Detection of Unicolor ECG Electrode Reversals in Standard 12-Lead ECG

被引:2
|
作者
Jekova, Irena [1 ]
Leber, Remo [2 ]
Krasteva, Vessela [1 ]
Schmid, Ramun [2 ]
机构
[1] Bulgarian Acad Sci, Inst Biophys & Biomed Engn, Acad G Bonchev Str,Bl 105, BU-1113 Sofia, Bulgaria
[2] Schiller AG, Signal Proc, Baar, Switzerland
关键词
AUTOMATIC DETECTION; LIMB-LEAD; MISPLACEMENT; ELECTROCARDIOGRAM; INTERCHANGE;
D O I
10.22489/CinC.2018.100
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper presents the performance of a commercial lead quality monitoring library (LQMLib, Schiller AG) for detection of reversals between 12-lead ECG cables with matching colors and proposes methods for improvements, where necessary. The study is performed on a large 12-lead ECG database with 1331 chest pain patients (646 training, 685 test recordings) and ECGs from 29 volunteers with swapped red, yellow, green, black and all unicolor cables. Relying on assessment of inter-lead correlations over continuous 4s episodes, LQMLib without adjustments achieves Se>97% for all unicolor cable reversals, except RC1 (Se=29%) and NC5 (Se=88%). The proposed additional analyses improve RC1 and NC5 swap detection by 63% points and 7% points, respectively.
引用
收藏
页数:4
相关论文
共 50 条
  • [21] Prehospital 12-lead ECG - Reply
    Selker, HP
    Zalenski, RJ
    Aufderheide, TP
    ANNALS OF EMERGENCY MEDICINE, 1997, 30 (03) : 353 - 353
  • [22] The 12-lead ECG in peripartum cardiomyopathy
    Tibazarwa, Kemi
    Lee, Geraldine
    Mayosi, Bongani
    Carrington, Melinda
    Stewart, Simon
    Sliwa, Karen
    CARDIOVASCULAR JOURNAL OF AFRICA, 2012, 23 (06) : 322 - 329
  • [23] Derived 12-lead ECG systems
    Nelwan, SP
    Meij, SH
    JOURNAL OF ELECTROCARDIOLOGY, 2006, 39 (01) : 29 - 30
  • [24] COMPARISON OF DIGITAL 12-LEAD ECG AND DIGITAL 12-LEAD HOLTER ECG RECORDINGS IN HEALTHY MALE SUBJECTS
    Taubel, J.
    Naseem, A.
    Shakeri-Nejad, K.
    Dingemanse, J.
    Ferber, G.
    Camm, A. J.
    CLINICAL PHARMACOLOGY & THERAPEUTICS, 2011, 89 : S11 - S11
  • [25] Left Atrial Appendage Activity Translation in the Standard 12-Lead ECG
    Platonov, Pyotr G.
    Nault, Isabelle
    Holmqvist, Fredrik
    Stridh, Martin
    Hocini, Meleze
    Haissaguerre, Michel
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 (06) : 706 - 710
  • [26] Accurate interpretation of the 12-lead ECG electrode placement: A systematic review
    Khunti, Kirti
    HEALTH EDUCATION JOURNAL, 2014, 73 (05) : 610 - 623
  • [27] Study on the use of standard 12-lead ECG data for rhythm-type ECG classification problems
    Park, Junsang
    An, Junho
    Kim, Jinkook
    Jung, Sunghoon
    Gil, Yeongjoon
    Jang, Yoojin
    Lee, Kwanglo
    Oh, Il-young
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2022, 214
  • [28] ENHANCEMENT OF AN OLD DIAGNOSTIC-TOOL, THE STANDARD 12-LEAD ECG
    HINDMAN, NB
    WAGNER, GS
    JOURNAL OF ELECTROCARDIOLOGY, 1987, 20 (02) : 93 - 97
  • [29] Risk stratification by means of standard 12-lead ECG in Brugada syndrome
    Atarashi, Hirotsugu
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (08) : E49 - E49
  • [30] 12-Lead ECG Reconstruction via Combinatoric Inclusion of Fewer Standard ECG Leads with Implications for Lead Information and Significance
    Jain, Utkars
    Butchy, Adam A.
    Leasure, Michael T.
    Covalesky, Veronica A.
    Mccormick, Daniel
    Mintz, Gary S.
    BIOSIGNALS: PROCEEDINGS OF THE 15TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES - VOL 4: BIOSIGNALS, 2022, : 133 - 141