An analysis of the accuracy of the Cyberknife: A robotic nameless stereotactic radiosurgical system

被引:257
|
作者
Chang, SD
Main, W
Martin, DP
Gibbs, IC
Heilbrun, MP
机构
[1] Stanford Univ, Sch Med, Dept Neurosurg, Stanford, CA 94305 USA
[2] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
[3] Accuray Inc, Sunnyvale, CA USA
关键词
accuracy; radiosurgery; stereotactic;
D O I
10.1097/00006123-200301000-00018
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: The use of stereotactic radiosurgical systems to treat intracranial and extracranial tumors and other lesions requires a high degree of accuracy in target identification and localization. The purpose of this study was to evaluate the total system accuracy of the CyberKnife (Accuray, Inc., Sunnyvale, CA), a frameless, image-guided, stereotactic radiosurgery system. METHODS: Clinically relevant accuracy or application accuracy of the CyberKnife radiosurgery system is based on 1) the beam delivery accuracy, which combines the robot and the camera image tracking system, and 2) target localization accuracy, which combines computed tomographic (CT) imaging and treatment planning. Clinically relevant accuracy can be measured by delivering a radiation dose to phantoms, in which the target is defined on a set of CT images using all components of the CyberKnife system, including the treatment planning software, the robot, the camera tracking system, and the linear accelerator. Clinically relevant accuracy was measured in head phantoms loaded with packs of radiochromic film. The accuracy measured is the displacement of the dose contours from the treatment plan to that measured in the radiosurgically exposed phantom. RESULTS: Measurements of mean errors of the second-generation CyberKnife system at Stanford University Medical Center, installed in 2001, ranged from 0.7 mm for a CT slice thickness of 0.625 mm to 1.97 mm for a CT slice thickness of 3.75 mm. CONCLUSION: The frameless, image-guided, second-generation CyberKnife radiosurgery system has a clinically relevant accuracy of 1.1 +/- 0.3 mm when CT slice thicknesses of 1.25 mm are used. CyberKnife precision is comparable to published localization errors in current frame-based radiosurgical systems.
引用
收藏
页码:140 / 146
页数:7
相关论文
共 50 条
  • [31] Stereotactic Body Radiotherapy using the Cyberknife Robotic Radiosurgery System for Lung Cancer Primaries and Metastases to the Lung
    Rashtian, A.
    Zhang, Y. R.
    Lee, C.
    Oh, D. S.
    Gitlitz, B. J.
    Liu, S. V.
    Hagen, J. A.
    JOURNAL OF THORACIC ONCOLOGY, 2012, 7 (09) : S294 - S294
  • [32] Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system
    Antypas, Christos
    Pantelis, Evaggelos
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (17): : 4697 - 4718
  • [33] Effect of Cardiac Motion On the Cyberknife Synchrony Tracking System for Radiosurgical Cardiac Ablation
    Teo, B. K.
    Dieterich, S.
    Blanck, O.
    Sumanaweera, T.
    Gardner, E.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [34] Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma
    Dutta, Debnarayan
    Subramanian, S. Balaji
    Murli, V.
    Sudahar, H.
    Kurup, P. G. Gopalakrishna
    Potharaju, Mahadev
    JOURNAL OF NEURO-ONCOLOGY, 2012, 106 (03) : 637 - 642
  • [35] Dose linearity and monitor unit stability of a G4 type cyberknife robotic stereotactic radiosurgery system
    Sudahar, H.
    Kurup, P. G. G.
    Murali, V.
    Velmurugan, J.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (01) : 4 - 7
  • [36] CyberKnife® robotic radiosurgery system for tumor treatment
    Hara, Wendy
    Soltys, Scott G.
    Gibbs, Iris C.
    EXPERT REVIEW OF ANTICANCER THERAPY, 2007, 7 (11) : 1507 - 1515
  • [37] Dosimetric comparison of Linac-based (BrainLAB®) and robotic radiosurgery (CyberKnife®) stereotactic system plans for acoustic schwannoma
    Debnarayan Dutta
    S. Balaji Subramanian
    V. Murli
    H. Sudahar
    P. G. Gopalakrishna Kurup
    Mahadev Potharaju
    Journal of Neuro-Oncology, 2012, 106 : 637 - 642
  • [38] Cyberknife robotic radiotherapy system for cancer treatment
    Kim, Yeon Sil
    JOURNAL OF THE KOREAN MEDICAL ASSOCIATION, 2008, 51 (07): : 630 - 637
  • [39] Long-term clinical outcomes of stereotactic radiosurgery and hypofractionated stereotactic radiotherapy using the CyberKnife(R) robotic radiosurgery system for vestibular schwannoma
    Puataweepong, Putipun
    Dhanachai, Mantana
    Swangsilpa, Thiti
    Sitathanee, Chomporn
    Ruangkanchanasetr, Rawee
    Hansasuta, Ake
    Pairat, Kumuthinee
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2022, 18 (05) : E247 - E254
  • [40] Radiosurgery and stereotactic radiotherapy with cyberknife system for meningioma treatment
    Di Franco, Rossella
    Borzillo, Valentina
    Ravo, Vincenzo
    Falivene, Sara
    Romano, Francesco Jacopo
    Muto, Matteo
    Cammarota, Fabrizio
    Totaro, Giuseppe
    Ametrano, Gianluca
    Rossetti, Sabrina
    Cavaliere, Carla
    D' Aniello, Carmine
    Iovane, Gelsomina
    Porricelli, Maria Assunta
    Berretta, Massimiliano
    Botti, Gerardo
    Starace, Luigi
    La Salvia, Enrico
    Facchini, Gaetano
    Muto, Paolo
    NEURORADIOLOGY JOURNAL, 2018, 31 (01): : 18 - 26