Comparison of three methods of extravascular lung water volume measurement in patients after cardiac surgery

被引:5
|
作者
Maddison, Benjamin [1 ]
Wolff, Christopher [1 ]
Findlay, George [2 ]
Radermacher, Peter [3 ]
Hinds, Charles [1 ]
Pearse, Rupert M. [1 ]
机构
[1] Queen Marys Univ London, Royal London Hosp, Barts & London Sch Med & Dent, London E1 1BB, England
[2] Univ Wales Hosp, Intens Care Unit, Cardiff CF14 4XW, S Glam, Wales
[3] Univ Ulm Klinikum, Sekt Anasthesiol Pathophysiol & Verfahrensentwick, D-89081 Ulm, Germany
来源
CRITICAL CARE | 2009年 / 13卷 / 04期
关键词
CRITICALLY ILL PATIENTS; POSTMORTEM GRAVIMETRY; OUTPUT MEASUREMENT; THERMAL DILUTION; INDICATOR; SINGLE; THERMODILUTION; BLOOD; CIRCULATION; MANAGEMENT;
D O I
10.1186/cc7948
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction Measurement of extravascular lung water (EVLW) by using the lithium-thermal (Li-thermal) and single-thermal indicator dilution methods was compared with the indocyanine green-thermal (ICG-thermal) method in humans. Methods Single-center observational study involving patients undergoing cardiac surgery with cardiopulmonary bypass. Paired measurements were taken 1, 2, 4, and 6 hours after surgery. Bland-Altman analysis was used to calculate bias and limits of agreement. Data are presented as mean (SD) or median (IQR). Results Seventeen patients were recruited (age, 69 years ( 54 to 87 years); Parsonnet score 10 (0 to 29)). Sixteen ICG-thermal measurements were excluded after blinded assessment because of poor-quality indicator dilution curves. EVLW volume as measured by the ICG-thermal technique was 4.6 (1.9) ml/kg, compared with 5.3 (1.4) ml/kg for the single-thermal method. Measurements taken with the Li-thermal method were clearly erroneous (-7.6 (7.4) ml/kg). In comparison with simultaneous measurements with the ICG-thermal method, single-thermal measurements had an acceptable degree of bias, but limits of agreement were poor (bias, -0.3 ml/kg (2.3)). Li-thermal measurements compared poorly with the ICG-thermal reference method (bias, 13.2 ml/kg (14.4)). Conclusions The principal finding of this study was that the prototype Li-thermal method did not provide reliable measurements of EVLW volume when compared with the ICG-thermal reference technique. Although minimal bias was associated with the single-thermal method, limits of agreement were approximately 45% of the normal value of EVLW volume. The Li-thermal method performed very poorly because of the overestimation of mean indicator transit time by using an external lithium ion electrode. These findings suggest that the assessment of lung water content by lithium-indicator dilution is not sufficiently reliable for clinical use in individual patients.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] BLOOD-VOLUME AND EXTRAVASCULAR WATER VOLUME IN CAT LUNG
    KARLSEN, J
    BO, G
    AARSETH, P
    MICROVASCULAR RESEARCH, 1973, 6 (02) : 257 - 257
  • [32] MEASUREMENT OF REGIONAL EXTRAVASCULAR LUNG WATER IN PATIENTS WITH HEART-FAILURE
    BOSSALLER, C
    SCHOBER, O
    MEYER, GJ
    HUNDESHAGEN, H
    LICHTLEN, PR
    ZEITSCHRIFT FUR KARDIOLOGIE, 1984, 73 (02): : 81 - 88
  • [33] Volume assessment in patients with necrotizing pancreatitis: A comparison of intrathoracic blood volume index, central venous pressure, and hematocrit, and their correlation to cardiac index and extravascular lung water index
    Huber, Wolfgang
    Umgelter, Andreas
    Reindl, Wolfgang
    Franzen, Michael
    Schmidt, Christian
    von Delius, Stefan
    Geisler, Fabian
    Eckel, Florian
    Fritsch, Ralph
    Siveke, Jens
    Henschel, Benedikt
    Schmid, Roland M.
    CRITICAL CARE MEDICINE, 2008, 36 (08) : 2348 - 2354
  • [34] MEASUREMENT OF EXTRAVASCULAR LUNG WATER - A TOY OR A TOOL
    NEIDHART, P
    SUTER, PM
    ANAESTHESIST, 1986, 35 (09): : 559 - 562
  • [35] Extravascular Lung Water Measurement: Proper Indexing
    Leibowitz, Andrew B.
    CRITICAL CARE MEDICINE, 2013, 41 (04) : 1143 - 1144
  • [36] EXTRAVASCULAR LUNG WATER MEASUREMENT IN SEPTIC SHEEP
    ANDREASSON, S
    BYLOCK, A
    SMITH, L
    RISBERG, B
    JOURNAL OF SURGICAL RESEARCH, 1986, 40 (02) : 95 - 104
  • [37] How important is the measurement of extravascular lung water?
    Fernandez-Mondejar, Enrique
    Guerrero-Lopez, Francisco
    Colmenero, Manuel
    CURRENT OPINION IN CRITICAL CARE, 2007, 13 (01) : 79 - 83
  • [38] Effects of Extravascular Lung Water on the Measurement of Transpulmonary Thermodilution Cardiac Output in Acute Respiratory Distress Syndrome Patients
    Huang, Chung-Chi
    Kao, Kuo-Chin
    Fu, Jui-Ying
    Hsieh, Meng-Jer
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (03) : 481 - 485
  • [39] Intravascular volume monitoring and extravascular lung water in septic patients with pulmonary edema
    Sandrine Boussat
    Thierry Jacques
    Bruno Levy
    Evelyne Laurent
    Antoine Gache
    Gilles Capellier
    Alain Neidhardt
    Intensive Care Medicine, 2002, 28 : 712 - 718
  • [40] Intravascular volume monitoring and extravascular lung water in septic patients with pulmonary edema
    Boussat, S
    Jacques, T
    Levy, B
    Laurent, E
    Gache, A
    Capellier, G
    Neidhardt, A
    INTENSIVE CARE MEDICINE, 2002, 28 (06) : 712 - 718