004, : 920 - 923
  • [25] METHOD OF MEASURING ADC ERROR CHARACTERISTICS UNDER DYNAMIC CONDITIONS
    BAKIROV, AA
    ZAIKO, AI
    MEASUREMENT TECHNIQUES USSR, 1985, 28 (01): : 10 - 11
  • [26] SOLUBILITY MEASUREMENTS USING A SEALED AMPOULE - DYNAMIC VS STATIC METHOD
    SATTERFIELD, RG
    HAULARD, M
    JOURNAL OF CHEMICAL AND ENGINEERING DATA, 1965, 10 (04): : 396 - +
  • [27] SIMULTANEOUS MEASUREMENTS OF DYNAMIC VALUES USING THE TRANSIT-TIME METHOD
    DAUBARIS, G
    KAUSINIS, S
    RAGAUSKAS, A
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1992, 41 (02) : 251 - 255
  • [28] Precision Dynamic Force Measurements using the Levitation Mass Method (LMM)
    Fujii, Yusaku
    PROCEEDINGS OF 2013 3RD INTERNATIONAL CONFERENCE ON INSTRUMENTATION, COMMUNICATIONS, INFORMATION TECHNOLOGY, AND BIOMEDICAL ENGINEERING (ICICI-BME), 2013, : 3 - 3
  • [29] A novel method for calibration of ADC using JAWS
    Sira, Martin
    Kieler, Oliver
    Behr, Ralf
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [30] ADC compensation using a sinewave histogram method
    Larrabee, JH
    Hummels, DM
    Irons, FH
    IMTC/97 - IEEE INSTRUMENTATION & MEASUREMENT TECHNOLOGY CONFERENCE: SENSING, PROCESSING, NETWORKING, PROCEEDINGS VOLS 1 AND 2, 1997, : 628 - 631