Portable gas chromatograph calibration with gases of varying viscosities

被引:1
|
作者
Tschedanoff, Vera [1 ]
Schildhauer, Tilman J. [1 ]
Biollaz, Serge M. A. [1 ]
Wokaun, Alexander [1 ]
机构
[1] Paul Scherrer Inst, Gen Energy Res Dept, CH-5232 Villigen, Switzerland
关键词
NATURAL-GAS; FLUIDIZED-BED;
D O I
10.1016/j.talanta.2020.121077
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The quantitative analysis of gas mixtures with gas chromatography is based on the calibration with certified standards and the determination of a response relationship for each species by regression analysis. The conventional assumption of a constant amount-of-substance injected onto the column, the sample size, for all standards analyzed represents one of the largest sources of uncertainies in this analysis technique. For systems using time-based microinjectors, the sample size injected onto the column is determined by the opening time of a pneumatically actuated microvalve and therefore depends upon the gas velocity developed in the injector's microchannel. For this reason the sample size is not necessarily constant for a given opening time, but strongly correlates - according to fluid mechanic theory - with the gas mixture's dynamic viscosity. Neglecting these sample size variations leads to errors in the analysis up to 30%, depending on the diversity of the standards' viscosities, especially in processes with strongly changing hydrogen contents. A mathematical correction exploiting the inverse proportionality of the sample size and the sample's dynamic viscosity normalizes the injection amounts and minimizes the influence of the sample variations on the calibration regression. The data analysis based on the corrected normalized calibration is no longer viscosity dependent and can be applied to gas mixtures of any composition.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Calibration of a portable tritium-in-air monitor for various radioactive gases
    Wood, MJ
    Hong, A
    Cross, WG
    Nunes, JC
    Leon, JW
    HEALTH PHYSICS, 1997, 72 (03): : 423 - 430
  • [42] RAPID ANALYSIS OF HYDROCARBONS AND INERT-GASES BY A MULTIDIMENSIONAL GAS-CHROMATOGRAPH
    TANI, H
    FURUNO, M
    JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1986, 9 (11): : 712 - 716
  • [43] GAS CHROMATOGRAPH
    DONNER, W
    JOURNAL OF GAS CHROMATOGRAPHY, 1965, 3 (02): : A14 - &
  • [44] INSTRUMENT CALIBRATION FOR MASS CHROMATOGRAPH
    LLOYD, RJ
    HENDERSON, DE
    UDEN, PC
    ANALYTICAL CHEMISTRY, 1976, 48 (11) : 1645 - 1646
  • [45] CALIBRATION OF GEL PERMEATION CHROMATOGRAPH
    PURDON, JR
    MATE, RD
    JOURNAL OF POLYMER SCIENCE PART A-1-POLYMER CHEMISTRY, 1968, 6 (1PA1): : 243 - &
  • [46] GAS CHROMATOGRAPH
    NORLEY, WN
    NEW ENGLAND JOURNAL OF MEDICINE, 1964, 271 (18): : 963 - &
  • [47] Gas chromatograph
    不详
    MICRO, 1995, 13 (02): : 54 - 54
  • [48] PORTABLE GAS CHROMATOGRAPH GIVES QUICK FIELD TEST FOR CHLORINATED HYDROCARBONS IN CRUDE.
    Bryant, M.R.
    Krottinger, D.L.
    Oil and Gas Journal, 1983, 81 (25): : 105 - 108
  • [49] A modular platform, portable gas chromatograph based on miniaturized chip injection and detection technologies
    de Zeeuw, J
    van Hove, R
    Curvers, J
    Verhoeve, L
    AMERICAN LABORATORY, 2002, 34 (25) : 25 - 27
  • [50] A method for on-site analysis of urinary benzene by means of a portable gas-chromatograph
    Suna, S
    Jitsunari, F
    Asakawa, F
    Hirao, T
    Mannami, T
    Suzue, T
    JOURNAL OF OCCUPATIONAL HEALTH, 2005, 47 (01) : 74 - 77