A biomechanical study comparing minimally invasive anterior pelvic ring fixation techniques to external fixation

被引:14
|
作者
MacCormicka, Lauren M. [1 ]
Chen, Frank [2 ]
Gilbertson, Jeff [3 ]
Khana, Sikandar [1 ]
Schroder, Lisa K. [3 ]
Bechtolda, Joan E. [1 ,2 ]
Cole, Peter A. [1 ,3 ]
机构
[1] Univ Minnesota, Dept Orthopaed Surg, 2512 S 7th St,R200, Minneapolis, MN 55454 USA
[2] Excelen Ctr Bone & Joint Res & Educ, 700 10th Ave South, Minneapolis, MN 55415 USA
[3] Reg Hosp, Dept Orthopaed Surg, 640 Jackson St MS 11503L, St Paul, MN 55101 USA
关键词
Anterior pelvic ring fixation; Subcutaneous; Minimally invasive; INFIX; Pelvic Bridge; SUBCUTANEOUS INTERNAL FIXATOR; COMPLICATIONS; INJURIES; REDUCTION; FRACTURES; LOADS;
D O I
10.1016/j.injury.2018.07.011
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: INFIX and Pelvic Bridge are two new minimally invasive surgical techniques for unstable pelvic ring injuries, and they have demonstrated early clinical success in small, single-center case-series. The primary objective of this study is to gather evidence speaking to the biomechanical stability of internal bridging methods relative to external fixation, with the expectation of biomechanical equivalence. Methods: Ten human cadaveric pelvic specimens were dissected free of all skin, fat, organs, and musculature and were prepared with a partially unstable pelvic ring injury (OTA/AO 61-B). The specimens were randomized to two groups and were repaired and tested with anterior pelvic external fixation (APEF) and INFIX sequentially, or APEF and Pelvic Bridge sequentially. Testing was performed with each specimen mounted onto a servo-hydraulic testing frame with axial compression applied to the superior base of the sacrum under five axial loading/unloading sinusoidal cycles between 10 N and 1000 N at 0.1 Hz. Relative translational motion and rotation across the osteotomy site was reported as our primary outcome measures. Outcome measures were further analyzed using a Wilcoxon signed-rank test to determine differences between non-parametric data sets with significance defined as a p value < 0.05. Results: We found no statistical difference in translation (p = 0.237, 0.228) or rotation (p = 0.278, 0.873) at the fracture site when comparing both new constructs to external fixation. Under the imposed loading protocol, no episodes of implant failure or failure at the bone-implant interface occurred. Discussion: Our study provides the biomechanical foundation necessary to support future clinical trial implementation for pelvic fracture patients. While biomechanical stability of these newer, subcutaneous techniques is equivalent to APEF, the surgeon must take into account their technical abilities and knowledge of pelvic anatomy, patient-specific factors including body habitus, and the potential complications associated with each implant and the ability to avoid them. (C) 2018 Elsevier Ltd. All rights reserved.
引用
收藏
页码:251 / 255
页数:5
相关论文
共 50 条
  • [1] Minimally-invasive fixation for anterior pelvic ring disruptions
    Cole, Peter A.
    Dyskin, Evgeny A.
    Gilbertson, Jeffrey A.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 : S27 - S34
  • [2] Minimally invasive anterior pelvic internal fixation: An anatomic study comparing Pelvic Bridge to INFIX
    Reichel, Lee M.
    MacCormick, Lauren M.
    Dugarte, Anthony J.
    Rizkala, Amir R.
    Graves, Sara C.
    Cole, Peter A.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2018, 49 (02): : 309 - 314
  • [3] Minimally invasive screw fixation of the anterior pelvic ring and the distal ilium
    Corbaz, Jocelyn
    Herteleer, Michiel
    Steinmetz, Sylvan
    Arand, Charlotte
    Nowak, Tobias
    Wagner, Daniel
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2025, 37 (01): : 23 - 33
  • [4] The Anterior Subcutaneous Pelvic Ring Fixator No Biomechanical Advantages Compared with External Fixation
    Jordan, Martin C.
    Brems, Anna-Christina
    Heintel, Timo
    Jansen, Hendrik
    Hoelscher-Doht, Stefanie
    Meffert, Rainer H.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2019, 101 (19): : 1724 - 1731
  • [5] A New Technique for Anterior Pelvic Ring Fixation Using a Minimally Invasive Approach
    Qoreishy, Mohamad
    Alamian, Aliasghar
    Movahedinia, Mohammad
    Keyhani, Sohrab
    TECHNIQUES IN ORTHOPAEDICS, 2022, 37 (04) : 218 - 223
  • [6] Intramedullary Fixation Techniques for the Anterior Pelvic Ring
    Eastman, Jonathan G.
    Routt, Milton L. Chip, Jr.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 : S4 - S13
  • [7] Mechanical testing of a device for subcutaneous internal anterior pelvic ring fixation versus external pelvic ring fixation
    Georg Osterhoff
    Simon Tiziani
    Stephen J Ferguson
    Gregor Spreiter
    Max J Scheyerer
    Gian-Leza Spinas
    Guido A Wanner
    Hans-Peter Simmen
    Clément ML Werner
    BMC Musculoskeletal Disorders, 15
  • [8] Mechanical testing of a device for subcutaneous internal anterior pelvic ring fixation versus external pelvic ring fixation
    Osterhoff, Georg
    Tiziani, Simon
    Ferguson, Stephen J.
    Spreiter, Gregor
    Scheyerer, Max J.
    Spinas, Gian-Leza
    Wanner, Guido A.
    Simmen, Hans-Peter
    Werner, Clement M. L.
    BMC MUSCULOSKELETAL DISORDERS, 2014, 15
  • [9] Minimally Invasive Fixation of the Pelvic Ring with a Transiliacal Internal Fixator
    Salasek, M.
    Pavelka, T.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2012, 79 (04) : 335 - 340
  • [10] A prospective case series for a minimally invasive internal fixation device for anterior pelvic ring fractures
    Hoskins, Wayne
    Bucknill, Andrew
    Wong, James
    Britton, Edward
    Judson, Rodney
    Gumm, Kellie
    Santos, Roselyn
    Sheehy, Rohan
    Griffin, Xavier
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2016, 11