Comparative Dose of Contrast-Enhanced Spectral Mammography (CESM), Digital Mammography, and Digital Breast Tomosynthesis

被引:42
|
作者
Phillips, Jordana [1 ,2 ]
Mihai, Georgeta [1 ,2 ]
Hassonjee, Sarah Esaa [1 ,2 ]
Raj, Sean D. [1 ,2 ]
Palmer, Matthew R. [1 ,2 ]
Brook, Alexander [1 ,2 ]
Zhang, Da [1 ,2 ]
机构
[1] Harvard Med Sch, Dept Radiol, 330 Brookline Ave,4th Fl, Boston, MA 02215 USA
[2] Beth Israel Deaconess Med Ctr, 330 Brookline Ave,4th Fl, Boston, MA 02215 USA
关键词
contrast-enhanced digital mammography; contrast-enhanced mammography; contrast-enhanced spectral mammography; radiation dose; DOSIMETRY PROTOCOLS; RADIATION-EXPOSURE; PERFORMANCE; MRI; UK;
D O I
10.2214/AJR.17.19036
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to provide a more accurate estimation of the radiation dose of contrast-enhanced spectral mammography (CESM) relative to that of 2D digital mammography and tomosynthesis using phantom and patient data and an accepted dosimetry protocol that eliminates vendor-specific average glandular dose (AGD) estimates while including breast density. MATERIALS AND METHODS. Patient and phantom AGD estimation was performed using two vendors (system 1 and system 2) in five imaging modes, including 2D, 3D, and CESM imaging. Patient AGD was retrospectively estimated from 45 patients who underwent mammography with all imaging modes during 2012-2016. Patient and phantom AGD were estimated using accepted European and International Atomic Energy Agency protocols for dosimetry and were compared across imaging modes using a paired t test with Bonferroni correction. RESULTS. Phantom data showed that the imaging modes with the lowest to highest AGDs were system 1 2D, followed by system 2 2D and system 2 3D, which had comparable values (p = 0.6), followed by system 1 CESM, and then by system 2 2D plus 3D. One hundred eighty views in 45 patients showed that the system 1 CESM AGD was 1.8 times greater than the system 1 21) AU) (p < 0.001), 1.2 times greater than the system 2 21) AO) (p < 0.001), 1.2 times greater than the system 2 3D AGD (p < 0.001), and 0.6 times less than the system 2 2D plus 3D AGD (p < 0.001). CONCLUSION. The CESM dose for system 1 is within an acceptable range as compared with other commonly performed mammographic examinations and should not preclude its use as a diagnostic breast imaging tool.
引用
收藏
页码:839 / 846
页数:8
相关论文
共 50 条
  • [31] Contrast-Enhanced Digital Mammography in the Surgical Management of Breast Cancer
    Ali-Mucheru, Mariam
    Pockaj, Barbara
    Patel, Bhavika
    Pizzitola, Victor
    Wasif, Nabil
    Stucky, Chee-Chee
    Gray, Richard
    ANNALS OF SURGICAL ONCOLOGY, 2016, 23 : S649 - S655
  • [33] Contrast-Enhanced Digital Mammography in the Surgical Management of Breast Cancer
    Ali-Mucheru, Mariam
    Patel, Bhavika
    Pockaj, Barbara
    Pizzitola, Victor
    Wasif, Nabil
    Stucky, Chee-Chee
    Gray, Richard
    ANNALS OF SURGICAL ONCOLOGY, 2016, 23 : 14 - 15
  • [34] Contrast-Enhanced Digital Mammography in the Surgical Management of Breast Cancer
    Mariam Ali-Mucheru
    Barbara Pockaj
    Bhavika Patel
    Victor Pizzitola
    Nabil Wasif
    Chee-Chee Stucky
    Richard Gray
    Annals of Surgical Oncology, 2016, 23 : 649 - 655
  • [35] Assessment of disease extent on contrast-enhanced MRI in breast cancer detected at digital breast tomosynthesis versus digital mammography alone
    Chudgar, A. V.
    Conant, E. F.
    Weinstein, P.
    Keller, M.
    Synnestvedt, M.
    Yamartino, P.
    McDonald, E. S.
    CLINICAL RADIOLOGY, 2017, 72 (07) : 573 - 579
  • [36] Dual-energy contrast-enhanced spectral mammography (CESM) for breast cancer screening
    Sorin, Vera
    Sklair-Levy, Miri
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (11) : 1914 - 1917
  • [37] Average Glandular Dose in Digital Mammography and Breast Tomosynthesis
    Olgar, T.
    Kahn, T.
    Gosch, D.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2012, 184 (10): : 911 - 918
  • [38] Contrast-enhanced spectral mammography (CESM) compared with breast MRI for breast cancer detection.
    Liao, Lydia
    Li, Luna
    Germaine, Pauline
    Tinney, Elizabeth
    JOURNAL OF CLINICAL ONCOLOGY, 2015, 33 (15)
  • [39] Does Contrast-Enhanced Mammography Outperform Digital Breast Tomosynthesis for Detection and Characterization of Breast Lesions or Vice Versa?
    Singla, Veenu
    Talasila, Pallavi
    Soni, Saumya
    Singh, Tulika
    Khare, Siddhant
    Bal, Amanjit
    INDIAN JOURNAL OF SURGICAL ONCOLOGY, 2025, 16 (01) : 333 - 343
  • [40] Radiation Exposure of Contrast-Enhanced Spectral Mammography Compared With Full-Field Digital Mammography
    Jeukens, Cecile R. L. P. N.
    Lalji, Ulrich C.
    Meijer, Eduard
    Bakija, Betina
    Theunissen, Robin
    Wildberger, Joachim E.
    Lobbes, Marc B. I.
    INVESTIGATIVE RADIOLOGY, 2014, 49 (10) : 659 - 665