Implementation of free breathing respiratory amplitude-gated treatments

被引:6
|
作者
Hickling, Susannah V. [1 ]
Veres, Andrew J. [1 ]
Moseley, Douglas J. [1 ]
Grams, Michael P. [1 ]
机构
[1] Mayo Clin, Dept Radiat Oncol, Rochester, MN 55905 USA
来源
关键词
4-DIMENSIONAL COMPUTED-TOMOGRAPHY; INDUCED TUMOR MOTION; QUALITY-ASSURANCE; LUNG; RADIOTHERAPY; 4DCT;
D O I
10.1002/acm2.13253
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purpose of this study was to provide guidance in developing and implementing a process for the accurate delivery of free breathing respiratory amplitude-gated treatments. Methods A phase-based 4DCT scan is acquired at time of simulation and motion is evaluated to determine the exhale phases that minimize respiratory motion to an acceptable level. A phase subset average CT is then generated for treatment planning and a tracking structure is contoured to indicate the location of the target or a suitable surrogate over the planning phases. Prior to treatment delivery, a 4DCBCT is acquired and a phase subset average is created to coincide with the planning phases for an initial match to the planning CT. Fluoroscopic imaging is then used to set amplitude gate thresholds corresponding to when the target or surrogate is in the tracking structure. The final imaging prior to treatment is an amplitude-gated CBCT to verify both the amplitude gate thresholds and patient positioning. An amplitude-gated treatment is then delivered. This technique was commissioned using an in-house lung motion phantom and film measurements of a simple two-field 3D plan. Results The accuracy of 4DCBCT motion and target position measurements were validated relative to 4DCT imaging. End to end testing showed strong agreement between planned and film measured dose distributions. Robustness to interuser variability and changes in respiratory motion were demonstrated through film measurements. Conclusions The developed workflow utilizes 4DCBCT, respiratory-correlated fluoroscopy, and gated CBCT imaging in an efficient and sequential process to ensure the accurate delivery of free breathing respiratory-gated treatments.
引用
收藏
页码:119 / 129
页数:11
相关论文
共 50 条
  • [31] A graphical and quantitative procedure for amplitude-gated treatment planned using phase-based 4DCT using the varian RPM system
    Hubenschmidt, J.
    Santanam, L.
    Goddu, S.
    Parikh, P.
    Mutic, S.
    Low, D.
    MEDICAL PHYSICS, 2006, 33 (06) : 2004 - 2004
  • [32] Respiratory Self-Gated Multiple Gradient Recalled Echo Sequence for Free-Breathing Abdominal R*2 Mapping
    Jin, Ning
    Zhang, Zhuoli
    Zhang, Longjiang
    Lu, Guangming
    Larson, Andrew C.
    MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (01) : 207 - 212
  • [33] CLINICAL COMPARISON OF THE FREE BREATHING INSPIRATION GATED AND DEEP INSPIRATION VOLUNTARY BREATH HOLD RESPIRATORY-GATED RADIOTHERAPY TECHNIQUES FOR EARLY STAGE BREAST CANCER PATIENTS
    Utehina, O.
    Popov, S.
    Purina, D.
    Siosberga, I.
    Vevere, I.
    Emzins, D.
    Berzins, J.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S195 - S196
  • [34] An affordable custom phantom for measurement of linac time delay in gated treatments with irregular breathing
    Santos, Alexandre M. Caraca
    Shepherd, Justin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (03) : 863 - 869
  • [35] An affordable custom phantom for measurement of linac time delay in gated treatments with irregular breathing
    Alexandre M. Caraça Santos
    Justin Shepherd
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 863 - 869
  • [36] Respiratory motion gated PET with end-expiratory, amplitude based gating
    Frey, Kirk
    Hamill, James
    Jones, Judson
    Koeppe, Robert
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [37] DEPENDENCIES OF RESPIRATORY SYSTEM RESISTANCE AND ELASTANCE ON AMPLITUDE AND FREQUENCY IN THE NORMAL RANGE OF BREATHING
    BARNAS, GM
    MILLS, PJ
    MACKENZIE, CF
    ASHBY, M
    SEXTON, WL
    IMLE, PC
    WILSON, PD
    AMERICAN REVIEW OF RESPIRATORY DISEASE, 1991, 143 (02): : 240 - 244
  • [38] IMPORTANCE OF RESPIRATORY PATHWAYS CROSSING THE MIDLINE AT CERVICAL LEVEL FOR AMPLITUDE AND RHYTHM OF BREATHING
    JANCZEWSKI, WA
    BULLETIN EUROPEEN DE PHYSIOPATHOLOGIE RESPIRATOIRE-CLINICAL RESPIRATORY PHYSIOLOGY, 1986, 22 : S19 - S19
  • [39] Free-Breathing Gated Proton Pencil Beam Scanning (PBS) Delivery - Proof of Concept Phantom Study with Respiratory Monitory System
    Qi, H.
    Chen, Q.
    Yu, F.
    Han, D.
    Kang, M.
    Zhou, J.
    Lazarev, S.
    Lee, N. Y.
    Garg, M. K.
    Hajj, C.
    Choi, I. J.
    Hasan, S.
    Chhabra, A. M.
    Simone, C. B., II
    Lin, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (02): : E178 - E178
  • [40] Comparing phase- and amplitude-gated volumetric modulated arc therapy for stereotactic body radiation therapy using 3D printed lung phantom
    Lee, Minsik
    Yoon, KyoungJun
    Cho, Byungchul
    Kim, Su Ssan
    Song, Si Yeol
    Choi, Eun Kyung
    Ahn, SeungDo
    Lee, Sang-Wook
    Kwak, JungWon
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (02): : 107 - 113