Development of implantable real time micro dosimeter system

被引:0
|
作者
Yamada, Shogo [1 ]
Nakata, Eiko [2 ]
Momose, Genki [3 ]
Kikuchi, Youhei [3 ]
Ogawa, Yoshihiro [1 ]
Mori, Issei [2 ]
Oishi, Mikio [2 ]
Ishii, Keizo [3 ]
机构
[1] Tohoku Univ, Grad Sch Med, Dept Radiat Oncol, Aoba Ku, 1-1 Seiryo Machi, Sendai, Miyagi 9810872, Japan
[2] Tohoku Univ, Sch Med, Dept Hlth Sci, Sendai, Miyagi, Japan
[3] Tohoku Univ, Grad Sch Engn, Dept Quantum Sci & Energy Engn, Sendai, Miyagi, Japan
关键词
radiation therapy; cancer; micro dosimeter; CdTe semiconductor; magnetic field; real time;
D O I
10.1142/9781860948800_0029
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Although progress of radiation therapy in recent years is rapid, medical accidents like the irradiation of over and under dose occur frequently in Japan. To eradicate these accidents and to increase accuracy of radiation therapy, we are developing an implantable real time micro dosimeter system. CdTe semiconductor is used as an ideal detector. And the doses of radiation are monitored by measuring the magnetic field converted from the current. We investigated the relationship between dose rate and current value from CdTe semiconductor(4mmx4mmx0.5mm). The data showed that 70 to 90 nA of electric current is proportionally generated from the CdTe semiconductor when it is irradiated by a LINAC under an air condition by the dose rate of I similar to 6Gy/min. In addition, the proportion of irradiated dose and the current was confirmed under a subaqueous condition(15cm in depth). To investigate the influence of a magnetic, field on a human body, we calculated the tolerance frequency for human body using FEM simulation method. The data showed that the sending and receiving of data as a frequency of magnetic field should be set 100MHz or less. We are now miniaturizing the dosimeter circuit up to the size of inserted level by a fine needle.
引用
收藏
页码:271 / +
页数:3
相关论文
共 50 条
  • [21] Real-time dosimeter targeted to nuclear applications
    Correia, Alexandre
    Rosa, Carla C.
    Santos, Pedro M. P.
    Falcao, Antonio N.
    Lorentz, Katharina
    SECOND INTERNATIONAL CONFERENCE ON APPLICATIONS OF OPTICS AND PHOTONICS, 2014, 9286
  • [23] Development of hybrid real time wide-band electronic radiation dosimeter for a mobile robot
    Lee, S. M.
    Lee, H. S.
    Lee, H. H.
    PROCEEDINGS OF THE 2006 IEEE SENSORS APPLICATIONS SYMPOSIUM, 2006, : 141 - 145
  • [24] DEVELOPMENT AND CHARACTERIZATION OF REAL-TIME WIDE-ENERGY RANGE PERSONAL NEUTRON DOSIMETER
    NAKAMURA, T
    TSUJIMURA, N
    YAMANO, T
    SUZUKI, T
    OKAMOTO, E
    JOURNAL OF THE ATOMIC ENERGY SOCIETY OF JAPAN, 1994, 36 (04): : 337 - 345
  • [25] A real time active pixel dosimeter for interventional radiology
    Paolucci, M.
    Battisti, D.
    Biasini, M.
    Checcucci, B.
    Di Lorenzo, R.
    Esposito, A.
    Fano, L.
    Passeri, D.
    Placidi, P.
    Servoli, L.
    RADIATION MEASUREMENTS, 2011, 46 (11) : 1271 - 1276
  • [26] Real-time radio-transparent dosimeter for X-ray imaging system
    Staub, D
    Cailleret, J
    Guyonnet, JL
    Le, TD
    Wurtz, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2): : 303 - 307
  • [27] Development of a system to estimate doses in real time
    de Camargo Lourenco, Jose Carlos
    Paschuk, Sergei Anatolyevich
    Schelin, Hugo Reuters
    Denyak, Valeriy
    Caggiano Santos, Cesar Antonio
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167 (167)
  • [28] Development of Real Time Virtual Aquarium System
    Lee, Hyun-Cheol
    Kim, Eun-Seok
    Joo, Nak-Keun
    Hur, Gi-Taek
    INTERNATIONAL JOURNAL OF COMPUTER SCIENCE AND NETWORK SECURITY, 2006, 6 (7A): : 58 - 63
  • [29] Predictability in real-time system development
    Huang, JF
    Voeten, J
    Florescu, O
    van der Putten, P
    Corporaal, H
    ADVANCES IN DESIGN AND SPECIFICATION LANGUAGES FOR SOCS: SELECTED CONTRIBUTIONS FROM FDL'04, 2005, : 123 - 139
  • [30] Development of a real-time monitoring system
    Marzi, H
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART B-JOURNAL OF ENGINEERING MANUFACTURE, 2002, 216 (06) : 933 - 937