Does minimally invasive external fixation of proximal humerus fractures provide adequate stability? A biomechanical in vitro study

被引:0
|
作者
Greinwald, Markus [1 ]
Matschl, Franz [1 ]
Regenbogen, Stephan [2 ,3 ]
Paetzold, Robert [2 ]
Blonna, Davide [4 ]
Augat, Peter [1 ,5 ]
机构
[1] BG Unfallklin Murnau, Inst Biomech, Prof Kuntscher Str 8, D-82418 Murnau, Germany
[2] BG Unfallklin Murnau, Dept Trauma & Orthopaed Surg, Murnau, Germany
[3] Heidelberg Univ, BG Klin Ludwigshafen, Dept Traumatol, Ludwigshafen, Germany
[4] Univ Torino, AO Ordine Mauriziano Hosp, Dept Orthopaed & Traumatol, Turin, Italy
[5] Paracelsus Med Univ, Inst Biomech, Salzburg, Austria
关键词
Proximal humeral fracture; External fixator; Internal fixator; Biomechanics; Shoulder fractures; In vitro; Semi-rigid; Galaxy fixator; Philos plate; PERCUTANEOUS FIXATION; WIRE MIGRATION; REDUCTION;
D O I
10.1186/s12891-024-07977-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Proximal humerus fractures are usually treated with rigid angle-stable plates or nails. As an alternative, semi-rigid, minimally invasive systems have been developed for supportive fixation. A new system uses this principle with 2.5 mm partially threaded pins and external fixation. Aim This study aimed to compare the fracture stability of the fixator with an angular stable plate about whether the mechanical stability is sufficient for early postoperative mobilisation. Materials and methods Human proximal humeri (paired, female, age >= 60, n = 8) were cut to a 3-part fracture and treated with an external fixator (Galaxy Shoulder, Orthofix) or an angular stable plate (Philos, DePuy Synthes). A physiological load was simulated in vitro based on a 45 degrees abduction motion. The load level was continuously increased from 10 to 320 N within 10,000 cycles simulating the loads during rehabilitation. This was followed by another 10,000 cycles at 320 N for fatigue testing. Fracture motion was recorded with a 3D camera. Fracture stability was evaluated at a partial load of 160 N and 240 N (simulating loads during early mobilisation). Results The fracture motion at the lateral aspect was significantly (p < 0.034) larger for the fixator compared to the plate both for axial and for shear movements, while at the medial aspects fracture motions were similar. Both fixation systems adequately stabilized the lateral tubercle fragment with no significant differences between fixation techniques. The migration of the fixator pins (1.1 +/- 1.0 mm) and valgus rotation of the head fragment were minimal (3.1 +/- 2.0 degrees). The maximum load level of 320 N was reached by all plate constructs and 3 fixator constructs. Conclusion Our results suggest that the external fixator system allows early but gentle postoperative mobilisation. More aggressive rehabilitation measures should be postponed after the healing of the fracture.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Treatment of proximal humerus fractures with external fixation
    Copuroglu, Cem
    Gurbuz, Hakan
    Eskin, Deniz
    TRAKYA UNIVERSITESI TIP FAKULTESI DERGISI, 2008, 25 (01): : 1 - 5
  • [2] Locking Plates With Computationally Enhanced Screw Trajectories Provide Superior Biomechanical Fixation Stability of Complex Proximal Humerus Fractures
    Mischler, Dominic
    Schader, Jana Felicitas
    Dauwe, Jan
    Tenisch, Lara
    Gueorguiev, Boyko
    Windolf, Markus
    Varga, Peter
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [3] Minimally Invasive Intramedullary Screw Versus Plate Fixation for Proximal Phalanx Fractures: A Biomechanical Study
    Miles, Megan R.
    Krul, Kevin P.
    Abbasi, Pooyan
    Thakkar, Madhuli Y.
    Giladi, Aviram M.
    Means, Kenneth R., Jr.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2021, 46 (06): : 518.e1 - 518.e8
  • [4] Closed Reduction and Minimally Invasive Percutaneous Fixation of Proximal Humerus Fractures Using the Humerusblock
    Brunner, Alexander
    Weller, Knut
    Thormann, Sebastian
    Joeckel, Jens-Arne
    Babst, Reto
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2010, 24 (07) : 407 - 413
  • [5] Percutaneous and minimally invasive treatment of proximal humerus fractures
    Wong, KL
    Williams, GR
    SPORTS MEDICINE AND ARTHROSCOPY REVIEW, 1999, 7 (02): : 134 - 144
  • [6] Minimal invasive flexible fixation for fractures of the proximal humerus
    Narvani, Ali
    Levy, Ofer
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2010, 92 (08) : 635 - 638
  • [7] Biomechanical cadaver study of proximal fixation in a minimally invasive bipolar construct
    M. Gaume
    S. Persohn
    C. Vergari
    C. Glorion
    W. Skalli
    L. Miladi
    Spine Deformity, 2020, 8 : 33 - 38
  • [8] Biomechanical cadaver study of proximal fixation in a minimally invasive bipolar construct
    Gaume, M.
    Persohn, S.
    Vergari, C.
    Glorion, C.
    Skalli, W.
    Miladi, L.
    SPINE DEFORMITY, 2020, 8 (01) : 33 - 38
  • [9] Plate fixation of proximal humerus fractures using the minimally invasive anterolateral delta split approach
    Acklin, Y. P.
    Sommer, C.
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2012, 24 (01): : 61 - 73
  • [10] Minimally invasive stabilization of distal humerus fractures: A pilot study with biomechanical evaluation
    Reising, K.
    Konstantinidis, L.
    Helwig, P.
    Wagner, F. C. L.
    Suedkamp, N. P.
    Strohm, P. C.
    TECHNOLOGY AND HEALTH CARE, 2014, 22 (06) : 909 - 913