Assessment of Reproducibility of Laser Electrospray Mass Spectrometry using Electrospray Deposition of Analyte

被引:6
|
作者
Sistani, Habiballah
Karki, Santosh
Archer, Jieutonne J.
Shi, Fengjian
Levis, Robert J. [1 ]
机构
[1] Temple Univ, Dept Chem, Philadelphia, PA 19122 USA
基金
美国国家科学基金会;
关键词
Femtosecond laser vaporization; Electrospray deposition; Quantitative analysis; QUANTITATIVE-ANALYSIS; SAMPLE PREPARATION; IONIZATION; DESORPTION/IONIZATION; VAPORIZATION; MALDI; MACROMOLECULES; SUPPRESSION; MIXTURES; SURFACES;
D O I
10.1007/s13361-017-1622-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A nonresonant, femtosecond (fs) laser is employed to desorb samples of Victoria blue deposited on stainless steel or indium tin oxide (ITO) slides using either electrospray deposition (ESD) or dried droplet deposition. The use of ESD resulted in uniform films of Victoria blue whereas the dried droplet method resulted in the formation of a ring pattern of the dye. Laser electrospray mass spectrometry (LEMS) measurements of the ESD-prepared films on either substrate were similar and revealed lower average relative standard deviations for measurements within-film (20.9%) and between-films (8.7%) in comparison to dried droplet (75.5% and 40.2%, respectively). The mass spectral response for ESD samples on both substrates was linear (R-2 > 0.99), enabling quantitative measurements over the selected range of 7.0 x 10(-11) to 2.8 x 10(-9) mol, as opposed to the dried droplet samples where quantitation was not possible (R-2 = 0.56). The limit of detection was measured to be 210 fmol.
引用
收藏
页码:880 / 886
页数:7
相关论文
共 50 条
  • [41] ELEMENTAL SPECIATION MEASUREMENTS WITH ELECTROSPRAY MASS-SPECTROMETRY - AN ASSESSMENT
    AGNES, GR
    STEWART, II
    HORLICK, G
    APPLIED SPECTROSCOPY, 1994, 48 (11) : 1347 - 1359
  • [42] ASSESSMENT OF METALS IN RECONSTITUTED METALLOTHIONEINS BY ELECTROSPRAY MASS-SPECTROMETRY
    YU, XL
    WOJCIECHOWSKI, M
    FENSELAU, C
    ANALYTICAL CHEMISTRY, 1993, 65 (10) : 1355 - 1359
  • [43] Dual Laser and Desorption Electrospray Ionization Mass Spectrometry Imaging Using the Same Interface
    Katz, Lauren
    Woolman, Michael
    Talbot, Francis
    Amara-Belgadi, Siham
    Wu, Megan
    Tortorella, Sara
    Das, Sunit
    Ginsberg, Howard J.
    Zarrine-Afsar, Arash
    ANALYTICAL CHEMISTRY, 2020, 92 (09) : 6349 - 6357
  • [44] Charge reduction electrospray mass spectrometry
    Scalf, M
    Westphall, MS
    Smith, LM
    ANALYTICAL CHEMISTRY, 2000, 72 (01) : 52 - 60
  • [45] Electrospray tandem mass spectrometry of aminochromes
    Lemos-Amado, F
    Domingues, P
    Ferrer-Correia, A
    Remiao, F
    Milhazes, N
    Borges, F
    Carvalho, FD
    Bastos, ML
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (24) : 2466 - 2471
  • [46] Electrospray tandem mass spectrometry of lexitropsins
    Domingues, MRM
    Domingues, P
    Gajewski, M
    Czuchajowski, L
    Ferrer-Correia, AJ
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (12) : 908 - 914
  • [47] Electrospray tandem mass spectrometry of epipolythiodioxopiperazines
    Wu, Zhi-Jun
    Li, Guo-You
    Fang, Dong-Mei
    Qi, Hua-Yi
    Ren, Wei-Jian
    Zhang, Guo-Lin
    JOURNAL OF MASS SPECTROMETRY, 2007, 42 (06): : 749 - 760
  • [48] Application of electrospray mass spectrometry in biology
    Pitt, AR
    NATURAL PRODUCT REPORTS, 1998, 15 (01) : 59 - 72
  • [49] Electrospray tandem mass spectrometry of nucleotides
    Boschenok, J
    Sheil, MM
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1996, 10 (01) : 144 - 149
  • [50] Miniaturization of electrospray ionization mass spectrometry
    Wood, TD
    Moy, MA
    Dolan, AR
    Bigwarfe, PM
    White, TP
    Smith, DR
    Higbee, DJ
    APPLIED SPECTROSCOPY REVIEWS, 2003, 38 (02) : 187 - 244