Fatigue lifetime prediction of a reduced-diameter dental implant system: Numerical and experimental study

被引:43
|
作者
Duan, Yuanyuan [1 ]
Gonzalez, Jorge A. [2 ]
Kulkarni, Pratim A. [1 ]
Nagy, William W. [3 ]
Griggs, Jason A. [1 ]
机构
[1] Univ Missouri, Med Ctr, Dept Biomed Mat Sci, 2500 N State St, Jackson, MS 39216 USA
[2] Texas A&M Hlth Sci Ctr, Dept Oral & Maxillofacial Surg, Bryan, TX USA
[3] Texas A&M Hlth Sci Ctr, Dept Restorat Sci, Bryan, TX USA
关键词
Cyclic fatigue; Accelerated lifetime testing; Finite element modeling; Dental implants; Fractography; Titanium; Simulated bone; Plastic strain; Brown-Miller criteria; FINITE-ELEMENT-ANALYSIS; MECHANICAL-PROPERTIES; MINI-IMPLANTS; CO-CR; COMPLICATIONS; TI-6AL-7NB; RESISTANCE; PROSTHESES; CASTINGS; MAXILLA;
D O I
10.1016/j.dental.2018.06.002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. To validate the fatigue lifetime of a reduced-diameter dental implant system predicted by three-dimensional finite element analysis (FEA) by testing physical implant specimens using an accelerated lifetime testing (ALT) strategy with the apparatus specified by ISO 14801. Methods. A commercially-available reduced-diameter titanium dental implant system (Straumann Standard Plus NN) was digitized using a micro-CT scanner. Axial slices were processed using an interactive medical image processing software (Mimics) to create 3D models. FEA analysis was performed in ABAQUS, and fatigue lifetime was predicted using fe-safe (R) software. The same implant specimens (n = 15) were tested at a frequency of 2 Hz on load frames using apparatus specified by ISO 14801 and ALT. Multiple step-stress load profiles with various aggressiveness were used to improve testing efficiency. Fatigue lifetime statistics of physical specimens were estimated in a reliability analysis software (ALTA PRO). Fractured specimens were examined using SEM with fractographic technique to determine the failure mode. Results. FEA predicted lifetime was within the 95% confidence interval of lifetime estimated by experimental results, which suggested that FEA prediction was accurate for this implant system. The highest probability of failure was located at the root of the implant body screw thread adjacent to the simulated bone level, which also agreed with the failure origin in physical specimens. Significance. Fatigue lifetime predictions based on finite element modeling could yield similar results in lieu of physical testing, allowing the use of virtual testing in the early stages of future research projects on implant fatigue. (C) 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1299 / 1309
页数:11
相关论文
共 50 条
  • [21] Finite Element Analysis and Fatigue Test of INTEGRA Dental Implant System
    Zielinski, Rafal
    Lipa, Sebastian
    Piechaczek, Martyna
    Sowinski, Jerzy
    Kolkowska, Agata
    Simka, Wojciech
    MATERIALS, 2024, 17 (05)
  • [22] Experimental study to establishing the insertion torque on dental implant
    Isabel Lopez-Gualdron, Clara
    Fernando Galeano-Arrieta, Carlos
    Cesar Pinillos, Julio
    Carlos Moreno, Juan
    REVISTA ITECKNE, 2014, 11 (01): : 17 - 26
  • [23] Experimental study of plasma nitriding dental implant surfaces
    Guerra Neto, C. L. B.
    da Silva, M. A. M.
    Alves, C., Jr.
    SURFACE ENGINEERING, 2009, 25 (06) : 430 - 433
  • [24] Experimental Study and Numerical Analysis for Prediction of the Fibre Diameter of Polylactic Acid (PLA) Spunbonded Nonwovens
    Bo, Zhao
    FIBRES & TEXTILES IN EASTERN EUROPE, 2009, 17 (05) : 82 - 86
  • [25] Influence of Narrow Titanium Dental Implant Diameter on Fatigue Behavior: A Comparison between Unitary and Splinted Implants
    Terrats, Rodrigo Gonzalez
    Linares, Maria Bonnin
    Punset, Miquel
    Molmeneu, Meritxell
    Nart Molina, Jose
    Ruiz Magaz, Vanessa
    Albertini, Matteo
    Manero, Jose Maria
    Gil Mur, Javier
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (06)
  • [26] Lifetime fatigue cracking behavior of weld defects in orthotropic steel bridge decks: Numerical and experimental study
    Luo, Yuan
    Liu, Xiaofan
    Cui, Jian
    Wang, Honghao
    Huang, Lian
    Zeng, Weiming
    Zhang, Haiping
    ENGINEERING FAILURE ANALYSIS, 2025, 167
  • [27] Impact of Implant Diameter on Success and Survival of Dental Implants: An Observational Cohort Study
    Romanos, Georgios E.
    Schesni, Aigerim
    Nentwig, Georg-Hubertus
    Winter, Anna
    Sader, Robert
    Brandt, Silvia
    PROSTHESIS, 2023, 5 (03): : 888 - 897
  • [28] Fatigue life prediction of commercial dental implant using analytical approach and verification by FEA
    Dhatrak P.
    Shirsat U.
    Deshmukh V.
    Sumanth S.
    Dhatrak, Pankaj (pankaj.dhatrak@mitpune.edu.in), 2018, Springer Heidelberg (00): : 203 - 212
  • [29] On the fatigue life of dental implants: Numerical and experimental investigation on configuration effect
    Abdoli, Zahrae
    Mohammadi, Bijan
    Karimi, Hamid Reza
    MEDICAL ENGINEERING & PHYSICS, 2024, 123
  • [30] Numerical analysis of a dental implant system pre-tensioned with a washer
    Versluis, A
    Cardoso, AC
    Korioth, TWP
    JOURNAL OF DENTAL RESEARCH, 1998, 77 : 836 - 836