摘要
Purpose: To develop and validate a real-time mitral valve annulus (MVA) tracking approach based on biplane transesophageal echocardiogram (TEE) data and magnetic tracking systems (MTS) to be used in minimally invasive off-pump beating heart mitral valve repair (MVR). Methods: The authors' guidance system consists of three major components: TEE, magnetic tracking system, and an image guidance software platform. TEE provides real-time intraoperative images to show the cardiac motion and intracardiac surgical tools. The magnetic tracking system tracks the TEE probe and the surgical tools. The software platform integrates the TEE image planes and the virtual model of the tools and the MVA model on the screen. The authors' MVA tracking approach, which aims to update the MVA model in near real-time, comprises of three steps: image based gating, predictive reinitialization, and registration based MVA tracking. The image based gating step uses a small patch centered at each MVA point in the TEE images to identify images at optimal cardiac phases for updating the position of the MVA. The predictive reinitialization step uses the position and orientation of the TEE probe provided by the magnetic tracking system to predict the position of the MVA points in the TEE images and uses them for the initialization of the registration component. The registration based MVA tracking step aims to locate the MVA points in the images selected by the image based gating component by performing image based registration. Results: The validation of the MVA tracking approach was performed in a phantom study and a retrospective study on porcine data. In the phantom study, controlled translations were applied to the phantom and the tracked MVA was compared to its "true" position estimated based on a magnetic sensor attached to the phantom. The MVA tracking accuracy was 1.29 +/- 0.58 mm when the translation distance is about 1 cm, and increased to 2.85 +/- 1.19 mm when the translation distance is about 3 cm. In the study on porcine data, the authors compared the tracked MVA to a manually segmented MVA. The overall accuracy is 2.37 +/- 1.67 mm for single plane images and 2.35 +/- 1.55 mm for biplane images. The interoperator variation in manual segmentation was 2.32 +/- 1.24 mm for single plane images and 1.73 +/- 1.18 mm for biplane images. The computational efficiency of the algorithm on a desktop computer with an Intel (R) Xeon (R) CPU @3.47 GHz and an NVIDIA GeForce 690 graphic card is such that the time required for registering four MVA points was about 60 ms. Conclusions: The authors developed a rapid MVA tracking algorithm for use in the guidance of off-pump beating heart transapical mitral valve repair. This approach uses 2D biplane TEE images and was tested on a dynamic heart phantom and interventional porcine image data. Results regarding the accuracy and efficiency of the authors' MVA tracking algorithm are promising, and fulfill the requirements for surgical navigation. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:456 / 468
页数:13
相关论文
共 50 条
  • [31] NEONAV: AUGMENTED REALITY ECHOCARDIOGRAPHIC NAVIGATION AND GUIDANCE FOR BEATING HEART MITRAL VALVE REPAIR
    Currie, M. E.
    Chu, M. W.
    Moore, J. T.
    Bainbridge, D.
    McCarty, D.
    Guiraudon, G.
    Wedlake, C.
    Daly, R.
    Patel, R.
    Peters, T. M.
    Kiaii, B.
    CANADIAN JOURNAL OF CARDIOLOGY, 2013, 29 (10) : S366 - S366
  • [32] Augmented Reality Image Guidance Improves Navigation for Beating Heart Mitral Valve Repair
    Chu, Michael W. A.
    Moore, John
    Peters, Terry
    Bainbridge, Daniel
    McCarty, David
    Guiraudon, Gerard M.
    Wedlake, Chris
    Lang, Pencilla
    Rajchl, Martin
    Currie, Maria E.
    Daly, Richard C.
    Kiaii, Bob
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2012, 7 (04) : 274 - 281
  • [33] Mitral valve repair with transapical, off-pump neochord implantation. Initial experience
    Perez-Camargo, Daniel
    Cobiella-Carnicer, Javier
    Mahia-Casado, Patricia
    Carnero-Alcazar, Manuel
    Montero-Cruces, Lourdes
    Campelos-Fernandez, Paula
    Nombela-Franco, Luis
    Castellanos, Luis C. Maroto
    CIRUGIA CARDIOVASCULAR, 2023, 30 : S21 - S27
  • [34] Safety, Feasibility And Efficacy Of Transapical Off-Pump Mitral Valve Repair With Neochord Implantation
    Colli, Andrea
    Manzan, Erica
    Zucchetta, Fabio
    Sarais, Cristiano
    Pittarello, Demetrio
    Gerosa, Gino
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 64 (11) : B234 - B235
  • [35] An early European experience with transapical off-pump mitral valve repair with NeoChord implantation
    Colli, Andrea
    Manzan, Erica
    Aidietis, Audrius
    Rucinskas, Kestutis
    Bizzotto, Eleonora
    Besola, Laura
    Pradegan, Nicola
    Pittarello, Demetrio
    Janusauskas, Vilius
    Zakarkaite, Diana
    Drasutiene, Agne
    Lipnevicius, Arturas
    Danner, Bernhard C.
    Sievert, Horst
    Vaskelyte, Laura
    Schnelle, Nalan
    Salizzoni, Stefano
    Marro, Massimo
    Rinaldi, Mauro
    Kurnicka, Katarzyna
    Wrobel, Kristof
    Ceffarelli, Mariano
    Savini, Carlo
    Pacini, Davide
    Gerosa, Gino
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2018, 54 (03) : 460 - 466
  • [36] Transapical off-pump mitral valve repair with Neochord implantation: Early clinical results
    Colli, Andrea
    Manzan, Erica
    Zucchetta, Fabio
    Bizzotto, Eleonora
    Besola, Laura
    Bagozzi, Lorenzo
    Bellu, Roberto
    Sarais, Cristiano
    Pittarello, Demetrio
    Gerosa, Gino
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 204 : 23 - 28
  • [37] Multicenter European Experience of the Echo Guided Off-Pump Transapical Mitral Valve Repair
    Colli, Andrea
    Manzan, Erica
    Besola, Laura
    Bizzotto, Eleonora
    Zucchetta, Fabio
    Pittarello, Demetrio
    Salizzoni, Stefano
    Rinaldi, Mauro
    Wrbel, Krzysztof
    Kurnicka, Katarzyna
    Aidietis, Audrius
    Rucinskas, Kestutis
    Janusauskas, Vilius
    Zakarkaite, Diana
    Drasutiene, Agne
    Lipnevicius, Arturas
    Pacini, Davide
    Savini, Carlo
    Sievert, Horst
    deBruijn, Sabine
    Danner, Bernd
    Obadia, Jean-Francois I.
    Gerosa, Gino
    CIRCULATION, 2016, 134
  • [38] Beating Heart Mitral Valve Surgery: How to Test Mitral Valve Competence After Repair?
    Cikirikcioglu, Mustafa
    Myers, Patrick O.
    Pektok, Erman
    Kalangos, Afksendiyos
    JOURNAL OF CARDIAC SURGERY, 2009, 24 (02) : 223 - 224
  • [39] Importance of stabilization of the mitral annulus in mitral valve repair
    Szentkiralyi, Istvan
    Peterffy, Arpad
    Galajda, Zoltan
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2008, 136 (04): : 1102 - 1103
  • [40] Minimally Invasive, Robotic, and Off-Pump Mitral Valve Surgery
    Woo, Y. Joseph
    Rodriguez, Evelio
    Atluri, Pavan
    Chitwood, W. Randolph, Jr.
    SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2006, 18 (02) : 139 - 147