Electrode microdislodgment as a cause of pseudocapture in a VDD pacemaker

被引:0
|
作者
Font, ER [1 ]
Prat, XV [1 ]
Alegret, JM [1 ]
Varela, MC [1 ]
Torner, P [1 ]
Oter, R [1 ]
机构
[1] Hosp Santa Creu & St Pau, Dept Cardiol & Cirugia Cardiaca, Barcelona 08025, Spain
来源
REVISTA ESPANOLA DE CARDIOLOGIA | 1997年 / 50卷 / 12期
关键词
pacemaker; complications; VDD; pseudocapture;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The new AV sequential pacemakers have improved the suitability for the election of the best pacing mode for each patient, The complexity of the systems may mask some dysfunctions. In the presented case, a failure to capture due to microdislodgment, may have been missed in a simple pacemaker control, because of the combination of several factors: the presence of normal AV conduction at that moment, the concordance between the pacemaker stimulus and the conduced QRS complex and the similar morphology of the conduced and paced QRS complex.
引用
收藏
页码:909 / 912
页数:4
相关论文
共 50 条
  • [41] INTRACARDIAC ECG IN PACEMAKER ELECTRODE INSERTION
    CHATTERJEE, K
    SUTTON, R
    BMJ-BRITISH MEDICAL JOURNAL, 1968, 1 (5595): : 838 - +
  • [42] Pacemaker failure as a cause of sudden death
    Morin, Daniel P.
    VIRCHOWS ARCHIV, 2018, 472 (02) : 293 - 293
  • [43] A rare cause of stroke in a pacemaker patient
    Laury, P
    Chabert, JP
    Blaise, C
    Pérotin, S
    Brasselet, C
    Schlick, C
    Tassan-Mangina, S
    Metz, D
    Elaerts, J
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 2002, 95 (03): : 219 - 222
  • [44] Mycobacterium abscessus as a cause of pacemaker infection
    Kessler, AT
    Kourtis, AP
    MEDICAL SCIENCE MONITOR, 2004, 10 (10): : CS60 - CS62
  • [45] Pacemaker failure as a cause of sudden death
    Daniel P. Morin
    Virchows Archiv, 2018, 472 : 293 - 293
  • [46] FRACTURE OF A TEMPORARY PACEMAKER ELECTRODE CATHETER
    MCRAVEN, DR
    LUCHI, RJ
    FUNK, DC
    SOUTHERN MEDICAL JOURNAL, 1978, 71 (12) : 1573 - 1575
  • [47] PACEMAKER ELECTRODE POLARIZATION - NEW DIRECTIONS
    ASTRINSKY, E
    FURMAN, S
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (02): : 298 - 298
  • [48] The pacemaker-twiddler's syndrome: An infrequent cause of pacemaker failure
    Salahuddin M.
    Cader F.A.
    Nasrin S.
    Chowdhury M.Z.
    BMC Research Notes, 9 (1)
  • [49] PACEMAKER ELECTRODE INFECTION PRESENTING WITH SPONDYLODISCITIS
    BONAL, J
    BOUCHIAT, C
    DEJAUREGUIBERRY, JP
    DUVAL, JL
    CARLI, P
    CHAGNON, A
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1994, 87 (12): : 1735 - 1738
  • [50] IMPROVED TRANSVENOUS PACEMAKER ELECTRODE STABILITY
    GORDON, S
    TIMMIS, GC
    GANGADHARAN, V
    RAMOS, R
    HAUSER, J
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (05): : A90 - A90