Evaluation of BD Vacutainer® Barricor™ blood collection tubes for routine chemistry testing on a Roche Cobas® 8000 Platform

被引:12
|
作者
Fournier, Jeffrey E. [1 ]
Northrup, Victoria [2 ,3 ]
Clark, Claudette [3 ]
Fraser, Jacqueline [4 ]
Howlett, Michael [4 ,5 ]
Atkinson, Paul [4 ,5 ]
Shea, Jennifer L. [3 ,6 ]
机构
[1] Dalhousie Med Sch, Dalhousie Med New Brunswick, St John, NB, Canada
[2] St Johns Hosp, Dept Res Serv, St John, NB, Canada
[3] St Johns Hosp, Dept Lab Med, St John, NB, Canada
[4] St Johns Hosp, Dept Emergency Med, St John, NB, Canada
[5] Dalhousie Univ, Dept Emergency Med, Halifax, NS, Canada
[6] Dalhousie Univ, Dept Pathol, Halifax, NS, Canada
关键词
Barricor (TM); PST (TM); Roche Cobas 8000; Routine chemistry testing; LITHIUM HEPARIN; CLINICAL-CHEMISTRY; GEL SEPARATOR; SERUM TUBES; II TUBES; STABILITY; ANALYTES; PLASMA; DRUGS; PST;
D O I
10.1016/j.clinbiochem.2018.06.002
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction: Barricor (TM) Vacutainers (R) are a novel non-gel separator blood collection tube. These tubes enable faster pre-analytical processing which could reduce turnaround time and be beneficial in an acute care setting. We sought to evaluate the bias, stability, and integrity of plasma generated from these tubes compared to Plasma Separator Tubes (TM) (PST) for 50 routine chemistry analyzes on a Roche Cobas (R) 8000 analyzer. Methods: Paired samples were collected from 150 patients originating in the emergency department and outpatient collections at the Saint John Regional Hospital. Barricor (TM) vacutainers were centrifuged for 3 min at 4000g and PST (TM) vacutainers for 10 min at 1300 g within two hours of collection. Plasma samples (n. = 126) were then analyzed for 50 chemistry analyzes and bias determined between tubes. Ten-day stability of AST, glucose, potassium, phosphate, and LDH was also assessed in a subset of paired samples (n. = 4). Lastly, the quality of plasma (n. = 20) was assessed through measurement of cell counts on a DxH Hematology Analyzer. Results: All 50 analyzes demonstrated comparable results across a broad concentration range between Barricor (TM) and PST (TM) vacutainers (average percent bias - 1.5% to 6.1%; Deming linear regression slopes 0.933-1.041; correlation coefficients >= 0.9144). AST, potassium, glucose and LDH were stable for 10 days in Barricor (TM) vacutainers (change from baseline < 10%) but < 5 days in PST (TM) vacutainers while phosphate was stable for 4 days in Barricor (TM) vs 2 days in PST (TM) vacutainers. Platelet counts were statistically lower in Barricor (TM) compared to PST (TM) vacutainers. Conclusion: Our data suggest that Barricor (TM) vacutainers are an acceptable alternative to PST (TM) vacutainers while offering the added benefit of decreased pre-analytical processing time, increased stability of certain analyzes, and possibly less cellular contamination.
引用
收藏
页码:94 / 99
页数:6
相关论文
共 33 条
  • [21] STAT AND ROUTINE CHEMISTRY ANALYSES UTILIZING THROMBIN CONTAINING BLOOD COLLECTION TUBES
    KIYASU, J
    NAPOLES, D
    GONZALES, MA
    HAUSCHILD, U
    INCE, G
    LIN, FC
    NARAYANAN, S
    CLINICAL CHEMISTRY, 1982, 28 (07) : 1656 - 1656
  • [22] Comparative evaluation of blood collection tubes for clinical chemistry analysis
    Ayala-Lopez, Nadia
    Conklin, Steven E.
    Tenney, Brandon J.
    Ness, Maryann
    Marzinke, Mark A.
    CLINICA CHIMICA ACTA, 2021, 520 : 118 - 125
  • [23] VALIDATION OF LOW-VOLUME BLOOD COLLECTION TUBES FOR ROUTINE HEMATOLOGIC TESTING
    Lippi, Giuseppe
    Martini, Loredana
    Cortivo, Barbara
    Zecchetto, Chiara
    Ferrari, Anna
    JOURNAL OF MEDICAL BIOCHEMISTRY, 2024, 43 (06) : 955 - 959
  • [24] Delayed testing for the diagnosis of fungi in the urines.: Evaluation of the BD Vacutainer® C&S tubes for the storage of urine samples at room temperature
    Baixench, MT
    Al-Sheikh, M
    Paugam, A
    ANNALES DE BIOLOGIE CLINIQUE, 2005, 63 (06) : 661 - 664
  • [25] Performance evaluation of Vacuette® blood collection systems for routine and special coagulation testing.
    Lietz, H
    Hoppensteadt, D
    Reinecke, C
    Bhatia, R
    Iqbal, O
    Fareed, J
    McClatchey, K
    CLINICAL CHEMISTRY, 1998, 44 : A53 - A53
  • [26] Comparison of Total Bilirubin Values Measured with ABL 735 Blood Gas Analyzer and Roche Cobas C8000 Chemistry Analyzer in Age-Segregated Pediatric Patients
    Acar, Esra
    Hunc, Fatih
    Kum, Tugba
    Eraldemir, Fatma Ceyla
    Kir, Hale Maral
    Baydemir, Canan
    REVISTA ROMANA DE MEDICINA DE LABORATOR, 2019, 27 (02): : 137 - 145
  • [27] Evaluation of VACUETTE® cat serum fast separator blood collection tube for routine chemistry assays
    Griebenow, S.
    CLINICA CHIMICA ACTA, 2019, 493 : S263 - S263
  • [28] Evaluation of BD serum separator tubes for potential interference with AEROSET/c8000 colorimetric end-point and rate Clinical Chemistry assays
    Colberg, S
    Craig, P
    Thillen, R
    Combs, B
    CLINICAL CHEMISTRY, 2005, 51 : A114 - A114
  • [29] Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study
    Fajardo, Emmanuel
    Metcalf, Carol
    Mbofana, Elton
    van Vyve, Charlotte
    Munyaradzi, Dhodho
    Simons, Sandra
    Kuhudzayi, Misheck
    Bygrave, Helen
    BMC INFECTIOUS DISEASES, 2014, 14
  • [30] Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study
    Emmanuel Fajardo
    Carol Metcalf
    Elton Mbofana
    Charlotte van Vyve
    Dhodho Munyaradzi
    Sandra Simons
    Misheck Kuhudzayi
    Helen Bygrave
    BMC Infectious Diseases, 14