Oscillator-Based Reactance Sensors With Injection Locking for High-Throughput Flow Cytometry Using Microwave Dielectric Spectroscopy

被引:68
|
作者
Chien, Jun-Chau [1 ]
Niknejad, Ali M. [1 ,2 ]
机构
[1] Univ Calif Berkeley, Dept Elect Engn & Comp Sci, Berkeley, CA 94720 USA
[2] Berkeley Wireless Res Ctr BWRC, Berkeley, CA USA
关键词
Capacitive sensor; dielectric spectroscopy; flow cytometry; injection-locked oscillators (ILOs); interferometry; microfluidics; nested chopping; permittivity; PHASE NOISE; IMPEDANCE CYTOMETRY; CHEMICAL SENSOR; COUPLED QVCO; SINGLE; CELLS; ARRAY;
D O I
10.1109/JSSC.2015.2500362
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents the analysis and design of oscillator-based reactance sensors employing injection locking for high-throughput label-free single-cell analysis using dielectric spectroscopy at microwave frequencies. By injection-locking two sensing LC-oscillators with an I/Q excitation source, the measurement of the sample-induced frequency shift caused by the interaction with the electromagnetic fields is performed through phase detection with injection-strength-dependent transducer gain. Such inherent phase amplification offered by the injection locking not only relaxes the design requirement for the readout circuits but also maintains the highest rejection against common-mode errors associated with the drift of the supply voltage and the environmental parameters. To reduce flicker noise contribution, a chopping technique employing phase modulation is exploited. In addition, this paper presents a novel ping-pong chopping approach to alleviate chopping-induced dc offset. In this prototype, four sensing channels, covering frequencies between 6.5 and 30 GHz, are distributed along a microfluidic channel fabricated with standard photolithography. Measurements show that the proposed microwave capacitive sensors achieve a sub-aF(rms) of noise sensitivity at 100 kHz filtering bandwidth, enabling measurement throughput exceeding 1 k cells/s. The sensor prototype is implemented in 65 nm CMOS technology and consumes 65 mW at 1 V supply.
引用
收藏
页码:457 / 472
页数:16
相关论文
共 50 条
  • [1] A high-throughput flow cytometry-on-a-CMOS platform for single-cell dielectric spectroscopy at microwave frequencies
    Chien, Jun-Chau
    Ameri, Ali
    Yeh, Erh-Chia
    Killilea, Alison N.
    Anwar, Mekhail
    Niknejad, Ali M.
    LAB ON A CHIP, 2018, 18 (14) : 2065 - 2076
  • [2] Cell-Based Screening Using High-Throughput Flow Cytometry
    Black, Christopher B.
    Duensing, Thomas D.
    Trinkle, Linda S.
    Dunlay, R. Terry
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2011, 9 (01) : 13 - 20
  • [3] A high-throughput method for detection of DNA in chloroplasts using flow cytometry
    Rowan, Beth A.
    Oldenburg, Delene J.
    Bendich, Arnold J.
    PLANT METHODS, 2007, 3 (1)
  • [4] High-throughput observation and quantification of ADCC using imaging flow cytometry
    Helguera, Gustavo
    Rodriguez, Jose
    Henery, Shannon
    Catterton, Paul
    Martinez-Maza, Otoniel
    Penichet, Manuel
    CANCER RESEARCH, 2009, 69
  • [5] A high-throughput method for detection of DNA in chloroplasts using flow cytometry
    Beth A Rowan
    Delene J Oldenburg
    Arnold J Bendich
    Plant Methods, 3
  • [6] A high-throughput cell-based toxicity analysis of drug metabolites using flow cytometry
    E. J. Buenz
    Cell Biology and Toxicology, 2007, 23 : 361 - 365
  • [7] High-Throughput Measurement of Small-Molecule Enantiopurity by Using Flow Cytometry
    Tan, Zhesen
    Heemstra, Jennifer M.
    CHEMBIOCHEM, 2018, 19 (17) : 1853 - 1857
  • [8] A high-throughput unscheduled DNA synthesis (UDS) assay using flow cytometry
    Kirk, M.
    Reeder, M.
    Ripper, T.
    Fischl, M.
    Cerven, D.
    DeGeorge, G.
    IN VITRO CELLULAR & DEVELOPMENTAL BIOLOGY-ANIMAL, 2005, 41 : 61A - 61A
  • [9] High-throughput small-molecule enantiopurity measurement using flow cytometry
    Tan, Zhesen
    Manna, Arunava
    Heemstra, Jennifer
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 256
  • [10] A high-throughput cell-based toxicity analysis of drug metabolites using flow cytometry.
    Buenz, E. J.
    CELL BIOLOGY AND TOXICOLOGY, 2007, 23 (05) : 361 - 365