DYNAMIC JAWS AND DYNAMIC COUCH IN HELICAL TOMOTHERAPY

被引:60
|
作者
Sterzing, Florian [1 ]
Uhl, Matthias [1 ]
Hauswald, Henrik [1 ]
Schubert, Kai [1 ]
Sroka-Perez, Gabriele [1 ]
Chen, Yu [2 ]
Lu, Weiguo [2 ]
Mackie, Rock [2 ]
Debus, Juergen [1 ]
Herfarth, Klaus [1 ]
Oliveira, Gustavo [2 ]
机构
[1] Univ Heidelberg, Dept Radiat Oncol, D-69120 Heidelberg, Germany
[2] Tomotherapy Inc, Madison, WI USA
关键词
Helical tomotherapy; Dynamic jaws; Dynamic couch; Longitudinal dose profile; Nasopharyngeal cancer; MODULATED RADIATION-THERAPY; NASOPHARYNX CANCER; CONFORMAL RADIOTHERAPY; NECK-CANCER; IMRT; HEAD; DELINEATION; IRRADIATION; EXPERIENCE; DELIVERY;
D O I
10.1016/j.ijrobp.2009.07.1686
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the next generation of helical tomotherapy delivery with dynamic jaw and dynamic couch movements. Methods and Materials: The new technique of dynamic jaw and dynamic couch movements is described, and a comparative planning study is performed. Ten nasopharyngeal cancer patients with skull base infiltration were chosen for this comparison of longitudinal dose profiles using regular tomotherapy delivery, running-start-stop treatment, and dynamic jaw and dynamic couch delivery. A multifocal simultaneous integrated boost concept was used (70.4Gy to the primary tumor and involved lymph nodes; 57.4Gy to the bilateral cervical lymphatic drainage pathways, 32 fractions). Target coverage, conformity, homogeneity, sparing of organs at risk, integral dose, and radiation delivery time were evaluated. Results: Mean parotid dose for all different deliveries was between 24.8 and 26.1Gy, without significant differences. The mean integral dose was lowered by 6.3% by using the dynamic technique, in comparison with a 2.5-cm-field width for regular delivery and 16.7% with 5-cm-field width for regular delivery. Dynamic jaw and couch movements reduced the calculated radiation time by 66% of the time required with regular 2.5-cm-field width delivery (199 sec vs. 595 sec, p<0.001). Conclusions: The current delivery mode of helical tomotherapy produces dose distributions with conformal avoidance of parotid glands, brain stem, and spinal cord. The new technology with dynamic jaw and couch movements improves the plan quality by reducing the dose penumbra and thereby reducing the integral dose. In addition, radiation time is reduced by 66% of the regular delivery time. (C) 2010 Elsevier Inc.
引用
收藏
页码:1266 / 1273
页数:8
相关论文
共 50 条
  • [21] QA for dynamic aspects of helical tomotherapy treatment delivery
    Schubert, K.
    Sroka-Perez, G.
    Wagenknecht, K.
    Caprile-Etchart, P.
    Sterzing, F.
    Debus, J.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S82 - S82
  • [22] ON THE DOSIMETRIC CHALLENGES OF DYNAMIC JAWS TOMOTHERAPY: A MONTE CARLO STUDY.
    Sterpin, E.
    Chen, Y.
    Chen, Q.
    Lu, W.
    Mackie, T.
    Vynckier, S.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S519 - S519
  • [23] Efficacy of Stereotactic Radiation Therapy for Brain Metastases Using Dynamic Jaws Technology in the Helical Tomotherapy System: Planning and Clinical Studies
    Murai, T.
    Hayashi, A.
    Mori, Y.
    Yoshihiko, M.
    Yanagi, T.
    Matsuo, M.
    Takaoka, T.
    Ogawa, Y.
    Kondo, T.
    Shibamoto, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E568 - E568
  • [24] Dynamic tangents and topotherapy: New delivery capabilities for helical tomotherapy
    Olivera, G
    Ruchala, K
    Lu, W
    Haimerl, J
    Schnarr, E
    Langen, K
    Kupelian, P
    Meeks, S
    Mackie, T
    MEDICAL PHYSICS, 2005, 32 (06) : 2034 - 2034
  • [25] Helical tomotherapy with dynamic running-start-stop delivery compared to conventional tomotherapy delivery
    Rong, Yi
    Chen, Yu
    Shang, Lu
    Zuo, Li
    Lu, Weiguo
    Chen, Quan
    MEDICAL PHYSICS, 2014, 41 (05)
  • [26] Dosimetric impact on different target separations in craniocaudal direction using dynamic jaws in Tomotherapy
    Lam, W. W.
    Geng, H.
    Kong, C. W.
    Ho, Y. W.
    Yang, B.
    Chiu, T. L.
    Choi, H. F.
    Cheung, K. Y.
    Yu, S. K.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S897 - S897
  • [27] A NEW METHOD FOR OUTPUT FACTORS MEASUREMENTS/MC COMPUTATIONS FOR STEREOTACTIC AND DYNAMIC JAWS TOMOTHERAPY
    Sterpin, E.
    Hundertmark, B.
    Mackie, T.
    Lu, W.
    Olivera, G.
    Vynckier, S.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S95 - S95
  • [28] Efficacy of the Dynamic Jaw Mode in Helical Tomotherapy With Static Ports for Breast Cancer
    Sugie, C.
    Manabe, Y.
    Hayashi, A.
    Murai, T.
    Takaoka, T.
    Hattori, Y.
    Takenaka, R.
    Shibamoto, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S222 - S222
  • [29] Efficacy of the Dynamic Jaw Mode in Helical Tomotherapy With Static Ports for Breast Cancer
    Sugie, Chikao
    Manabe, Yoshihiko
    Hayashi, Akihiro
    Murai, Taro
    Takaoka, Taiki
    Hattori, Yukiko
    Iwata, Hiromitsu
    Takenaka, Ran
    Shibamoto, Yuta
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2015, 14 (04) : 459 - 465
  • [30] Monte Carlo Evaluation of the Dose Calculation Algorithm of TomoTherapy for Clinical Cases in Dynamic Jaws Mode
    Sterpin, E.
    MEDICAL PHYSICS, 2015, 42 (06) : 3463 - 3463