Biomechanical cadaver study of proximal fixation in a minimally invasive bipolar construct

被引:12
|
作者
Gaume, M. [1 ,2 ]
Persohn, S. [1 ]
Vergari, C. [1 ]
Glorion, C. [2 ]
Skalli, W. [1 ]
Miladi, L. [2 ]
机构
[1] LBM Inst Biomecan Humaine Georges Charpak, Arts & Metiers ParisTech, 151 Bd Hop, F-75013 Paris, France
[2] Paris Descartes Univ, Necker Hosp, AP HP, Pediat Orthoped Dept, Paris, France
关键词
Thoracic fixation; Biomechanical cadaver study; Neuromuscular scoliosis; Minimally invasive surgery; Bipolar construct; PEDICLE SCREW FIXATION; THORACIC SPINE; INSTRUMENTATION; KINEMATICS; STABILITY; SEGMENT; DEVICE; MOTION; HOOKS;
D O I
10.1007/s43390-019-00014-2
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study designBiomechanical human cadaver study.ObjectiveTo determine the three-dimensional intervertebral ranges of motion (ROMs) of intact and hook-instrumented thoracic spine specimens subjected to physiological loads, using an in vitro experimental protocol with EOS biplane radiography.Summary of background dataPedicle screws are commonly used in thoracic instrumentation constructs, and their biomechanical properties have been widely studied. Promising clinical results have been reported using a T1-T5 thoracic hook-claw construct for proximal rod anchoring. Instrumentation stability is a crucial factor in minimizing mechanical complications rates but had not been assessed for this construct in a biomechanical study.MethodsSix fresh-frozen human cadaver C6-T7 thoracic spines were studied. The first thoracic vertebrae were instrumented using two claws of supra-laminar and pedicle hooks, each fixed on two adjacent vertebrae, on either side of a single free vertebra. Quasi-static pure-moment loads up to 5 Nm were applied to each specimen before and after instrumentation, in flexion-extension, right and left bending, and axial rotation. Five steel beads impacted in each vertebra allowed 3D tracking of vertebral movements on EOS biplanar radiographs acquired after each loading step. The relative ranges of motion (ROMs) of each pair of vertebras were computed.ResultsMean ROMs with the intact specimens were 17 degrees in flexion-extension, 27.9 degrees in lateral bending, and 29.5 degrees in axial rotation. Corresponding values with the instrumented specimens were 0.9 degrees, 2.6 degrees, and 7.3 degrees, respectively. Instrumentation significantly (P<0.05) decreased flexion-extension (by 92-98%), lateral bending (by 87-96%), and axial rotation (by 68-84%).ConclusionThis study establishes the biomechanical stability of a double claw-hook construct in the upper thoracic spine, which may well explain the low mechanical complication rate in previous clinical studies.Level of evidenceNot applicable, experimental cadaver study.
引用
收藏
页码:33 / 38
页数:6
相关论文
共 50 条
  • [1] Biomechanical cadaver study of proximal fixation in a minimally invasive bipolar construct
    M. Gaume
    S. Persohn
    C. Vergari
    C. Glorion
    W. Skalli
    L. Miladi
    Spine Deformity, 2020, 8 : 33 - 38
  • [2] Minimally Invasive Intramedullary Screw Versus Plate Fixation for Proximal Phalanx Fractures: A Biomechanical Study
    Miles, Megan R.
    Krul, Kevin P.
    Abbasi, Pooyan
    Thakkar, Madhuli Y.
    Giladi, Aviram M.
    Means, Kenneth R., Jr.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2021, 46 (06): : 518.e1 - 518.e8
  • [3] Stability of Fixation of Proximal Phalanx Unicondylar Fractures of the Hand: A Biomechanical Cadaver Study
    Sirota, Michael A.
    Parks, Brent G.
    Higgins, James P.
    Means, Kenneth R., Jr.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (01): : 77 - 81
  • [4] Does minimally invasive external fixation of proximal humerus fractures provide adequate stability? A biomechanical in vitro study
    Greinwald, Markus
    Matschl, Franz
    Regenbogen, Stephan
    Paetzold, Robert
    Blonna, Davide
    Augat, Peter
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
  • [5] Biomechanical Assessment of Minimally Invasive Decompression for Lumbar Spinal Canal Stenosis A Cadaver Study
    Hamasaki, Takahiko
    Tanaka, Nobuhiro
    Kim, JinHwan
    Okada, Motohiro
    Ochi, Mitsuo
    Hutton, William C.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2009, 22 (07): : 486 - 491
  • [6] Anatomy Notes on Minimally Invasive Plate Osteosynthesis of the Proximal Humerus. A Cadaver Study
    Krivohlavek, M.
    Taller, S.
    Lukas, R.
    Drac, P.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2014, 81 (01) : 63 - 69
  • [7] A biomechanical study comparing minimally invasive anterior pelvic ring fixation techniques to external fixation
    MacCormicka, Lauren M.
    Chen, Frank
    Gilbertson, Jeff
    Khana, Sikandar
    Schroder, Lisa K.
    Bechtolda, Joan E.
    Cole, Peter A.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2019, 50 (02): : 251 - 255
  • [8] Biomechanical study on surgical fixation methods for minimally invasive treatment of hallux valgus
    Mao, Rui
    Guo, Junchao
    Luo, Chenyu
    Fan, Yubo
    Wen, Jianmin
    Wang, Lizhen
    MEDICAL ENGINEERING & PHYSICS, 2017, 46 : 21 - 26
  • [9] Minimally Invasive Polyaxial Locking Plate Fixation of Proximal Humeral Fractures: A Prospective Study
    Ruchholtz, Steffen
    Hauk, Carsten
    Lewan, Ulrike
    Franz, Daniel
    Kuehne, Christian
    Zettl, Ralph
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 71 (06): : 1737 - 1744
  • [10] Open versus minimally invasive fixation of a simulated syndesmotic injury in a cadaver model
    Adam C. Shaner
    Norachart Sirisreetreerux
    Babar Shafiq
    Lynne C. Jones
    Erik A. Hasenboehler
    Journal of Orthopaedic Surgery and Research, 12