MR-Guided Vertebroplasty With Augmented Reality Image Overlay Navigation

被引:24
|
作者
Fritz, Jan [1 ]
U-Thainual, Paweena [2 ,3 ,4 ,5 ]
Ungi, Tamas [4 ]
Flammang, Aaron J. [6 ]
Kathuria, Sudhir [1 ]
Fichtinger, Gabor [4 ]
Iordachita, Iulian I. [2 ,3 ]
Carrino, John A. [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21287 USA
[2] Johns Hopkins Univ, Dept Mech Engn, Baltimore, MD 21218 USA
[3] Johns Hopkins Univ, Lab Computat Sensing & Robot, Baltimore, MD 21218 USA
[4] Queens Univ, Sch Comp, Kingston, ON K7L 2N8, Canada
[5] Queens Univ, Dept Mech & Mat Engn, Kingston, ON K7L 3N6, Canada
[6] Siemens Healthcare USA Inc, Malvern, PA 19355 USA
关键词
Vertebroplasty; Interventional MR imaging; MR imaging guidance; MR-guided; SPINAL INJECTION PROCEDURES; PERCUTANEOUS VERTEBROPLASTY; BONE-BIOPSY; INTERVENTIONS; ABLATION; VISUALIZATION; FEASIBILITY; ACCURACY; PERFORMANCE; TECHNOLOGY;
D O I
10.1007/s00270-014-0885-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To evaluate the feasibility of magnetic resonance imaging (MRI)-guided vertebroplasty at 1.5 Tesla using augmented reality image overlay navigation. Twenty-five unilateral vertebroplasties [5 of 25 (20 %) thoracic, 20 of 25 (80 %) lumbar] were prospectively planned in 5 human cadavers. A clinical 1.5-Teslan MRI system was used. An augmented reality image overlay navigation system and 3D Slicer visualization software were used for MRI display, planning, and needle navigation. Intermittent MRI was used to monitor placement of the MRI-compatible vertebroplasty needle. Cement injections (3 ml of polymethylmethacrylate) were performed outside the bore. The cement deposits were assessed on intermediate-weighted MR images. Outcome variables included type of vertebral body access, number of required intermittent MRI control steps, location of final needle tip position, cement deposit location, and vertebroplasty time. All planned procedures (25 of 25, 100 %) were performed. Sixteen of 25 (64 %) transpedicular and 9 of 25 (36 %) parapedicular access routes were used. Six (range 3-9) MRI control steps were required for needle placement. No inadvertent punctures were visualized. Final needle tip position and cement location were adequate in all cases (25 of 25, 100 %) with a target error of the final needle tip position of 6.1 +/- A 1.9 mm (range 0.3-8.7 mm) and a distance between the planned needle tip position and the center of the cement deposit of 4.3 mm (range 0.8-6.8 mm). Time requirement for one level was 16 (range 11-21) min. MRI-guided vertebroplasty using image overlay navigation is feasible allowing for accurate vertebral body access and cement deposition in cadaveric thoracic and lumbar vertebral bodies.
引用
收藏
页码:1589 / 1596
页数:8
相关论文
共 50 条
  • [21] Augmented Reality Navigation With Automatic Marker-Free Image Registration Using 3-D Image Overlay for Dental Surgery
    Wang, Junchen
    Suenaga, Hideyuki
    Hoshi, Kazuto
    Yang, Liangjing
    Kobayashi, Etsuko
    Sakuma, Ichiro
    Liao, Hongen
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (04) : 1295 - 1304
  • [22] Augmented Reality Visualization with Use of Image Overlay Technology for MR Imaging-guided Interventions: Assessment of Performance in Cadaveric Shoulder and Hip Arthrography at 1.5 T
    Fritz, Jan
    U-Thainual, Paweena
    Ungi, Tamas
    Flammang, Aaron J.
    Fichtinger, Gabor
    Iordachita, Iulian I.
    Carrino, John A.
    RADIOLOGY, 2012, 265 (01) : 254 - 259
  • [23] Moving-Tolerant Augmented Reality Surgical Navigation System Using Autostereoscopic Three-Dimensional Image Overlay
    Ma, Cong
    Chen, Guowen
    Zhang, Xinran
    Ning, Guochen
    Liao, Hongen
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2019, 23 (06) : 2483 - 2493
  • [24] Application of open source image guided therapy software in MR-guided therapies
    Hata, Nobuhiko
    Piper, Steve
    Jolesz, Ferenc A.
    Tempany, Clare M. C.
    Black, Peter McL
    Morikawa, Shigehiro
    Iseki, Horoshi
    Hashizume, Makoto
    Kikinis, Ron
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2007, PT 1, PROCEEDINGS, 2007, 4791 : 491 - +
  • [25] Image fusion for catheter tracking in MR-guided endovascular therapy
    Draper, J. N.
    Sabati, M.
    Lauzon, M. L.
    Frayne, R.
    MEDICAL PHYSICS, 2006, 33 (07) : 2669 - 2669
  • [26] Multi-parametric MR-guided PET Image Reconstruction
    Mehranian, Abolfazl
    Reader, Andrew J.
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [27] MR-guided MR arthrography of the shoulder
    Trattnig, S
    Breitenseher, M
    Pretterklieber, M
    Kontaxis, G
    Rand, T
    Imhof, H
    RADIOLOGE, 1996, 36 (09): : 709 - 712
  • [28] An augmented reality system for image-guided surgery
    Marmulla, R
    Hoppe, H
    Mühling, J
    Eggers, G
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 34 (06) : 594 - 596
  • [29] 3-D Augmented Reality for MRI-Guided Surgery Using Integral Videography Autostereoscopic Image Overlay
    Liao, Hongen
    Inomata, Takashi
    Sakuma, Ichiro
    Dohi, Takeyoshi
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (06) : 1476 - 1486
  • [30] An augmented reality system for MR image-guided needle biopsy: Initial results in a swine model
    Wacker, FK
    Vogt, S
    Khamene, A
    Jesberger, JA
    Nour, SG
    Elgort, DR
    Sauer, F
    Duerk, JL
    Lewin, JS
    RADIOLOGY, 2006, 238 (02) : 497 - 504