Real-Time Analysis of the Log Files of the HeartWare Continuous-Flow Left Ventricular Assist Device for the Early Diagnosis of Pump Thrombosis: a Step Forward Toward Clinical Translation

被引:5
|
作者
Consolo, Filippo [1 ,2 ]
Pappalardo, Federico [3 ]
机构
[1] Univ Vita Salute San Raffaele, Milan, Italy
[2] Ist Sci San Raffaele, Anesthesia & Intens Care, Milan, Italy
[3] UPMC Italy, IRCCS ISMETT, Anesthesia & Intens Care, Palermo, Italy
关键词
Left ventricular assist device; Pump thrombosis; Log files; Circadian rhythm; Time-frequency analysis; MANAGEMENT;
D O I
10.1007/s12265-021-10157-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the real-time diagnostic capability of a new tool enhancing early diagnosis of pump thrombosis (PT) of the HeartWare left ventricular assist device via time-frequency analysis (TFA) of the log files. We analyzed 173 log files, including 24 (14%) associated with a clinical diagnosis of PT and 149 (86%) controls. The 30-day log file records were discretized into consecutive windows of a 24-h duration, which were iteratively acquired and processed via TFA. This way, we simulated longitudinal acquisition of pump parameters and provided real-time analysis of consecutive data, thus resembling the clinical scenario. Sensitivity and specificity of the tool were 79% and 84%, respectively. Sensitivity against PT events with progressive "build-up" thrombus increased up to 95%, and early signs of a forthcoming PT were identified 10 +/- 8 days prior to its clinical manifestation. This study demonstrates high reliability and the potential for effective clinical translation of this prognostic tool.
引用
收藏
页码:408 / 415
页数:8
相关论文
共 23 条
  • [1] Real-Time Analysis of the Log Files of the HeartWare Continuous-Flow Left Ventricular Assist Device for the Early Diagnosis of Pump Thrombosis: a Step Forward Toward Clinical Translation
    Filippo Consolo
    Federico Pappalardo
    Journal of Cardiovascular Translational Research, 2022, 15 : 408 - 415
  • [2] Log files analysis and evaluation of circadian patterns for the early diagnosis of pump thrombosis with a centrifugal continuous-flow left ventricular assist device
    Consolo, Filippo
    Esposti, Federico
    Gustar, Adrian
    De Bonis, Michele
    Pappalardo, Federico
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (10): : 1077 - 1086
  • [3] Echocardiographic Findings in Continuous-Flow Left Ventricular Assist Device Pump Thrombosis
    Ortega-Legaspi, Juan M.
    Airhart, Sophia
    Hoang Nguyen
    Masri, S. Carolina
    Kirkpatrick, James
    JOURNAL OF CARDIAC FAILURE, 2016, 22 (08) : S117 - S117
  • [4] Acoustic Analysis of a Continuous-Flow Left Ventricular Assist Device before and after Suspected Pump Thrombosis
    Castagna, F.
    Pan, S.
    Garan, A. R.
    Yuzefpolskaya, M.
    Takeda, K.
    Takayama, H.
    Ross, K.
    Torres, M.
    Blum, R. A.
    Singh, S.
    Mancini, D. M.
    Naka, Y.
    Colombo, P. C.
    Bulek, S.
    Topkara, V. K.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (04): : S326 - S326
  • [5] Recurrent pump thrombosis is common after axial continuous-flow left ventricular assist device exchange
    Urban, Marian
    Um, John
    Moulton, Michael
    Stoller, Douglas
    Zolty, Ronald
    Lowes, Brian
    Macdonald, Collin
    McCain, Natalia
    Siddique, Aleem
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (02): : 109 - 118
  • [6] Clinical Durability and Low Incidence of Pump Replacement of a Continuous-Flow Left Ventricular Assist Device
    Moazami, N.
    Ewald, G. A.
    Pagant, F. D.
    John, R.
    MacGillivray, T. E.
    Chen, L.
    Dembitsky, W. P.
    Farrar, D. J.
    Frazier, O. H.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (02): : S207 - S207
  • [7] Log Files of Continuous-Flow Left Ventricular Assist Devices Reveal Diurnal Changes of Pump Parameters Beyond Circadian Variations
    Cervi, Federico
    Vismara, Emanuele
    Pappalardo, Federico
    Consolo, Filippo
    ASAIO JOURNAL, 2021, 67 (02) : E62 - E63
  • [8] A simplified echocardiographic technique for detecting continuous-flow left ventricular assist device malfunction due to pump thrombosis
    Estep, Jerry D.
    Vivo, Rey P.
    Cordero-Reyes, Andrea M.
    Bhimaraj, Arvind
    Trachtenberg, Barry H.
    Torre-Amione, Guillermo
    Chang, Su Min
    Elias, Barbara
    Bruckner, Brian A.
    Suarez, Erik E.
    Loebe, Matthias
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (06): : 575 - 586
  • [9] Clinical Significance of Early Hospital Readmission in Continuous-Flow Left Ventricular Assist Device Patients
    Kutyifa, Valentina
    Naqvi, Syed Y.
    Vidula, Himabindu
    Thomas, Sabu
    Sherazi, Saadia
    McNitt, Scott
    Alexis, Jeffrey D.
    ASAIO JOURNAL, 2020, 66 (07) : 760 - 765
  • [10] Clinical implications of LDH isoenzymes in hemolysis and continuous-flow left ventricular assist device-induced thrombosis
    Gordon, Jonathan S.
    Wood, Chelsey T.
    Luc, Jessica G. Y.
    Watson, Ryan A.
    Maynes, Elizabeth J.
    Choi, Jae Hwan
    Morris, Rohinton J.
    Massey, Howard Todd
    Throckmorton, Amy L.
    Tchantchaleishvili, Vakhtang
    ARTIFICIAL ORGANS, 2020, 44 (03) : 231 - 238