Optimal Configuration Design of MRI REBCO Magnet Taking into Account Superconducting Layer

被引:0
|
作者
Noguchi, So [1 ]
Miyazaki, Hiroshi [2 ]
Tosaka, Taizo [2 ]
Nomura, Shunji [2 ]
Kurusu, Tsutomu [2 ]
Ueda, Hiroshi [3 ]
Ishiyama, Atsushi [4 ]
Urayama, Shinichi [5 ]
Fukuyama, Hidenao [5 ]
机构
[1] Hokkaido Univ, Grad Sch Informat Sci & Technol, Sapporo, Hokkaido, Japan
[2] Toshiba Co Ltd, Yokohama, Kanagawa, Japan
[3] Okayama Univ, Grad Sch Nat Sci & Technol, Okayama, Japan
[4] Waseda Univ, Sch Adv Sci & Engn, Tokyo, Japan
[5] Kyoto Univ, Grad Sch Med, Kyoto, Japan
关键词
Field homogeneity; MRI; optimal design; REBCO coated conductor;
D O I
暂无
中图分类号
TP301 [理论、方法];
学科分类号
081202 ;
摘要
The development of ultra-high-field (similar to 10 T magnetic resonance imaging (MRI) device is underway. It is, therefore, required to achieve the extremely high field homogeneity at a magnet design stage, where the field homogeneity is evaluated using the Legendre polynomials. Commonly, it is supposed that the current uniformly flows on the cross section of coils. However, for REBCO magnet, the no-current-flowing region has to be taken into account, such as a copper stabilizer. In addition, the REBCO layer inside REBCO tape is very thin, 1 - 2 mu m, so the current-flowing region is 1/100 of the REBCO tape area. Therefore, it is necessary to consider that the current flows only in the REBCO layer, for accurate evaluation of field homogeneity.
引用
收藏
页数:1
相关论文
共 50 条
  • [31] Evaluation of influence of external magnetic disturbance on superconducting magnet taking conservation of linkage flux into account
    Takechi, M
    Kuroda, S
    ELECTRICAL ENGINEERING IN JAPAN, 1996, 116 (02) : 112 - 123
  • [33] OPTIMAL ECONOMICAL DESIGN TAKING ACCOUNT OF STATIONARY RANDOM FACTORS
    EVSTIGNEEV, IV
    DOKLADY AKADEMII NAUK SSSR, 1972, 206 (05): : 1040 - +
  • [34] Optimal reactor design and operation taking catalyst deactivation into account
    Xie, Mingquan
    Freund, Hannsjoerg
    CHEMICAL ENGINEERING SCIENCE, 2018, 175 : 405 - 415
  • [35] Pareto optimal design of sectored toroidal superconducting magnet for SMES
    Bhunia, Uttam
    Saha, Subimal
    Chakrabarti, Alok
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2014, 505 : 6 - 13
  • [36] Multiple Layer Superconducting Magnet Design for Magnetic Resonance Imaging
    Wang, Chunzhong
    Wang, Qiuliang
    Zhang, Quan
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2010, 20 (03) : 706 - 709
  • [37] Design of a Cooling System for a REBCO Insert Coil in a Cryogen-Free 25 T Superconducting Magnet
    Iwai, S.
    Takahashi, M.
    Miyazaki, H.
    Tosaka, T.
    Tasaki, K.
    Hanai, S.
    Ioka, S.
    Watanabe, K.
    Awaji, S.
    Oguro, H.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2015, 25 (03)
  • [38] Design and Performance Analysis of Gradient Coil for High-Field HTS REBCO MRI Magnet
    Park, Hyunsoo
    Kim, Geonyoung
    Choi, Kibum
    Park, Jeonghwan
    Kim, Jaemin
    Choi, Seyong
    Sim, Kideok
    Hahn, Seungyong
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2024, 34 (05) : 1 - 5
  • [39] Design, Fabrication, and Test of a 12 T REBCO Insert for a 27 T All-Superconducting Magnet
    Liu, Jianhua
    Wang, Lei
    Qin, Lang
    Wang, Qiuliang
    Dai, Yinming
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2020, 30 (05)
  • [40] A study on optimal coil configurations in a split-type superconducting MRI magnet
    Kakugawa, S
    Hino, N
    Komura, A
    Kitamura, M
    Takeshima, H
    Honmei, T
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 366 - 369