Evaluation of GAFCHROMIC® EBT film for CyberKnife® dosimetry

被引:111
|
作者
Wilcox, Ellen E. [1 ]
Daskalov, George M. [1 ]
机构
[1] St Francis Hosp & Med Ctr, Dept Radiat Oncol, Hartford, CT 06105 USA
关键词
EBT film dosimetry; stereotactic radiosurgery;
D O I
10.1118/1.2734384
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
External beam therapy (EBT) GAFCHROMIC (R) film is evaluated for dosimetry and characterization of the CyberKnife (R) radiation beams. Percentage depth doses, lateral beam profiles, and output factors are measured in solid water using EBT GAFCHROMIC (R) film (International Specialty Products, Wayne, NJ) for the 6 MV radiation beams of diameter 5 to 60 rum produced by the CyberKnife (R) (Accuray, Sunnyvale, CA). The data are compared to those measured with the PTW 60008 diode and the Wellhofer CC01 ion chamber in water. For the small radiation field sizes used in stereotactic radiosurgery, lateral electronic disequilibrium and steep dose gradients exist in a large portion of these fields, requiring the use of high-resolution measurement techniques. For small beams, the detector size approaches the dimensions of the beam and adversely affects measurement accuracy in regions where the gradient varies across the detector. When film is the detector, the scanning system is usually the resolution-limiting component. Radiographic films based upon silver halide (AgH) emulsions are widely used for relative dosimetry of external radiation treatment beams in the megavoltage energy range, because of their good spatial resolution and capability to provide integrated dosimetry over two dimensions. Film dosimetry, however, has drawbacks due to its steep energy dependence at low photon energies as well as film processor and densitometer artifacts. EBT radiochromic film, introduced in 2004 specifically for IMRT dosimetry, may be a detector of choice for the characterization of small radiosurgical beams, because of its near-tissue equivalence, radiation beam energy independence, high spatial resolution, and self developing properties. For radiation beam sizes greater than 10 rum, the film measurements were identical to those of the diode and ion chamber. For the smaller beam diameters of 7.5 and 5 rum, however, there were differences in the data measured with the different detectors, which are attributed to their different spatial resolution and non-water-equivalence. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1967 / 1974
页数:8
相关论文
共 50 条
  • [41] Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry
    Ferreira, B. C.
    Lopes, M. C.
    Capela, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (04): : 1073 - 1085
  • [42] I-125 Seed Dosimetry in the Near Field Using Gafchromic® EBT Film
    Chiu-Tsao, S.
    Hanley, J.
    Napoli, J.
    Fan, J.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [43] Dose delivery verification for Cyberknife image guided frameless intracranial treatments using GAFChromic EBT film
    Marijnissen, H
    de Jong, A
    van De Baan, P
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S214 - S214
  • [44] Polarized dosimetry method for Gafchromic EBT3
    Perez Azorin, Jose Fernando
    Ramos Garcia, Luis Isaac
    Maza Ozcoidi, Diego
    Almansa, Julio F.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (08): : 972 - 980
  • [45] Systematic Evaluation of Uncertainties Associated with GAFCHROMIC EBT2 Film Dosimetry for 6MV Photon Beams
    Kim, J.
    Kim, S.
    Shaikh, M.
    Li, H.
    Huang, Y.
    Wen, N.
    Glide-Hurst, C.
    Jin, J.
    Nurushev, T.
    Chetty, I. J.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [46] Evaluation of EBT-2 model GAFCHROMIC™ film performance in water
    Aldelaijan, Saad
    Devic, Slobodan
    Mohammed, Huriyyah
    Tomic, Nada
    Liang, Li-Heng
    DeBlois, Francois
    Seuntjens, Jan
    MEDICAL PHYSICS, 2010, 37 (07) : 3687 - 3693
  • [47] CALIBRATION OF GAFCHROMIC EBT3 FILM FOR DOSIMETRY OF SCANNING PROTON PENCIL BEAM (PBS)
    Krzempek, D.
    Mianowska, G.
    Bassler, N.
    Stolarczyk, L.
    Kopec, R.
    Sas-Korczynska, B.
    Olko, P.
    RADIATION PROTECTION DOSIMETRY, 2018, 180 (1-4) : 324 - 328
  • [48] Characterisation of EPSONV700 flatbed scanner for EBT3 Gafchromic film dosimetry
    Nelson, V.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S737 - S738
  • [49] High dose per fraction dosimetry of small fields with Gafchromic EBT2 film
    Hardcastle, Nicholas
    Basavatia, Amar
    Bayliss, Adam
    Tome, Wolfgang A.
    MEDICAL PHYSICS, 2011, 38 (07) : 4081 - 4085
  • [50] Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system
    Aldelaijan, Saad
    Devic, Slobodan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 49 : 112 - 118