Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria

被引:88
|
作者
Lenters-Westra, Erna [1 ,2 ]
Slingerland, Robbert J. [1 ,2 ]
机构
[1] Isala, Dept Clin Chem, NL-8025 AB Zwolle, Netherlands
[2] European Reference Lab Glycohemoglobin, Zwolle, Netherlands
关键词
GLYCOHEMOGLOBIN STANDARDIZATION PROGRAM; BIOLOGICAL VARIATION;
D O I
10.1373/clinchem.2014.224311
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS: Clinical and Laboratory Standards Institute protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using the mean of 3 certified secondary reference measurement procedures (SRMPs). Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria was also evaluated. Interference of common Hb variants was investigated for methods that could work with hemolysed material. RESULTS: The total CVs for all instruments, except for the DCA Vantage at a high Hb A(1c) value, were <= 3.1% in SI units and <= 2.1% in Diabetes Control and Complications Trial (DCCT) units. Afinion, DCA Vantage, B-analyst, and Cobas B101 instruments passed the NGSP criteria with 2 different reagent lot numbers. Quo-Test, Quo-Lab, and InnovaStar instruments had a negative bias compared to the mean of the 3 SRMPs and failed NGSP criteria. Most of the common Hb variants did not interfere with the investigated instruments, except Hb AE for the Cobas B101. CONCLUSIONS: Afinion, DCA Vantage, Cobas B101, and B-analyst instruments met the generally accepted performance criteria for Hb A(1c). Quo-Test, Quo-Lab, and InnovaStar met the criteria for precision but not for bias. Proficiency testing should be mandated for users of Hb A(1c) POC assays to ensure quality. (C) 2014 American Association for Clinical Chemistry
引用
收藏
页码:1062 / 1072
页数:11
相关论文
共 50 条
  • [31] One in Five Laboratories Using Various Hemoglobin A1c Methods Do Not Meet the Criteria for Optimal Diabetes Care Management
    Lenters-Westra, Erna
    Weykamp, Cas
    Schindhelm, Roger K.
    Siebelder, Carla
    Bilo, Henk J.
    Slingerland, Robbert J.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2011, 13 (04) : 429 - 433
  • [32] Evaluation of Afinion AS100 hemoglobin A1c point-of-care testing analyzer
    Hui, S. R.
    Charnot-Katsikas, A.
    Yeo, K. T.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (03) : 490 - 491
  • [33] Comparability of hemoglobin A1c level measured in capillary versus venous blood sample applying two point-of-care instruments
    Keramati T.
    Razi F.
    Tootee A.
    Larijani B.
    Journal of Diabetes & Metabolic Disorders, 13 (1):
  • [34] Assessment of Two Diabetes Point-of-care Analyzers Measuring Hemoglobin A1c in the Peruvian Amazon
    Saxton, Anthony T.
    Jaime Miranda, J.
    Ortiz, Ernesto J.
    Pan, William
    ANNALS OF GLOBAL HEALTH, 2018, 84 (04): : 618 - 624
  • [35] Evaluation of the Quo-Test Hemoglobin A1c Point-of-Care Instrument: Second Chance
    Lenters-Westra, Erna
    Slingerland, Robbert J.
    CLINICAL CHEMISTRY, 2010, 56 (07) : 1191 - 1193
  • [36] Analytical performance evaluation of the GreenCare A1c and Cera-Stat HbA1c point-of-care testing assays
    Lee, Joon Hee
    Jun, Sun-Hee
    Lee, Jikyo
    Song, Sang Hoon
    Lee, Kyunghoon
    ANNALS OF CLINICAL BIOCHEMISTRY, 2025, 62 (01) : 46 - 56
  • [37] Analytical evaluation and human factors assessment of two point of care hemoglobin A1c devices
    Thayer, Johanna
    Carson, Kes
    Fuezery, Anna
    Pouliot, Rhonda
    Tessier, Rachelle
    VanSpronsen, Amanda
    Parrish, Megan
    CLINICAL BIOCHEMISTRY, 2019, 74 : 89 - 90
  • [38] Hemoglobin A1c Measurement Using Point of Care Testing
    Erdal, Gulcin Sahingoz
    Isiksacan, Nilgun
    Koser, Murat
    Kocamaz, Nursel
    ISTANBUL MEDICAL JOURNAL, 2020, 21 (01): : 37 - 41
  • [39] Assessment of Patient Satisfaction with On-Site Point-of-Care Hemoglobin A1c Testing: An Observational Study
    Ayman A. Al Hayek
    Abdulghani H. Al-Saeed
    Wael M. Alzahrani
    Mohamed A. Al Dawish
    Diabetes Therapy, 2021, 12 : 2531 - 2544
  • [40] Point-of-Care Hemoglobin A1C Testing System in Community Settings: Challenges, Opportunities, and Measurement Characteristics
    Smith, Morgan
    Rosenmoss, Sophie
    Seligman, Hilary K.
    PROGRESS IN COMMUNITY HEALTH PARTNERSHIPS-RESEARCH EDUCATION AND ACTION, 2020, 14 (03) : 327 - 335