Electrical impedance measurements of cochlear structures using the four-electrode reflection-coefficient technique

被引:9
|
作者
Kumar, Gagan [1 ]
Chokshi, Moulin [1 ]
Richter, Claus-Peter [1 ,2 ,3 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Dept Otolaryngol Head & Neck Surg, Chicago, IL 60611 USA
[2] Northwestern Univ, Dept Biomed Engn, Evanston, IL 60208 USA
[3] Northwestern Univ, Dept Commun Sci & Disorders, Hugh Knowles Ctr, Evanston, IL 60208 USA
基金
美国国家科学基金会;
关键词
Cochlea; Hemicochlea; Electrical impedance; Four-electrode reflection-coefficient method; GUINEA-PIG COCHLEA; INNER-EAR; RESISTANCE CHANGES; HEMICOCHLEA;
D O I
10.1016/j.heares.2009.10.010
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
In individuals with severe-to-profound hearing loss, cochlear implants (CIs) bypass normal inner ear function by applying electrical current directly into the cochlea, thereby stimulating surviving auditory nerve fibers. Although cochlear implants are able to restore some auditory sensation, they are far from providing normal hearing. It has been estimated that up to 75% of the current injected via a Cl is shunted along scala tympani and is not available to stimulate auditory neurons. The path of the injected current and the consequent population of stimulated spiral ganglion cells are dependent upon the positions of the electrode contacts within the cochlea and the impedances of cochlear structures. However, characterization of the current path remains one of the most critical, yet least understood, aspects of cochlear implantation. In particular, the impedances of cochlear structures, including the modiolus, are either unknown or based upon estimates derived from circuit models. Impedance values for many cochlear structures have never been measured. By combining the hemicochlea preparation, a cochlea cut in half along its mid-modiolar plane, and the four-electrode reflection-coefficient technique, impedances can be measured for cochlear tissues in a cochlear cross section including the modiolus. Advantages and disadvantages of the method are discussed in detail and electrical impedance measurements obtained in the gerbil hemicochlea are presented. The resistivity values for the cochlear wall in Omega cm are, 528 (range: 432-708) for scala media 3rd turn, 502 (range: 421-616) for scala tympani 3rd turn and scala vestibuli 2nd turn, 627 (range: 531-759) for scala media 2nd turn, 434 (range: 353-555) for scala tympani 2nd turn and scala vestibuli basal turn, 434 (range: 373-514) for scala media basal turn, and 590 (range: 546-643) for scala tympani basal turn. The resistivity was 455 Omega cm (range: 426-487) for the modiolus. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:86 / 94
页数:9
相关论文
共 40 条
  • [31] A Simple Instrumentation Calibration Technique for Electrical Impedance Tomography (EIT) Using A 16-Electrode Phantom
    Bera, Tushar Kanti
    Nagaraju, J.
    2009 IEEE INTERNATIONAL CONFERENCE ON AUTOMATION SCIENCE AND ENGINEERING, 2009, : 347 - 352
  • [32] Robotic pullback technique of a precurved cochlear-implant electrode array using real-time impedance sensing feedback
    Katherine E. Riojas
    Trevor L. Bruns
    Josephine Granna
    Robert J. Webster
    Robert F. Labadie
    International Journal of Computer Assisted Radiology and Surgery, 2023, 18 : 413 - 421
  • [33] Robotic pullback technique of a precurved cochlear-implant electrode array using real-time impedance sensing feedback
    Riojas, Katherine E.
    Bruns, Trevor L.
    Granna, Josephine
    Webster, Robert J., III
    Labadie, Robert F.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2023, 18 (03) : 413 - 421
  • [34] Four versus two-electrode measurement strategies for cell growing and differentiation monitoring using electrical impedance spectroscopy
    Bragos, R.
    Sarro, E.
    Fontova, A.
    Soley, A.
    Cairo, J.
    Bayes-Genis, A.
    Rosell, J.
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 4544 - +
  • [35] Three and Four-Electrode Electrochemical Impedance Spectroscopy Studies Using Embedded Composite Thin Film Pseudo-Reference Electrodes in Proton Exchange Membrane Fuel Cells
    Szendrei, Alex
    Sparks, Taylor D.
    Virkar, Anil
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2019, 166 (12) : F784 - F795
  • [36] Changes over time in electrical stimulation levels and electrode impedance values in children using the Nucleus 24M cochlear implant
    Henkin, Y
    Kaplan-Neeman, R
    Muchnik, C
    Kronenberg, J
    Hildesheimer, M
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2003, 67 (08) : 873 - 880
  • [38] Modiolar coiling, electrical thresholds, and speech perception after cochlear implantation using the nucleus contour advance electrode with the advance off stylet technique
    Huang, TC
    Reitzen, SD
    Marrinan, MS
    Waltzman, SB
    Roland, JT
    OTOLOGY & NEUROTOLOGY, 2006, 27 (02) : 159 - 166
  • [39] Electrical and dielectric properties of Al/(PVP: Zn-TeO2)/p-Si heterojunction structures using current–voltage (I–V) and impedance-frequency (Z–f) measurements
    Yashar Azizian-Kalandaragh
    Javid Farazin
    Şemsettin Altindal
    Mehdi Shahedi Asl
    Gholamreza Pirgholi-Givi
    Seyed Ali Delbari
    Abbas Sabahi Namini
    Applied Physics A, 2020, 126
  • [40] Electrical and dielectric properties of Al/(PVP: Zn-TeO2)/p-Si heterojunction structures using current-voltage (I-V) and impedance-frequency (Z-f) measurements
    Azizian-Kalandaragh, Yashar
    Farazin, Javid
    Altindal, Semsettin
    Asl, Mehdi Shahedi
    Pirgholi-Givi, Gholamreza
    Delbari, Seyed Ali
    Namini, Abbas Sabahi
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2020, 126 (08):