Research of Low Resistance Calibration Method

被引:0
|
作者
Huang Xiaoding [1 ]
Cai Jianzhen [1 ]
机构
[1] Beijing Orient Inst Measurement & Test, Beijing 100086, Peoples R China
关键词
Low resistance; simulated resistance; calibration; DC current comparator bridge;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this paper, the low resistance calibration methods are introduced. We have analyzed some technical difficulties of low resistance calibration in this paper. A simulated resistance calibration method is suggested. The uncertainty of 10 mu Omega resistance with this method is estimated as 2x10(-4) (k=2). We also put forward the technique for spot test.
引用
收藏
页码:1541 / 1543
页数:3
相关论文
共 50 条
  • [21] A Method of A Method of Low-Cost IMU Calibration and Alignment
    Lv, Jixin
    Ravankar, Ankit A.
    Kobayashi, Yukinori
    Emaru, Takanori
    2016 IEEE/SICE INTERNATIONAL SYMPOSIUM ON SYSTEM INTEGRATION (SII), 2016, : 373 - 378
  • [22] Resistance Calibration Method Without External Precision Elements
    Melikyan, Vazgen
    Sahakyan, Arthur
    Piloyan, Mikayel
    Hovhannisyan, Gagik
    Shishmanyan, Aram
    Hovhannisyan, Taron
    Trdatyan, Davit
    2014 IEEE EAST-WEST DESIGN & TEST SYMPOSIUM (EWDTS), 2014,
  • [23] Research on the calibration method of radio interference measurement result at the low-voltage side of corona cage
    Wang, Wenzhuo
    Guo, Jian
    Zhang, Bo
    He, Jinliang
    Gaodianya Jishu/High Voltage Engineering, 2015, 41 (03): : 1042 - 1048
  • [24] Research on Calibration Method of Infrared Test Equipment (Calibration Method of Infrared Radiometer and Infrared Thermal Imager)
    Guan, Wei
    Guo, Ling
    Fu, Li
    26TH CHINESE CONTROL AND DECISION CONFERENCE (2014 CCDC), 2014, : 1863 - 1868
  • [25] Research on calibration technology of ship model resistance in towing tank
    Jin, Zhongjia
    Liu, Sheng
    Wu, Baoshan
    Qiu, Chengchang
    Wei, Naxin
    Ship Building of China, 2015, 56 (03) : 162 - 171
  • [26] Experimental and numerical research on a novel underwater launching method with high efficiency and low resistance
    Zhang, Xinwei
    Yu, Yonggang
    AIP ADVANCES, 2021, 11 (02)
  • [27] CALIBRATION OF INSTRUMENTS COUPLED TO LOW-IMPEDANCE RESISTANCE THERMOMETERS
    BOZHENKO, LI
    ZANICHKOVSKAYA, LV
    KOCHAN, VA
    LAKH, VI
    PAVLIKEVICH, AN
    MEASUREMENT TECHNIQUES-USSR, 1971, 14 (06): : 932 - +
  • [28] Research of dataglove calibration method based on genetic algorithms
    Sun, Zhongsheng
    Bao, Gang
    Li, Jun
    Wang, Zuwen
    WCICA 2006: SIXTH WORLD CONGRESS ON INTELLIGENT CONTROL AND AUTOMATION, VOLS 1-12, CONFERENCE PROCEEDINGS, 2006, : 305 - 305
  • [29] Research on calibration method for space rover vision system
    Shenyang Institute of Automation, Chinese Academy of Sciences, Shenyang 110016, China
    Yi Qi Yi Biao Xue Bao, 2008, SUPPL. 2 (417-420):
  • [30] Research on the radiation calibration method of infrared target simulator
    Zhang, Jianping
    Yang, Ciyin
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2013, 34 (09): : 2045 - 2050