Development of implantable real time micro dosimeter system

被引:0
|
作者
Yamada, Shogo [1 ]
Nakata, Eiko [2 ]
Momose, Genki [3 ]
Kikuchi, Youhei [3 ]
Ogawa, Yoshihiro [1 ]
Mori, Issei [2 ]
Oishi, Mikio [2 ]
Ishii, Keizo [3 ]
机构
[1] Tohoku Univ, Grad Sch Med, Dept Radiat Oncol, Aoba Ku, 1-1 Seiryo Machi, Sendai, Miyagi 9810872, Japan
[2] Tohoku Univ, Sch Med, Dept Hlth Sci, Sendai, Miyagi, Japan
[3] Tohoku Univ, Grad Sch Engn, Dept Quantum Sci & Energy Engn, Sendai, Miyagi, Japan
关键词
radiation therapy; cancer; micro dosimeter; CdTe semiconductor; magnetic field; real time;
D O I
10.1142/9781860948800_0029
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Although progress of radiation therapy in recent years is rapid, medical accidents like the irradiation of over and under dose occur frequently in Japan. To eradicate these accidents and to increase accuracy of radiation therapy, we are developing an implantable real time micro dosimeter system. CdTe semiconductor is used as an ideal detector. And the doses of radiation are monitored by measuring the magnetic field converted from the current. We investigated the relationship between dose rate and current value from CdTe semiconductor(4mmx4mmx0.5mm). The data showed that 70 to 90 nA of electric current is proportionally generated from the CdTe semiconductor when it is irradiated by a LINAC under an air condition by the dose rate of I similar to 6Gy/min. In addition, the proportion of irradiated dose and the current was confirmed under a subaqueous condition(15cm in depth). To investigate the influence of a magnetic, field on a human body, we calculated the tolerance frequency for human body using FEM simulation method. The data showed that the sending and receiving of data as a frequency of magnetic field should be set 100MHz or less. We are now miniaturizing the dosimeter circuit up to the size of inserted level by a fine needle.
引用
收藏
页码:271 / +
页数:3
相关论文
共 50 条
  • [31] System development of real-time navigation
    Incicco, Sebastian
    De Ferraris, Facundo Sala
    Real, Nicolas
    2015 XVI WORKSHOP ON INFORMATION PROCESSING AND CONTROL (RPIC), 2015,
  • [32] Evaluation and performance characteristics of an implantable dosimeter with the TomoTherapy hi-art system
    Esquivel, C.
    Rassiah, P.
    Beyer, G.
    Childress, C.
    Papanikolaou, N.
    MEDICAL PHYSICS, 2007, 34 (06) : 2448 - 2448
  • [33] Micro-Sized Implantable Electrocardiogram (ECG) Sensor for Real-Time Monitoring of A trial Fibrillation
    Lee, J.
    Park, H.
    Son, C.
    Kim, Y.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [34] Development of a Real-Time Deformable Dosimeter That Simultaneously Measures Dose and Tracks Deformation Vector Fields
    Cloutier, E.
    Archambault, L.
    Beaulieu, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E280 - E280
  • [35] Development of a Dental Implantable Temperature Sensor for Real-Time Diagnosis of Infectious Disease
    Kim, Jeffrey J.
    Stafford, Gery R.
    Beauchamp, Carlos
    Kim, Shin Ae
    SENSORS, 2020, 20 (14) : 1 - 18
  • [36] Evaluation of Biofouling for Implantable Micro Dialysis System
    Ota, T.
    To, N.
    Kanno, Y.
    Miki, N.
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 1942 - 1945
  • [37] Characterization of Phosphorescent Strontium Aluminate as a Real-Time Dosimeter
    Drehmel, S.
    Mathew, D.
    Ehler, E.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [38] An implantable dosimeter study of patients with prostate cancer
    Beyer, G.
    Scarantino, C.
    Mann, G.
    Carrea, T.
    Anscher, M.
    Prestidge, B.
    Black, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 2151 - 2151
  • [39] Commissioning and implementation of an implantable dosimeter for radiation therapy
    Buzurovic, Ivan
    Showalter, Timothy N.
    Studenski, Matthew T.
    Den, Robert B.
    Dicker, Adam P.
    Cao, Junsheng
    Xiao, Ying
    Yu, Yan
    Harrison, Amy
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 234 - 252
  • [40] Micro core based double satellite real time simulation system
    Wang, Ji-He
    Wang, Feng
    Lan, Sheng-Chang
    Cao, Xi-Bin
    Xitong Fangzhen Xuebao / Journal of System Simulation, 2008, 20 (02): : 328 - 331