Evaluation of BD Vacutainer® Barricor™ blood collection tubes for routine chemistry testing on a Roche Cobas® 8000 Platform

被引:12
|
作者
Fournier, Jeffrey E. [1 ]
Northrup, Victoria [2 ,3 ]
Clark, Claudette [3 ]
Fraser, Jacqueline [4 ]
Howlett, Michael [4 ,5 ]
Atkinson, Paul [4 ,5 ]
Shea, Jennifer L. [3 ,6 ]
机构
[1] Dalhousie Med Sch, Dalhousie Med New Brunswick, St John, NB, Canada
[2] St Johns Hosp, Dept Res Serv, St John, NB, Canada
[3] St Johns Hosp, Dept Lab Med, St John, NB, Canada
[4] St Johns Hosp, Dept Emergency Med, St John, NB, Canada
[5] Dalhousie Univ, Dept Emergency Med, Halifax, NS, Canada
[6] Dalhousie Univ, Dept Pathol, Halifax, NS, Canada
关键词
Barricor (TM); PST (TM); Roche Cobas 8000; Routine chemistry testing; LITHIUM HEPARIN; CLINICAL-CHEMISTRY; GEL SEPARATOR; SERUM TUBES; II TUBES; STABILITY; ANALYTES; PLASMA; DRUGS; PST;
D O I
10.1016/j.clinbiochem.2018.06.002
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction: Barricor (TM) Vacutainers (R) are a novel non-gel separator blood collection tube. These tubes enable faster pre-analytical processing which could reduce turnaround time and be beneficial in an acute care setting. We sought to evaluate the bias, stability, and integrity of plasma generated from these tubes compared to Plasma Separator Tubes (TM) (PST) for 50 routine chemistry analyzes on a Roche Cobas (R) 8000 analyzer. Methods: Paired samples were collected from 150 patients originating in the emergency department and outpatient collections at the Saint John Regional Hospital. Barricor (TM) vacutainers were centrifuged for 3 min at 4000g and PST (TM) vacutainers for 10 min at 1300 g within two hours of collection. Plasma samples (n. = 126) were then analyzed for 50 chemistry analyzes and bias determined between tubes. Ten-day stability of AST, glucose, potassium, phosphate, and LDH was also assessed in a subset of paired samples (n. = 4). Lastly, the quality of plasma (n. = 20) was assessed through measurement of cell counts on a DxH Hematology Analyzer. Results: All 50 analyzes demonstrated comparable results across a broad concentration range between Barricor (TM) and PST (TM) vacutainers (average percent bias - 1.5% to 6.1%; Deming linear regression slopes 0.933-1.041; correlation coefficients >= 0.9144). AST, potassium, glucose and LDH were stable for 10 days in Barricor (TM) vacutainers (change from baseline < 10%) but < 5 days in PST (TM) vacutainers while phosphate was stable for 4 days in Barricor (TM) vs 2 days in PST (TM) vacutainers. Platelet counts were statistically lower in Barricor (TM) compared to PST (TM) vacutainers. Conclusion: Our data suggest that Barricor (TM) vacutainers are an acceptable alternative to PST (TM) vacutainers while offering the added benefit of decreased pre-analytical processing time, increased stability of certain analyzes, and possibly less cellular contamination.