Electrode microdislodgment as a cause of pseudocapture in a VDD pacemaker

被引:0
|
作者
Font, ER [1 ]
Prat, XV [1 ]
Alegret, JM [1 ]
Varela, MC [1 ]
Torner, P [1 ]
Oter, R [1 ]
机构
[1] Hosp Santa Creu & St Pau, Dept Cardiol & Cirugia Cardiaca, Barcelona 08025, Spain
来源
REVISTA ESPANOLA DE CARDIOLOGIA | 1997年 / 50卷 / 12期
关键词
pacemaker; complications; VDD; pseudocapture;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The new AV sequential pacemakers have improved the suitability for the election of the best pacing mode for each patient, The complexity of the systems may mask some dysfunctions. In the presented case, a failure to capture due to microdislodgment, may have been missed in a simple pacemaker control, because of the combination of several factors: the presence of normal AV conduction at that moment, the concordance between the pacemaker stimulus and the conduced QRS complex and the similar morphology of the conduced and paced QRS complex.
引用
收藏
页码:909 / 912
页数:4
相关论文
共 50 条
  • [21] ATRIAL SENSING PERFORMANCE OF THE SINGLE-LEAD VDD PACEMAKER DURING EXERCISE
    VARRIALE, P
    CHRYSSOS, BE
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (07) : 1854 - 1857
  • [22] A NEW PACEMAKER COMBINE VDD PERFORMANCE WITH OUTPUT AND TIMING AUTOMATIC-CONTROL
    AUDOGLIO, R
    MOSCHINI, GF
    IARLORI, A
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03): : 637 - 637
  • [23] An uncommon cause of pacemaker malfunction
    Whymark, C
    Wilson, J
    ANAESTHESIA, 2002, 57 (07) : 728 - 728
  • [24] HANDLING PACEMAKER ELECTRODE DISPLACEMENT
    AZZAM, EG
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1974, 228 (11): : 1368 - 1368
  • [25] PERCUTANEOUS PACEMAKER ELECTRODE REMOVAL
    SMITH, CW
    MESSENGER, JC
    WEINBERG, DM
    SHANER, SP
    RENNER, JW
    RADIOLOGY, 1992, 185 : 310 - 311
  • [26] ELECTRODE CAUSATION OF PACEMAKER INHIBITION
    FURMAN, S
    DECAPRIO, V
    CHEST, 1977, 72 (01) : 117 - 119
  • [27] FAILURE OF PACEMAKER ELECTRODE LEADS
    KAUL, TK
    GREEN, GD
    BAIN, WH
    EUROPEAN JOURNAL OF CARDIOLOGY, 1979, 10 (05): : 385 - 394
  • [28] ELECTRODE FRACTURE IN PACEMAKER PATIENTS
    ALSUKHUN, M
    BARDOS, P
    MESSMER, J
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1980, 105 (15) : 520 - 522
  • [29] Development and preliminary clinical tests of an impedance sensing VDD recording pacemaker for diagnosis and research
    Edgar, DR
    Horwood, JMK
    Woollons, DJ
    Baker, J
    Shaw, DB
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (04) : 517 - 522
  • [30] Development and preliminary clinical tests of an impedance sensing VDD recording pacemaker for diagnosis and research
    D. R. Edgar
    J. M. K. Horwood
    D. J. Woollons
    J. Baker
    D. B. Shaw
    Medical & Biological Engineering & Computing, 1999, 37 : 517 - 522