Potential of Stroke Imaging Using a New Prototype of Low-Field MRI: A Prospective Direct 0.55 T/1.5 T Scanner Comparison

被引:23
|
作者
Rusche, Thilo [1 ]
Breit, Hanns-Christian [1 ]
Bach, Michael [1 ]
Wasserthal, Jakob [1 ]
Gehweiler, Julian [1 ]
Manneck, Sebastian [1 ]
Lieb, Johanna Maria [1 ]
De Marchis, Gian Marco [2 ]
Psychogios, Marios Nikos [1 ]
Sporns, Peter B. [1 ,3 ]
机构
[1] Univ Basel, Univ Hosp Basel, Dept Radiol, Clin Radiol & Nucl Med, CH-4031 Basel, Switzerland
[2] Univ Basel, Univ Hosp Basel, Dept Neurol, CH-4031 Basel, Switzerland
[3] Univ Med Ctr Hamburg Eppendorf, Dept Diagnost & Intervent Neuroradiol, D-20246 Hamburg, Germany
关键词
stroke imaging; low-field MRI; reading study; scanner comparison; GLOBAL BURDEN;
D O I
10.3390/jcm11102798
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: Ischemic stroke is a leading cause of mortality and acquired disability worldwide and thus plays an enormous health-economic role. Imaging of choice is computed-tomographic (CT) or magnetic resonance imaging (MRI), especially diffusion-weighted (DW) sequences. However, MR imaging is associated with high costs and therefore has a limited availability leading to low-field-MRI techniques increasingly coming into focus. Thus, the aim of our study was to assess the potential of stroke imaging with low-field MRI. Material and Methods: A scanner comparison was performed including 27 patients (17 stroke cohort, 10 control group). For each patient, a brain scan was performed first with a 1.5T scanner and afterwards with a 0.55T scanner. Scan protocols were as identical as possible and optimized. Data analysis was performed in three steps: All DWI/ADC (apparent diffusion coefficient) and FLAIR (fluid attenuated inversion recovery) sequences underwent Likert rating with respect to image impression, resolution, noise, contrast, and diagnostic quality and were evaluated by two radiologists regarding number and localization of DWI and FLAIR lesions in a blinded fashion. Then segmentation of lesion volumes was performed by two other radiologists on DWI/ADC and FLAIR. Results: DWI/ADC lesions could be diagnosed with the same reliability by the most experienced reader in the 0.55T and 1.5T sequences (specificity 100% and sensitivity 92.9%, respectively). False positive findings did not occur. Detection of number/location of FLAIR lesions was mostly equivalent between 0.55T and 1.5T sequences. No significant difference (p = 0.789-0.104) for FLAIR resolution and contrast was observed regarding Likert scaling. For DWI/ADC noise, the 0.55T sequences were significantly superior (p < 0.026). Otherwise, the 1.5T sequences were significantly superior (p < 0.029). There was no significant difference in infarct volume and volume of infarct demarcation between the 0.55T and 1.5T sequences, when detectable. Conclusions: Low-field MRI stroke imaging at 0.55T may not be inferior to scanners with higher field strengths and thus has great potential as a low-cost alternative in future stroke diagnostics. However, there are limitations in the detection of very small infarcts. Further technical developments with follow-up studies must show whether this problem can be solved.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] New-generation 0.55T MRI in patients with total hip arthroplasty: a comparison with 1.5T MRI
    Breit, H. -C
    Obmann, M.
    Schlicht, F.
    Vosshenrich, J.
    Segeroth, M.
    Bach, M.
    Clauss, M.
    Harder, D.
    Donners, R.
    CLINICAL RADIOLOGY, 2025, 81
  • [22] Initial experience with a next-generation low-field MRI scanner: Potential for breast imaging?
    Dietzel, Matthias
    Laun, Frederik B.
    Heiss, Rafael
    Wenkel, Evelyn
    Bickelhaupt, Sebastian
    Hack, Carolin
    Uder, Michael
    Ohlmeyer, Sabine
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 173
  • [23] Usability of magnetic resonance images acquired at a novel low-field 0.55 T scanner for brain radiotherapy treatment planning
    Grigo, Johanna
    Masitho, Siti
    Fautz, Hans-Peter
    Voigt, Raphaela
    Schonath, Miriam
    Oleszczuk, Annika
    Uder, Michael
    Heiss, Rafael
    Fietkau, Rainer
    Putz, Florian
    Bert, Christoph
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 25
  • [24] Track density imaging using diffusion tensor imaging data from 1.5 T MRI scanner
    Kosker, Fatma Betul
    Gumus, Kazim Ziya
    Tokmakci, Mahmut
    Acer, Niyazi
    Senol, Serkan
    Bilgen, Mehmet
    TURKISH JOURNAL OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCES, 2022, 30 (06) : 2044 - 2053
  • [25] Toward Next-Generation Molecular Imaging with a Clinical Low-Field (0.064 T) Point-of-Care MRI Scanner
    Iqbal, Nadiya
    Brittin, Drew O.
    Daluwathumullagamage, Praveen J.
    Alam, Md Shahabuddin
    Senanayake, Ishani M.
    Gafar, A. Tobi
    Siraj, Zahid
    Petrilla, Anthony
    Pugh, Margaret
    Tonazzi, Brockton
    Ragunathan, Sudarshan
    Poorman, Megan E.
    Sacolick, Laura
    Theis, Thomas
    Rosen, Matthew S.
    Chekmenev, Eduard Y.
    Goodson, Boyd M.
    ANALYTICAL CHEMISTRY, 2024, 96 (25) : 10348 - 10355
  • [26] Comparison of high-field (1.5 T) and low-field (0.2 T) MR imaging for the reliability of meniscal diagnosis (vol 37, pg 802, 1997)
    Rand, T
    Imhof, H
    Breitenseher, M
    Happel, B
    Turetschek, K
    Schneider, B
    Vogele, T
    Kukla, C
    Trattnig, S
    RADIOLOGE, 1998, 38 (03): : 193 - 193
  • [27] Deep Learning Reconstructed New-Generation 0.55 T MRI of the Knee-A Prospective Comparison With Conventional 3 T MRI
    Donners, Ricardo
    Vosshenrich, Jan
    Segeroth, Martin
    Seng, Magdalena
    Fenchel, Matthias
    Nickel, Marcel Dominik
    Bach, Michael
    Schmaranzer, Florian
    Todorski, Inga
    Obmann, Markus M.
    Harder, Dorothee
    Breit, Hanns-Christian
    INVESTIGATIVE RADIOLOGY, 2024, 59 (12) : 823 - 830
  • [28] Economic aspects of low-field magnetic resonance imaging Acquisition, installation, and maintenance costs of 0.55 T systems
    Vosshenrich, Jan
    Breit, Hanns-Christian
    Bach, Michael
    Merkle, Elmar M.
    RADIOLOGE, 2022, 62 (05): : 400 - 404
  • [29] Magnetic resonance imaging on a low-field system (0.2 T): comparison with a 1.0-T system
    Trattnig, S
    Kontaxis, G
    Breitenseher, M
    Czerny, C
    Rand, T
    Turetschek, K
    Barth, M
    Imhof, H
    RADIOLOGE, 1997, 37 (10): : 773 - 777
  • [30] Comparison Between 1.5-T and 3-T MRI for Fetal Imaging: Is There an Advantage to Imaging With a Higher Field Strength?
    Victoria, Teresa
    Johnson, Ann M.
    Edgar, J. Christopher
    Zarnow, Deborah M.
    Vossough, Arastoo
    Jaramillo, Diego
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 206 (01) : 195 - 201