Fatigue lifetime prediction of a reduced-diameter dental implant system: Numerical and experimental study

被引:43
|
作者
Duan, Yuanyuan [1 ]
Gonzalez, Jorge A. [2 ]
Kulkarni, Pratim A. [1 ]
Nagy, William W. [3 ]
Griggs, Jason A. [1 ]
机构
[1] Univ Missouri, Med Ctr, Dept Biomed Mat Sci, 2500 N State St, Jackson, MS 39216 USA
[2] Texas A&M Hlth Sci Ctr, Dept Oral & Maxillofacial Surg, Bryan, TX USA
[3] Texas A&M Hlth Sci Ctr, Dept Restorat Sci, Bryan, TX USA
关键词
Cyclic fatigue; Accelerated lifetime testing; Finite element modeling; Dental implants; Fractography; Titanium; Simulated bone; Plastic strain; Brown-Miller criteria; FINITE-ELEMENT-ANALYSIS; MECHANICAL-PROPERTIES; MINI-IMPLANTS; CO-CR; COMPLICATIONS; TI-6AL-7NB; RESISTANCE; PROSTHESES; CASTINGS; MAXILLA;
D O I
10.1016/j.dental.2018.06.002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. To validate the fatigue lifetime of a reduced-diameter dental implant system predicted by three-dimensional finite element analysis (FEA) by testing physical implant specimens using an accelerated lifetime testing (ALT) strategy with the apparatus specified by ISO 14801. Methods. A commercially-available reduced-diameter titanium dental implant system (Straumann Standard Plus NN) was digitized using a micro-CT scanner. Axial slices were processed using an interactive medical image processing software (Mimics) to create 3D models. FEA analysis was performed in ABAQUS, and fatigue lifetime was predicted using fe-safe (R) software. The same implant specimens (n = 15) were tested at a frequency of 2 Hz on load frames using apparatus specified by ISO 14801 and ALT. Multiple step-stress load profiles with various aggressiveness were used to improve testing efficiency. Fatigue lifetime statistics of physical specimens were estimated in a reliability analysis software (ALTA PRO). Fractured specimens were examined using SEM with fractographic technique to determine the failure mode. Results. FEA predicted lifetime was within the 95% confidence interval of lifetime estimated by experimental results, which suggested that FEA prediction was accurate for this implant system. The highest probability of failure was located at the root of the implant body screw thread adjacent to the simulated bone level, which also agreed with the failure origin in physical specimens. Significance. Fatigue lifetime predictions based on finite element modeling could yield similar results in lieu of physical testing, allowing the use of virtual testing in the early stages of future research projects on implant fatigue. (C) 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1299 / 1309
页数:11
相关论文
共 50 条
  • [41] Numerical verification of two-component dental implant in the context of fatigue life for various load cases
    Szajek, Krzysztof
    Wierszycki, Marcin
    ACTA OF BIOENGINEERING AND BIOMECHANICS, 2016, 18 (01) : 103 - 113
  • [42] Deformation behaviour and numerical fatigue lifetime prediction of metallic materials under multiaxial nonproportional loading
    Döring, R
    Hoffmeyer, J
    Vormwald, M
    Seeger, T
    MATERIALWISSENSCHAFT UND WERKSTOFFTECHNIK, 2002, 33 (05) : 280 - 288
  • [43] Numerical and experimental study of the fatigue of threaded pipe couplings
    Van Wittenberghe, J.
    De Baets, P.
    De Waele, W.
    Van Autreve, S.
    SURFACE EFFECTS AND CONTACT MECHANICS IX, 2009, : 163 - 174
  • [44] Experimental Study on Displacement and Strain Distributions of Bone Model with Dental Implant
    Morita, Yasuyuki
    Matsushita, Yasuyuki
    Todo, Mitsugu
    Koyano, Kiyoshi
    EXPERIMENTAL MECHANICS AND MATERIALS, 2011, 83 : 73 - +
  • [45] The barrier between the keratinized mucosa and the dental implant - An experimental study in the dog
    Moon, IS
    Berglundh, T
    Abrahamsson, I
    Linder, E
    Lindhe, J
    JOURNAL OF CLINICAL PERIODONTOLOGY, 1999, 26 (10) : 658 - 663
  • [46] Load fatigue performance of a single-tooth implant-abutment system: Effect of diameter.
    Quek, CEY
    Tan, KBC
    Nicholls, JI
    JOURNAL OF DENTAL RESEARCH, 2003, 82 : 663 - 663
  • [47] Biomechanical effects of dental implant diameter, connection type, and bone density on microgap formation and fatigue failure: A finite element analysis
    Lee, Hyeonjong
    Jo, Minhye
    Noh, Gunwoo
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 200
  • [48] Contact analysis of gap formation at dental implant-abutment interface under oblique loading: A numerical-experimental study
    He, Yiting
    Fok, Alex
    Aparicio, Conrado
    Teng, Wei
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (04) : 741 - 752
  • [49] Numerical and experimental method for the prediction of the propagation life of fatigue crack on metallic materials
    Ni, Chen
    Hua, Lin
    Wang, Xiaokai
    Wang, Zhou
    Ma, Zhikui
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2018, 32 (09) : 4183 - 4190
  • [50] Numerical and experimental method for the prediction of the propagation life of fatigue crack on metallic materials
    Chen Ni
    Lin Hua
    Xiaokai Wang
    Zhou Wang
    Zhikui Ma
    Journal of Mechanical Science and Technology, 2018, 32 : 4183 - 4190