Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria

被引:88
|
作者
Lenters-Westra, Erna [1 ,2 ]
Slingerland, Robbert J. [1 ,2 ]
机构
[1] Isala, Dept Clin Chem, NL-8025 AB Zwolle, Netherlands
[2] European Reference Lab Glycohemoglobin, Zwolle, Netherlands
关键词
GLYCOHEMOGLOBIN STANDARDIZATION PROGRAM; BIOLOGICAL VARIATION;
D O I
10.1373/clinchem.2014.224311
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS: Clinical and Laboratory Standards Institute protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using the mean of 3 certified secondary reference measurement procedures (SRMPs). Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria was also evaluated. Interference of common Hb variants was investigated for methods that could work with hemolysed material. RESULTS: The total CVs for all instruments, except for the DCA Vantage at a high Hb A(1c) value, were <= 3.1% in SI units and <= 2.1% in Diabetes Control and Complications Trial (DCCT) units. Afinion, DCA Vantage, B-analyst, and Cobas B101 instruments passed the NGSP criteria with 2 different reagent lot numbers. Quo-Test, Quo-Lab, and InnovaStar instruments had a negative bias compared to the mean of the 3 SRMPs and failed NGSP criteria. Most of the common Hb variants did not interfere with the investigated instruments, except Hb AE for the Cobas B101. CONCLUSIONS: Afinion, DCA Vantage, Cobas B101, and B-analyst instruments met the generally accepted performance criteria for Hb A(1c). Quo-Test, Quo-Lab, and InnovaStar met the criteria for precision but not for bias. Proficiency testing should be mandated for users of Hb A(1c) POC assays to ensure quality. (C) 2014 American Association for Clinical Chemistry
引用
收藏
页码:1062 / 1072
页数:11
相关论文
共 50 条
  • [21] Evaluation of the Siemens DCA Vantage hemoglobin A1c point-of-care analyzer
    Charnot-Katsikos, A.
    Hui, S. R.
    Yeo, K. T.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (03) : 481 - 481
  • [22] Are hemoglobin A1c point-of-care analyzers fit for purpose? The story continues
    Lenters-Westra, Erna
    English, Emma
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2021, 59 (04) : 765 - 774
  • [23] Evaluation of the TRIstat point-of-care hemoglobin A1c device versus HPLC
    Shea, E. N.
    Wu, A.
    Wong, A.
    Munoz, V.
    Smith, A. C.
    Midyett, K.
    Garg, U.
    Reddig, N.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A196 - A196
  • [24] Implementing point-of-care hemoglobin A1C testing in an obstetrics outpatient clinic
    Weli, Homayemem
    Farnsworth, Christopher W.
    LABORATORY MEDICINE, 2024, 55 (04) : 464 - 470
  • [25] Point-of-Care Hemoglobin A1c Testing in a Community Diabetes Prevention Program
    Mcdonnell, Marie E.
    Jahreis, Katie
    Khan, Waheed
    Bourland, Ashley C.
    Cullum-Dugan, Diana
    Hanrahan, Patricia
    Apovian, Caroline M.
    DIABETES, 2012, 61 : A308 - A309
  • [26] Retrospective Accuracy Study of Point-of-Care Hemoglobin A1c in Field Conditions
    Manthei, David M.
    Wesener, Nicholas
    Twarkowski, Denise
    Giacherio, Donald A.
    Schroeder, Lee F.
    CLINICAL CHEMISTRY, 2017, 63 (03) : 780 - 781
  • [27] Discordant Point-of-Care and Laboratory Hemoglobin A1c Concentrations in Ambulatory Settings
    Korpi-Steiner, Nichole
    Cotten, Steven W.
    Little, Randie R.
    Kirk, Deepa
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2024, 10 (02): : 519 - 522
  • [28] Analytical performance evaluation of three methods for hemoglobin A1c quantification
    Unger, Gisela
    Ruiz, Gustavo
    Milano, Pablo
    Fabiana Benozzi, Silvia
    Laura Pennacchiotti, Graciela
    ACTA BIOQUIMICA CLINICA LATINOAMERICANA, 2014, 48 (02): : 183 - 189
  • [29] Hemoglobin A1C determination by point-of-care testing: Its correlation to standard method
    Wiwanitkit, Viroj
    DIABETES & METABOLIC SYNDROME-CLINICAL RESEARCH & REVIEWS, 2012, 6 (02) : 110 - 111
  • [30] Multicenter assessment of a hemoglobin A1c point-of-care device for diagnosis of diabetes mellitus
    Sobolesky, Philip M.
    Smith, Breland E.
    Saenger, Amy K.
    Schulz, Karen
    Apple, Fred S.
    Scott, Mitchell G.
    Wu, Alan H. B.
    Little, Randie R.
    Fitzgerald, Robert L.
    CLINICAL BIOCHEMISTRY, 2018, 61 : 18 - 22