Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria

被引:88
|
作者
Lenters-Westra, Erna [1 ,2 ]
Slingerland, Robbert J. [1 ,2 ]
机构
[1] Isala, Dept Clin Chem, NL-8025 AB Zwolle, Netherlands
[2] European Reference Lab Glycohemoglobin, Zwolle, Netherlands
关键词
GLYCOHEMOGLOBIN STANDARDIZATION PROGRAM; BIOLOGICAL VARIATION;
D O I
10.1373/clinchem.2014.224311
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS: Clinical and Laboratory Standards Institute protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using the mean of 3 certified secondary reference measurement procedures (SRMPs). Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria was also evaluated. Interference of common Hb variants was investigated for methods that could work with hemolysed material. RESULTS: The total CVs for all instruments, except for the DCA Vantage at a high Hb A(1c) value, were <= 3.1% in SI units and <= 2.1% in Diabetes Control and Complications Trial (DCCT) units. Afinion, DCA Vantage, B-analyst, and Cobas B101 instruments passed the NGSP criteria with 2 different reagent lot numbers. Quo-Test, Quo-Lab, and InnovaStar instruments had a negative bias compared to the mean of the 3 SRMPs and failed NGSP criteria. Most of the common Hb variants did not interfere with the investigated instruments, except Hb AE for the Cobas B101. CONCLUSIONS: Afinion, DCA Vantage, Cobas B101, and B-analyst instruments met the generally accepted performance criteria for Hb A(1c). Quo-Test, Quo-Lab, and InnovaStar met the criteria for precision but not for bias. Proficiency testing should be mandated for users of Hb A(1c) POC assays to ensure quality. (C) 2014 American Association for Clinical Chemistry
引用
收藏
页码:1062 / 1072
页数:11
相关论文
共 50 条
  • [1] Six of Eight Hemoglobin A1c Point-of-Care Instruments Do Not Meet the General Accepted Analytical Performance Criteria
    Lenters-Westra, Erna
    Slingerland, Robbert J.
    CLINICAL CHEMISTRY, 2010, 56 (01) : 44 - 52
  • [2] Evaluation of three hemoglobin A1c point-of-care instruments
    Criel, M.
    Jonckheere, S.
    Langlois, M.
    ACTA CLINICA BELGICA, 2015, 70 : S3 - S3
  • [3] Evaluation of Three Hemoglobin A1c Point-of-Care Instruments
    Criel, Marieke
    Jonckheere, Stijn
    Langlois, Michel
    CLINICAL LABORATORY, 2016, 62 (03) : 285 - 291
  • [4] Point-of-Care Hemoglobin A1c
    O'Brien, Matthew J.
    Sacks, David B.
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2019, 322 (14): : 1404 - 1405
  • [5] Assessing the Performance of Point-of-Care Hemoglobin A1c Systems
    Irvin, Ben
    Knaebel, Jennifer
    Simmons, David
    CLINICAL CHEMISTRY, 2010, 56 (08) : 1359 - 1360
  • [6] Point-of-Care Assays for Hemoglobin A1c: Is Performance Adequate?
    Little, Randie R.
    Lenters-Westra, Erna
    Rohlfing, Curt L.
    Slingerland, Robbert
    CLINICAL CHEMISTRY, 2011, 57 (09) : 1333 - 1334
  • [7] Diagnosing Diabetes Mellitus: Performance of Hemoglobin A1c Point-of-Care Instruments in General Practice Offices
    Solvik, Una Orvim
    Roraas, Thomas
    Christensen, Nina Gade
    Sandberg, Sverre
    CLINICAL CHEMISTRY, 2013, 59 (12) : 1790 - 1801
  • [8] Assessing the Performance of Point-of-Care Hemoglobin A1c Systems Reply
    Lenters-Westra, Erna
    Slingerland, Robbert J.
    CLINICAL CHEMISTRY, 2010, 56 (08) : 1360 - 1361
  • [9] Few Point-of-Care Hemoglobin A1c Assay Methods Meet Clinical Needs
    Bruns, David E.
    Boyd, James C.
    CLINICAL CHEMISTRY, 2010, 56 (01) : 4 - 6
  • [10] Point-of-care hemoglobin A1c testing in postmortem examination
    Joo-Young Na
    Forensic Science, Medicine and Pathology, 2018, 14 : 322 - 326