High-throughput flow cytometry: Validation in microvolume bioassays

被引:44
|
作者
Ramirez, S [1 ]
Aiken, CT [1 ]
Andrzejewski, B [1 ]
Sklar, LA [1 ]
Edwards, BS [1 ]
机构
[1] Univ New Mexico, Hlth Sci Ctr, Dept Pathol, Canc Res & Treatment Ctr, Albuquerque, NM 87131 USA
关键词
flow cytometry; automation; drug discovery; biomolecular screening; SELECTIN GLYCOPROTEIN LIGAND-1; P-SELECTIN; DRUG DISCOVERY;
D O I
10.1002/cyto.a.10035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: We recently reported an automated sample handling system, designated HyperCyt, by which samples are aspirated from microplate wells and delivered to the flow cytometer for analysis at rates approaching 100 samples per minute. In this approach, an autosampler and peristaltic pump introduce samples into a tubing line that directly connects to the flow cytometer. Air bubbles are inserted between samples to prevent sample dispersion. In the present work, we compare results of HyperCyt with those of conventional manual flow cytometric analysis in representative flow cytometric bioassays and describe a cell suspension method in which HyperCyt exploits the use of microvolume wells. Methods: Human eosinophils and neutrophils were treated with trypsin to generate a wide (>25-fold) range of membrane P-selectin glycoprotein ligand-1 (PSGL-1) expression and then stained with fluorescent anti-PSGL-1 antibodies. Human peripheral blood mononuclear cells were stained with fluorescein isothiocyanate- and phycoerythrin-conjugated monoclonal antibodies for multiparameter immunophenotype analysis. U937 cells labeled with PKH62GL were used to assess cell settling in microplate wells. Results: Differences in PSGL-1 expression levels were detected by HyperCyt autosampling of leukocytes from 96-well plates at an analysis rate of approximately 1.5 s/well. Hyper-Cyt measurements linearly correlated with parallel manual measurements (r(2) = 0.98). Lymphocyte subpopulations were accurately distinguished and reproducibly quantified in multiparameter immunophenotyping assays performed over a range of HyperCyt analysis rates (1.4-5.5 s/sample). When assay volumes were reduced to 10 mul/well in 60-well Terasaki plates, cells could be maintained in uniform suspension for up to 30 min by periodically inverting plates on a rotating carousel before HyperCyt analysis. HyperCyt analysis of five fluorescence-level Cyto-Plex beads sampled from Temsaki plate microwells at 2.5 s/well produced highly reproducible results over a wide range of input bead concentrations (from 7 X 10(5) to 20 X 10(6) beads/ml) that linearly correlated with manual analysis results. Conclusions: The HyperCyt autosampling system enabled a 10-fold or greater increase in sample thoughput compared with conventional manual flow cytometric sample analysis, with comparable analysis results. Assays were performed efficiently in 10-mul volumes to enable significant reagent cost savings, use of quantity-limited reagents at otherwise prohibitive concentrations, and maintenance of uniform suspensions of cells for prolonged periods. Cytometry Part A 53A:55-65, 2003. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:55 / 65
页数:11
相关论文
共 50 条
  • [31] Cameraless high-throughput three-dimensional imaging flow cytometry
    Han, Yuanyuan
    Tang, Rui
    Gu, Yi
    Zhang, Alex Ce
    Cai, Wei
    Castor, Violet
    Cho, Sung Hwan
    Alaynick, William
    Lo, Yu-Hwa
    OPTICA, 2019, 6 (10): : 1297 - 1304
  • [32] Inertial microfluidics for sheath-less high-throughput flow cytometry
    Ali Asgar S. Bhagat
    Sathyakumar S. Kuntaegowdanahalli
    Necati Kaval
    Carl J. Seliskar
    Ian Papautsky
    Biomedical Microdevices, 2010, 12 : 187 - 195
  • [33] High-Throughput Quantitation of Intracellular Trafficking and Organelle Disruption by Flow Cytometry
    Chia, Pei Zhi Cheryl
    Ramdzan, Yasmin M.
    Houghton, Fiona J.
    Hatters, Danny M.
    Gleeson, Paul A.
    TRAFFIC, 2014, 15 (05) : 572 - 582
  • [34] WebFlow: A Software Package for High-Throughput Analysis of Flow Cytometry Data
    Hammer, Mark M.
    Kotecha, Nikesh
    Irish, Jonathan M.
    Nolan, Garry P.
    Krutzik, Peter O.
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2009, 7 (01) : 44 - 55
  • [35] Inertial microfluidics for sheath-less high-throughput flow cytometry
    Bhagat, Ali Asgar S.
    Kuntaegowdanahalli, Sathyakumar S.
    Kaval, Necati
    Seliskar, Carl J.
    Papautsky, Ian
    BIOMEDICAL MICRODEVICES, 2010, 12 (02) : 187 - 195
  • [36] A high-throughput method for detection of DNA in chloroplasts using flow cytometry
    Rowan, Beth A.
    Oldenburg, Delene J.
    Bendich, Arnold J.
    PLANT METHODS, 2007, 3 (1)
  • [37] Cell-Based Screening Using High-Throughput Flow Cytometry
    Black, Christopher B.
    Duensing, Thomas D.
    Trinkle, Linda S.
    Dunlay, R. Terry
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2011, 9 (01) : 13 - 20
  • [38] High-throughput observation and quantification of ADCC using imaging flow cytometry
    Helguera, Gustavo
    Rodriguez, Jose
    Henery, Shannon
    Catterton, Paul
    Martinez-Maza, Otoniel
    Penichet, Manuel
    CANCER RESEARCH, 2009, 69
  • [39] A high-throughput method for detection of DNA in chloroplasts using flow cytometry
    Beth A Rowan
    Delene J Oldenburg
    Arnold J Bendich
    Plant Methods, 3
  • [40] A HIGH-THROUGHPUT FLOW CYTOMETRY METHOD FOR PLOIDY DETERMINATION IN OIL PALM
    Nasution, Olviyani
    Sitorus, Andrew Christian
    Nelson, Stephen P. C.
    Forster, Brian P.
    Caligari, Peter D. S.
    JOURNAL OF OIL PALM RESEARCH, 2013, 25 (02): : 265 - 271