An analysis of the accuracy of the Cyberknife: A robotic nameless stereotactic radiosurgical system

被引:257
|
作者
Chang, SD
Main, W
Martin, DP
Gibbs, IC
Heilbrun, MP
机构
[1] Stanford Univ, Sch Med, Dept Neurosurg, Stanford, CA 94305 USA
[2] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
[3] Accuray Inc, Sunnyvale, CA USA
关键词
accuracy; radiosurgery; stereotactic;
D O I
10.1097/00006123-200301000-00018
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: The use of stereotactic radiosurgical systems to treat intracranial and extracranial tumors and other lesions requires a high degree of accuracy in target identification and localization. The purpose of this study was to evaluate the total system accuracy of the CyberKnife (Accuray, Inc., Sunnyvale, CA), a frameless, image-guided, stereotactic radiosurgery system. METHODS: Clinically relevant accuracy or application accuracy of the CyberKnife radiosurgery system is based on 1) the beam delivery accuracy, which combines the robot and the camera image tracking system, and 2) target localization accuracy, which combines computed tomographic (CT) imaging and treatment planning. Clinically relevant accuracy can be measured by delivering a radiation dose to phantoms, in which the target is defined on a set of CT images using all components of the CyberKnife system, including the treatment planning software, the robot, the camera tracking system, and the linear accelerator. Clinically relevant accuracy was measured in head phantoms loaded with packs of radiochromic film. The accuracy measured is the displacement of the dose contours from the treatment plan to that measured in the radiosurgically exposed phantom. RESULTS: Measurements of mean errors of the second-generation CyberKnife system at Stanford University Medical Center, installed in 2001, ranged from 0.7 mm for a CT slice thickness of 0.625 mm to 1.97 mm for a CT slice thickness of 3.75 mm. CONCLUSION: The frameless, image-guided, second-generation CyberKnife radiosurgery system has a clinically relevant accuracy of 1.1 +/- 0.3 mm when CT slice thicknesses of 1.25 mm are used. CyberKnife precision is comparable to published localization errors in current frame-based radiosurgical systems.
引用
收藏
页码:140 / 146
页数:7
相关论文
共 50 条
  • [21] CyberKnife® robotic stereotactic radiotherapy:: technical aspects and medical indications
    Bondiau, P. -Y.
    Benezery, K.
    Beckendorf, V.
    Peiffert, D.
    Gerard, J. -P
    Mirabel, X.
    Noel, A.
    Marchesi, V.
    Lacornerie, T.
    Dubus, F.
    Sarrazin, T.
    Herault, J.
    Marcie, S.
    Angellier, G.
    Lartigau, E.
    CANCER RADIOTHERAPIE, 2007, 11 (6-7): : 338 - 344
  • [22] Cyberknife® robotic stereotactic radiotherapy: technical aspects and recent developments
    Thariat, J.
    Marcie, S.
    Marcy, P. -Y.
    Trimaud, R.
    Angellier, G.
    Mammar, H.
    Bondiau, P. -Y.
    Gerard, J. -P.
    BULLETIN DU CANCER, 2010, 97 (07) : 807 - 818
  • [23] Image guidance quality assurance of a G4 CyberKnife robotic stereotactic radiosurgery system
    Pantelis, E.
    Petrokokkinos, L.
    Antypas, C.
    JOURNAL OF INSTRUMENTATION, 2009, 4
  • [24] Accuracy of Robotic Radiosurgical Liver Treatment Throughout the Respiratory Cycle
    Winter, Jeff D.
    Wong, Raimond
    Swaminath, Anand
    Chow, Tom
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (04): : 916 - 924
  • [25] The cyberknife: A frameless robotic system for radiosurgery
    Adler, JR
    Chang, SD
    Murphy, MJ
    Doty, J
    Geis, P
    Hancock, SL
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 69 (1-4) : 124 - 128
  • [26] DOSIMETRIC ACCURACY OF CYBERKNIFE STEREOTACTIC RADIOSURGERY FOR BENIGN PERIOPTIC TUMOR
    Yoon, K.
    Jeong, C.
    Park, M.
    Goh, Y.
    Kim, S.
    Cho, B.
    Kwak, J.
    Cho, Y.
    NEURO-ONCOLOGY, 2023, 25
  • [27] The CyberKnife® Robotic Radiosurgery System in 2010
    Kilby, W.
    Dooley, J. R.
    Kuduvalli, G.
    Sayeh, S.
    Maurer, C. R., Jr.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2010, 9 (05) : 433 - 452
  • [28] An Analysis of CyberKnife Prediction Accuracy
    Pepin, E.
    Wu, H.
    Zhang, Y.
    Lord, B.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [29] Treatment of hepatic metastases of colorectal cancer by robotic stereotactic radiation (Cyberknife ®)
    Peiffert, D.
    Baumann, A. -S.
    Marchesi, V.
    JOURNAL OF VISCERAL SURGERY, 2014, 151 : S45 - S49
  • [30] Unresectable Foramen Magnum Meningioma Treated With CyberKnife Robotic Stereotactic Radiosurgery
    Malone, Janna
    Gaviolli, Eduardo
    Doody, Janice
    Sinclair, John
    Malone, Shawn
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2020, 12 (06)