Comparison of Tissue Doppler Velocities Obtained by Different Types of Echocardiography Systems: Are They Compatible?

被引:20
|
作者
Denes, Monika [1 ]
Farkas, Katalin [2 ]
Erdei, Tamas [1 ]
Lengyel, Maria [1 ]
机构
[1] Gottsegen Gyorgy Hungarian Inst Cardiol, H-1096 Budapest, Hungary
[2] Szent Imre Teaching Hosp, Budapest, Hungary
关键词
tissue Doppler imaging; diastolic function; systolic function; left ventricular filling pressures; VENTRICULAR DIASTOLIC FUNCTION; HEART-FAILURE; PROGNOSTIC VALUE; IMPACT;
D O I
10.1111/j.1540-8175.2009.01018.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Both systolic and diastolic tissue Doppler (TD) velocities have an important diagnostic and prognostic role in cardiology. We aimed to compare TD velocities between two different echocardiography systems. Patients: Thirty-one consecutive patients (mean age: 65.2 +/- 17.5 years; 12 males) were enrolled. Methods: Systolic (Sa), early (Ea), and late (Aa) diastolic velocities were measured by TD at the lateral mitral annulus by a Sonos 2000 (Hewlett-Packard, Andover, MA, USA) and a Philips iE33 system. The E/Ea ratio was calculated. Results: Ea, Aa, and Sa velocities were higher when measured by the Sonos system (Ea: 13.2 +/- 4.1 cm/s vs. 8.3 +/- 3.6 cm/s; Aa: 14.8 +/- 3.8 cm/s vs. 9.3 +/- 2.3 cm/s; Sa: 15.2 +/- 3.6 cm/s vs. 8.4 +/- 2.0 cm/s; P < 0.0001 all). A significant correlation was found in Ea and in Ea/Aa (r = 0.84 and r = 0.85 resp; P < 0.0001 for both), and a weaker in Aa (r = 0.43; P = 0.02) between the machines. The Bland-Altman analysis showed broad limits of agreement between the measurements for Ea, Aa, and Sa (mean difference: 4.95 cm/s; 5.52 cm/s; 6.73 cm/s, respectively; limits: 0.64-9.25 cm/s; -1.39-12.39 cm/s; -0.37-13.83 cm/s, respectively). An E/Ea ratio > 5.6 by the Sonos system showed 75% sensitivity and 79% specificity for elevated left ventricular filling pressure, defined as E/Ea > 10 by the reference Philips system. Conclusions: Although diastolic TD velocities had excellent correlations between the two machines, there was a systematic overestimation by the Sonos system. Since the limits of agreement do not allow replacing the measurements, we suggest using the same echocardiographic equipment at patient follow-up. (Echocardiography 2010;27:230-235).
引用
收藏
页码:230 / 235
页数:6
相关论文
共 50 条
  • [1] Comparison of Two Different Ultrasound Systems for the Evaluation of Tissue Doppler Velocities in Fetuses
    Cruz-Lemini, Monica
    Valenzuela-Alcaraz, Brenda
    Figueras, Francesc
    Sitges, Marta
    Gomez, Olga
    Maria Martinez, Josep
    Bijnens, Bart
    Gratacos, Eduard
    Crispi, Fatima
    FETAL DIAGNOSIS AND THERAPY, 2016, 40 (01) : 35 - 40
  • [2] Discrepancies between Tissue Doppler Velocities on Different Echocardiography Machines: Comparing Apples to Oranges
    Halpern, Dan G.
    Sanchez-Ross, Monica
    Joshi, Sandeep
    Cantales, Deborah
    Chaudhry, Farooq A.
    Sherrid, Mark V.
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2013, 30 (06): : 730 - 732
  • [3] Intervendor Variabilities of Left and Right Ventricular Myocardial Velocities among Three Tissue Doppler Echocardiography Systems
    Abe, Kiyoshi
    Yuda, Satoshi
    Sato, Yasumi
    Yasui, Kenji
    Nishi, Ryosuke
    Hanada, Kyosuke
    Hashimoto, Naoko
    Kawamukai, Mina
    Kouzu, Hidemichi
    Muranaka, Atsuko
    Hashimoto, Akiyoshi
    Tsuchihashi, Kazufumi
    Watanabe, Naoki
    Miura, Tetsuji
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2015, 32 (12): : 1790 - 1801
  • [4] Are measurements of systolic myocardial velocities and displacement with colour and spectral Tissue Doppler compatible?
    Aristomenis Manouras
    Arben Shala
    Evangelia Nyktari
    Kambiz Shahgaldi
    Reidar Winter
    Panagiotis Vardas
    Lars-Åke Brodin
    Jacek Nowak
    Cardiovascular Ultrasound, 7
  • [5] Are measurements of systolic myocardial velocities and displacement with colour and spectral Tissue Doppler compatible?
    Manouras, Aristomenis
    Shala, Arben
    Nyktari, Evangelia
    Shahgaldi, Kambiz
    Winter, Reidar
    Vardas, Panagiotis
    Brodin, Lars-Ake
    Nowak, Jacek
    CARDIOVASCULAR ULTRASOUND, 2009, 7
  • [6] Tissue Doppler echocardiography reveals distinct patterns of impaired myocardial velocities in different degrees of coronary artery disease
    Hoffmann, Soren
    Mogelvang, Rasmus
    Olsen, Niels Thue
    Sogaard, Peter
    Fritz-Hansen, Thomas
    Bech, Jan
    Galatius, Soren
    Madsen, Jan Kyst
    Jensen, Jan Skov
    EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2010, 11 (06): : 544 - 549
  • [7] Comparison of myocardial performance index obtained either by conventional echocardiography or tissue Doppler echocardiography in healthy subjects and patients with heart failure
    Duzenli, Mehmet Akif
    Ozdemir, Kurtulus
    Aygul, Nazif
    Soylu, Ahmet
    Aygul, Meryem Ulku
    Gok, Hasan
    HEART AND VESSELS, 2009, 24 (01) : 8 - 15
  • [8] Comparison of myocardial performance index obtained either by conventional echocardiography or tissue Doppler echocardiography in healthy subjects and patients with heart failure
    Mehmet Akif Duzenli
    Kurtulus Ozdemir
    Nazif Aygul
    Ahmet Soylu
    Meryem Ulku Aygul
    Hasan Gök
    Heart and Vessels, 2009, 24 : 8 - 15
  • [9] Method of myocardial velocities measurements based on combination of echocardiography and Doppler tissue imaging
    Fitton, I
    Shen, J
    Perron, JM
    Roudaut, R
    Barat, JL
    ISPA 2001: PROCEEDINGS OF THE 2ND INTERNATIONAL SYMPOSIUM ON IMAGE AND SIGNAL PROCESSING AND ANALYSIS, 2001, : 232 - 237
  • [10] Quantification of longitudinal myocardial velocities by Doppler tissue imaging during clobutamine echocardiography
    Cosson, S
    Carrat, F
    Lamisse, N
    Aouad, A
    Albo, C
    Cohen, A
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 2004, 97 (04): : 311 - 319