Real-Time Analysis of the Log Files of the HeartWare Continuous-Flow Left Ventricular Assist Device for the Early Diagnosis of Pump Thrombosis: a Step Forward Toward Clinical Translation

被引:5
|
作者
Consolo, Filippo [1 ,2 ]
Pappalardo, Federico [3 ]
机构
[1] Univ Vita Salute San Raffaele, Milan, Italy
[2] Ist Sci San Raffaele, Anesthesia & Intens Care, Milan, Italy
[3] UPMC Italy, IRCCS ISMETT, Anesthesia & Intens Care, Palermo, Italy
关键词
Left ventricular assist device; Pump thrombosis; Log files; Circadian rhythm; Time-frequency analysis; MANAGEMENT;
D O I
10.1007/s12265-021-10157-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the real-time diagnostic capability of a new tool enhancing early diagnosis of pump thrombosis (PT) of the HeartWare left ventricular assist device via time-frequency analysis (TFA) of the log files. We analyzed 173 log files, including 24 (14%) associated with a clinical diagnosis of PT and 149 (86%) controls. The 30-day log file records were discretized into consecutive windows of a 24-h duration, which were iteratively acquired and processed via TFA. This way, we simulated longitudinal acquisition of pump parameters and provided real-time analysis of consecutive data, thus resembling the clinical scenario. Sensitivity and specificity of the tool were 79% and 84%, respectively. Sensitivity against PT events with progressive "build-up" thrombus increased up to 95%, and early signs of a forthcoming PT were identified 10 +/- 8 days prior to its clinical manifestation. This study demonstrates high reliability and the potential for effective clinical translation of this prognostic tool.
引用
收藏
页码:408 / 415
页数:8
相关论文
共 23 条
  • [11] Increased Platelet Glycoprotein IIb/IIIa Activation Precedes Continuous-Flow Left Ventricular Assist Device Pump Thrombosis Events
    Weiland, M.
    Schuitema, J.
    Jovinge, S.
    Lee, S.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2020, 39 (04): : S148 - S148
  • [12] Increased platelet glycoprotein IIb/IIIa activation precedes continuous-flow left ventricular assist device pump thrombosis events
    Weiland, Matthew
    Schuitema, Jennifer
    Jovinge, Stefan
    Lee, Sangjin
    THROMBOSIS RESEARCH, 2021, 201 : 143 - 146
  • [13] Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device HeartMate II
    S. Akin
    O.I. Soliman
    A.A. Constantinescu
    F. Akca
    O. Birim
    R.T. van Domburg
    O. Manintveld
    K. Caliskan
    Netherlands Heart Journal, 2016, 24 : 134 - 142
  • [14] Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device HeartMate II
    Akin, S.
    Soliman, O. I.
    Constantinescu, A. A.
    Akca, F.
    Birim, O.
    van Domburg, R. T.
    Manintveld, O.
    Caliskan, K.
    NETHERLANDS HEART JOURNAL, 2016, 24 (02) : 134 - 142
  • [15] Surgical device exchange provides improved clinical outcomes compared to medical therapy in treating continuous-flow left ventricular assist device thrombosis
    Koda, Yojiro
    Kitahara, Hiroto
    Kalantari, Sara
    Chung, Bow
    Smith, Bryan
    Raikhelkar, Jayant
    Kim, Gene
    Sarswat, Nitasha
    Sayer, Gabriel
    Onsager, David
    Song, Tae
    Uriel, Nir
    Jeevanandam, Valluvan
    Ota, Takeyoshi
    ARTIFICIAL ORGANS, 2020, 44 (04) : 367 - 374
  • [16] Characteristics, Therapeutic Efficacy, and Clinical Implications of Ventricular Tachycardias Detected Early After Implant of Continuous-flow Left Ventricular Assist Device
    Chyou, Janice Y.
    Whang, William
    Dizon, Jose
    Biviano, Angelo
    Garan, Hasan
    Naka, Yoshifumi
    Colombo, Paolo
    Garan, Arthur R.
    CIRCULATION, 2016, 134
  • [17] Real-Time 3-Dimensional Echocardiographic Assessment of Current Continuous-Flow Rotary Left Ventricular Assist Devices
    Castillo, Javier G.
    Anyanwu, Anelechi C.
    Adams, David H.
    Nyirenda, Thandiwe
    Fischer, Gregory W.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2009, 23 (05) : 702 - 710
  • [18] Meta-Analysis of Time in Therapeutic Range in Continuous-Flow Left Ventricular Assist Device Patients Receiving Warfarin
    Martinez, Brandon K.
    Yik, Brenda
    Tran, Raymond
    Ilham, Sabrina
    Coleman, Craig I.
    Jennings, Douglas L.
    Baker, William L.
    ARTIFICIAL ORGANS, 2018, 42 (07) : 700 - 704
  • [19] Increased Aortic Stiffness is Associated with Higher Rates of Stroke, GI-bleeding and Pump Thrombosis in Continuous-Flow Left Ventricular Assist Device Patients
    Rosenblum, H.
    Pinsino, A.
    Demmer, R. T.
    Colombo, P. C.
    Yuzefpolskaya, M.
    Garan, A. R.
    Topkara, V. K.
    Gaudig, A.
    Pineda, M. T.
    Naka, Y.
    Takayama, H.
    Takeda, K.
    Castagna, F.
    Willey, J. Z.
    McDonnell, B. J.
    Cockcroft, J. R.
    Stohr, E. J.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2018, 37 (04): : S32 - S32
  • [20] Medical Management of Pump-Related Thrombosis in Patients with Continuous-Flow Left Ventricular Assist Devices: A Systematic Review and Meta-Analysis
    Dang, Geetanjali
    Epperla, Narendranath
    Muppidi, Vijayadershan
    Sahr, Natasha
    Pan, Amy
    Simpson, Pippa
    Kreuziger, Lisa Baumann
    ASAIO JOURNAL, 2017, 63 (04) : 373 - 385