An Efficient Algorithm Based on Spectrum Migration for High Frame Rate Ultrasound Imaging

被引:2
|
作者
Feng, Shuai [1 ]
Jin, Liu [1 ]
Wang, Yadan [1 ]
Zhao, Wei [1 ,2 ]
Peng, Hu [1 ]
Qiao, Heyuan [1 ]
机构
[1] Hefei Univ Technol, Dept Biomed Engn, Hefei 230009, Peoples R China
[2] Yijishan Hosp, Wannan Med Coll, Wuhu 241001, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
High frame rate imaging; spectrum migration; Fourier migration; DIFFRACTION; SYSTEM; FIELDS;
D O I
10.32604/cmes.2021.014027
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The high frame rate (HFR) imaging technique requires only one emission event for imaging. Therefore, it can achieve ultrafast imaging with frame rates up to the kHz regime, which satisfies the frame rate requirements for imaging moving tissues in scientific research and clinics. Lu's Fourier migration method is based on a non diffraction beam to obtain HFR images and can improve computational speed and efficiency. However, in order to obtain high-quality images, Fourier migration needs to make full use of the spectrum of echo signals for imaging, which requires a large number of Fast Fourier Transform (FFT) points and increases the complexity of the hardware when the echo frequency is high. Here, an efficient algorithm using the spectrum migration technique based on the spectrum's distribution characteristics is proposed to improve the imaging efficiency in HFR imaging. Since the actual echo signal spectrum is of limited bandwidth, low-frequency and high-frequency parts with low-energy have little contribution to the imaging spectrum. We transform the effective part that provides the main energy in the signal spectrum to the imaging spectrum while the ineffective spectrum components are not utilized for imaging. This can significantly reduce the number of Fourier transform points, improve Fourier imaging efficiency, and ensure the imaging quality. The proposed method is evaluated on simulated and experimental datasets. Results demonstrated that the proposed method could achieve equivalent image quality with a reduced point number for FFT compared to the complete spectrum migration. In this paper, it only requires a quarter of the FFT points used in the complete spectrum migration, which can improve the computational efficiency; thus, it is more suitable for real-time data processing. The proposed spectrum migration method has a specific significance for the study and clinical application of HFR imaging.
引用
收藏
页码:739 / 754
页数:16
相关论文
共 50 条
  • [41] High Frame Rate 3-D Ultrasound Imaging Using Separable Beamforming
    Ming Yang
    Richard Sampson
    Siyuan Wei
    Thomas F. Wenisch
    Chaitali Chakrabarti
    Journal of Signal Processing Systems, 2015, 78 : 73 - 84
  • [42] A high-frame rate duplex ultrasound biomicroscopy for small animal imaging in vivo
    Sun, Lei
    Xu, Xiaochen
    Richard, William D.
    Feng, Ching
    Johnson, Jeffrey A.
    Shung, K. Kirk
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (08) : 2039 - 2049
  • [43] Speckle-Enhanced Cardiac Blood Flow Imaging with High Frame Rate Ultrasound
    Takahashi, Hiroki
    Hasegawa, Hideyuki
    Kanai, Hiroshi
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 2030 - 2033
  • [44] Adaptive Sparse Representation Beamformer for High-Frame-Rate Ultrasound Imaging Instrument
    Shen, Minfen
    Zhang, Qiong
    Li, Delai
    Yang, Jinyao
    Li, Bin
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2012, 61 (05) : 1323 - 1333
  • [45] Measuring Submicrometer Displacement Vectors Using High-Frame-Rate Ultrasound Imaging
    Kruizinga, Pieter
    Mastik, Frits
    Bosch, Johan G.
    de Jong, Nico
    van der Steen, Anton F. W.
    van Soest, Gijs
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (10) : 1733 - 1744
  • [46] High Frame Rate 3-D Ultrasound Imaging Using Separable Beamforming
    Yang, Ming
    Sampson, Richard
    Wei, Siyuan
    Wenisch, Thomas F.
    Chakrabarti, Chaitali
    JOURNAL OF SIGNAL PROCESSING SYSTEMS FOR SIGNAL IMAGE AND VIDEO TECHNOLOGY, 2015, 78 (01): : 73 - 84
  • [47] A High-Frequency, High Frame Rate Duplex Ultrasound Linear Array Imaging System for Small Animal Imaging
    Zhang, Lequan
    Xu, Xiaochen
    Hu, Changhong
    Sun, Lei
    Yen, Jesse T.
    Cannata, Jonathan M.
    Shung, K. Kirk
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2010, 57 (07) : 1548 - 1557
  • [48] Improved high-frequency high frame rate duplex ultrasound linear array imaging system
    Zhang, Lequan
    Xu, Xiaochen
    Hu, Changhong
    Sun, Lei
    Yen, Jesse T.
    Cannata, Jonathan M.
    Shung, K. Kirk
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 1730 - +
  • [49] Efficient Frame Rate Control algorithm and its implementation
    Shen, Haibin
    Fan, Junfeng
    FIRST INTERNATIONAL MULTI-SYMPOSIUMS ON COMPUTER AND COMPUTATIONAL SCIENCES (IMSCCS 2006), PROCEEDINGS, VOL 1, 2006, : 799 - +
  • [50] An Efficient Frame-Level Rate Control Algorithm for High Efficiency Video Coding
    Lin, Yubei
    Zhang, Xingming
    Xiao, Jianen
    Su, Shengkai
    KSII TRANSACTIONS ON INTERNET AND INFORMATION SYSTEMS, 2016, 10 (04): : 1877 - 1891