Virtual patient 3D dose reconstruction using in air EPID measurements and a back-projection algorithm for IMRT and VMAT treatments

被引:37
|
作者
Olaciregui-Ruiz, Igor [1 ]
Rozendaal, Roel [1 ]
van Oers, Rene F. M. [1 ]
Mijnheer, Ben [1 ]
Mans, Anton [1 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Plesmanlaan 121, NL-1066 CX Amsterdam, Netherlands
关键词
EPID dosimetry; Pre-treatment; In vivo; Verification; IMRT; VMAT; PRETREATMENT VERIFICATION; PASSING RATES; GAMMA INDEX; DOSIMETRY; QA; THERAPY;
D O I
10.1016/j.ejmp.2017.04.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: At our institute, a transit back-projection algorithm is used clinically to reconstruct in vivo patient and in phantom 3D dose distributions using EPID measurements behind a patient or a polystyrene slab phantom, respectively. In this study, an extension to this algorithm is presented whereby in air EPID measurements are used in combination with CT data to reconstruct 'virtual' 3D dose distributions. By combining virtual and in vivo patient verification data for the same treatment, patient-related errors can be separated from machine, planning and model errors. Methods and materials: The virtual back-projection algorithm is described and verified against the transit algorithm with measurements made behind a slab phantom, against dose measurements made with an ionization chamber and with the OCTAVIUS 4D system, as well as against TPS patient data. Virtual and in vivo patient dose verification results are also compared. Results: Virtual dose reconstructions agree within 1% with ionization chamber measurements. The average c-pass rate values (3% global dose/3 mm) in the 3D dose comparison with the OCTAVIUS 4D system and the TPS patient data are 98.5 +/- 1.9%(1SD) and 97.1 +/- 2.9%(1SD), respectively. For virtual patient dose reconstructions, the differences with the TPS in median dose to the PTV remain within 4%. Conclusions: Virtual patient dose reconstruction makes pre-treatment verification based on deviations of DVH parameters feasible and eliminates the need for phantom positioning and re-planning. Virtual patient dose reconstructions have additional value in the inspection of in vivo deviations, particularly in situations where CBCT data is not available (or not conclusive). (C) 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:49 / 57
页数:9
相关论文
共 50 条
  • [21] 3D Tomography Back-Projection Parallelization on Intel FPGAs Using OpenCL
    Maxime Martelli
    Nicolas Gac
    Alain Mérigot
    Cyrille Enderli
    Journal of Signal Processing Systems, 2019, 91 : 731 - 743
  • [22] Fluorescent optical imaging of small animals using filtered back-projection 3D surface reconstruction method
    Hu, Gang
    Yao, Junjie
    BMEI 2008: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOL 2, 2008, : 76 - 80
  • [23] On the Sensitivity of An EPID-Based 3D Dose Verification System to Detect Delivery Errors in VMAT Treatments
    Gonzalez, P.
    Olaciregui-Ruiz, I.
    Mijnheer, B.
    Mans, A.
    Rozendaal, R.
    MEDICAL PHYSICS, 2016, 43 (06) : 3633 - 3633
  • [24] 3D dose verification for stereotactic lung and spine treatments using the EPID
    McCowan, Peter
    vanBeek, Timothy
    van Uytven, Eric
    McCurdy, Boyd
    MEDICAL PHYSICS, 2017, 44 (08) : 4390 - 4390
  • [25] An Efficient Method of 3D Patient Dose Reconstruction Based On EPID Measurements for Pre-Treatment Patient Specific QA
    David, R.
    Zwan, B.
    Hindmarsh, J.
    Seymour, E.
    Kandasamy, K.
    Arthur, G.
    Lee, C.
    Greer, P.
    MEDICAL PHYSICS, 2016, 43 (06) : 3718 - 3718
  • [26] An EPID-based 3D patient dose verification method for SBRT-VMAT delivery
    McCowan, P.
    vanUytven, E.
    vanBeek, T.
    McCurdy, B.
    MEDICAL PHYSICS, 2014, 41 (08) : 2 - 2
  • [27] In Vivo 3D Dose Verification for IMRT Using Electronic Portal Imaging Device (EPID)
    Chang, C.
    Anderson, N. G.
    Zhu, T. C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [28] Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma
    Lin H.
    Huang S.
    Deng X.
    Zhu J.
    Chen L.
    Radiation Oncology, 9 (1)
  • [29] 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT
    McDermott, Leah N.
    Wendling, Markus
    Nijkamp, Jasper
    Mans, Anton
    Sonke, Jan-Jakob
    Mijnheer, Ben J.
    van Herk, Marcel
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 35 - 42
  • [30] Investigating the effectiveness of monitoring relevant variations during IMRT and VMAT treatments by EPID-based 3D in vivo verification performed using planning CTs
    Li, Yinghui
    Zhu, Jinhan
    Shi, Jinping
    Chen, Lixin
    Liu, Xiaowei
    PLOS ONE, 2019, 14 (06):