A neutron track etch detector for electron linear accelerators in radiotherapy

被引:10
|
作者
Vukovic, Branko [1 ]
Faj, Dario [2 ]
Poje, Marina [1 ]
Varga, Maja [1 ]
Radolic, Vanja [1 ]
Miklavcic, Igor [1 ]
Ivkovic, Ana [2 ]
Planinic, Josip [1 ]
机构
[1] Univ Osijek, Dept Phys, HR-31000 Osijek, Croatia
[2] Clin Hosp Osijek, Osijek, Croatia
关键词
electron linear accelerator; photoneutron; track etch detector; neutron dose equivalent; SPECTRA;
D O I
10.2478/v10019-010-0003-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. Electron linear accelerators in medical radiotherapy have replaced cobalt and caesium sources of radiation. However, medical accelerators with photon energies over 10 MeV generate undesired fast neutron contamination in a therapeutic X-ray photon beam. Photons with energies above 10 MeV can interact with the atomic nucleus of a high-Z material, of which the target and the head of an accelerator consist, and lead to the neutron ejection. Results and conclusions. Our neutron dosimeter, composed of the LR-115 track etch detector and boron foil BN-1 converter, was calibrated on thermal neutrons generated in the nuclear reactor of the Josef Stefan Institute (Slovenia), and applied to dosimetry of undesirable neutrons in photon radiotherapy by the linear accelerator 15 MV Siemens Mevatron. Having considered a high dependence of a cross-section between neutron and boron on neutron energy, and broad neutron spectrum in a photon beam, as well as outside the entrance door to maze of the Mevatron, we developed a method for determining the effective neutron detector response. A neutron dose rate in the photon beam was measured to be 1.96 Sv/h. Outside the Mevatron room the neutron dose rate was 0.62 mu Sv/h. PACS: 87.52. Ga; 87.53.St; 29.40.Wk.
引用
收藏
页码:62 / 66
页数:5
相关论文
共 50 条
  • [1] NEW LINEAR ELECTRON ACCELERATORS FOR RADIOTHERAPY
    NYGARD, JC
    KELLIHER, MG
    PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1955, 43 (03): : 374 - 374
  • [2] Cosmic radiation dose in aircraft - a neutron track etch detector
    Vukovic, B.
    Radolic, V.
    Miklavcic, I.
    Poje, M.
    Varga, M.
    Planinic, J.
    JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2007, 98 (03) : 264 - 273
  • [3] HISTORY OF MICROWAVE ELECTRON LINEAR ACCELERATORS FOR RADIOTHERAPY
    GINZTON, EL
    NUNAN, CS
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (02): : 205 - 216
  • [4] RADIOTHERAPY ELECTRON LINEAR ACCELERATORS - HISTORY AND CONTEMPORARY DEVELOPMENTS
    KARZMARK, CJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (01): : 114 - &
  • [5] Neutron detection on the Foton-M2 satellite by a track etch detector stack
    Palfalvi, J. K.
    Szabo, J.
    Dudas, B.
    RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) : 590 - 594
  • [6] THE TRACK-ETCH DETECTOR OF THE MACRO EXPERIMENT
    不详
    NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1988, 15 (1-4): : 331 - 336
  • [7] Neutron track detector for photo neutron measurements from high energy medical linear accelerator
    Cyriac, S.
    Kummali, A.
    Mm, M.
    Raman, G. R.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S150 - S150
  • [8] DOSIMETRIC PROPERTIES OF DIFFERENT CR-39 PLASTICS USED AS NEUTRON RECOIL TRACK ETCH DETECTOR
    PIESCH, E
    URBAN, M
    NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1986, 12 (1-6): : 661 - 664
  • [9] Properties of TNF-1 track etch detector
    Ogura, K
    Asano, M
    Yasuda, N
    Yoshida, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2001, 185 : 222 - 227
  • [10] Calculation of photoneutrons produced in the targets of electron linear accelerators for radiography and radiotherapy applications
    Huang, WL
    Li, QF
    Lin, YZ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 229 (3-4): : 339 - 347